1.
J Appl Clin Med Phys ; 25(4): e14315, 2024 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-38415897

RESUMEN

AIM: To report on the performance characteristics of the 5-ring GE Discovery MI PET/CT systems using the AAPM TG-126 report and compare these results to NEMA NU 2-2012 where applicable. MATERIALS AND METHODS: TG-126 testing was performed on two GE 5-Rings Discovery MI scanners. Tests performed included spatial resolution, PET/CT image-registration accuracy, sensitivity, count rate performance, accuracy of corrections, image contrast, scatter/attenuation correction, and image uniformity. All acquired data were analyzed using scanner console or free software tools as described by TG-126 and the results were then compared to published NEMA NU 2-2012 values. RESULTS: Both scanners gave similar resolution results for TG-126 and NEMA NU 2-2012 and were within manufacturer specifications. Image-registration accuracy between PET and CT using our clinical protocol showed excellent results with values ≤1 mm. Sensitivity using TG-126 was 19.43 cps/kBq while for NEMA the value was 20.73 cps/kBq. The peak noise-equivalent counting rate was 2174 kcps at 63.1 kBq/mL and is not comparable to NEMA NU 2-2012 due to differences in phantoms and methods used to measure and calculate this parameter. The accuracy of corrections for count losses for TG-126 were expressed in SUV values and found to be within 10% of the expected SUV measurement of 1. Image contrast and scatter/attenuation correction using the TG-126 method gave acceptable results. Image uniformity assessment resulted in values within the recommended ± 5% limits. CONCLUSION: These results show that the 5-ring GE Discovery MI PET/CT scanner testing using TG-126 is reproducible and has similar results to NEMA NU 2-2012 tests where applicable. We hope these results start to form the basis to compare PET/CT systems using TG-126.


Asunto(s)
Tomografía Computarizada por Tomografía de Emisión de Positrones , Tomografía de Emisión de Positrones , Humanos , Tomografía de Emisión de Positrones/métodos , Tomógrafos Computarizados por Rayos X , Fantasmas de Imagen , Programas Informáticos
2.
Eur J Nucl Med Mol Imaging ; 49(9): 3023-3032, 2022 07.
Artículo en Inglés | MEDLINE | ID: mdl-35284970

RESUMEN

PURPOSE: The DMI PET/CT is a modular silicon photomultiplier-based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an additional 6th ring, extending the axial FOV to 30 cm. We report on the performance evaluation of the 6-ring upgraded Generation 2 (Gen2) system while values are also reported for the 5-ring configuration of the very same system prior to the upgrade. METHODS: PET performance was evaluated using the NEMA NU 2-2018 standard for spatial resolution, sensitivity, image quality, count rate performance, timing resolution, and image co-registration accuracy. Patient images were used to assess image quality. RESULTS: The average system sensitivity was measured at 32.76 cps/kBq (~ 47% increase to 5 rings at 22.29 cps/kBq) while noise equivalent count rate peaked at 434.3 kcps corresponding to 23.6 kBq/mL (~ 60% increase to Generation 1 (Gen1) and 39% to Gen2 5 rings). Contrast recovery ranged between 54.5 and 85.8% similar to 5 rings, while the 6 rings provided lower background variability (2.3-8.5% for 5 rings vs 1.9-6.8% for 6 rings) and lower lung error (4.0% for the 5 rings and 3.16% for the 6 rings). Transverse/axial full width at half-maximum (FWHM) at 1 cm (3.79/4.26 mm) and 10 cm (4.29/4.55 mm), scatter fraction (40.2%), energy resolution (9.63%), and time-of-flight (TOF) resolution (389.6 ps at 0 kBq/mL) were in line to previously reported values measured across different system configurations. Improved patient image quality is obtained with the 6 rings compared to the 5 rings, while image quality is retained even at reduced scan times, enabling WB dynamic acquisitions. CONCLUSIONS: The higher sensitivity of the 6-ring DMI compared to the 5-ring configuration may lead to improved image quality of clinical images at reduced scan time. Additionally, it could equally be used to allow improved temporal sampling and/or reduced overall scan time in dynamic acquisitions. Conversely, temporal sampling and scan time could be traded per application to further drive injected dose at lower levels.


Asunto(s)
Tomografía Computarizada por Tomografía de Emisión de Positrones , Tomografía de Emisión de Positrones , Humanos , Fantasmas de Imagen , Tomografía de Emisión de Positrones/métodos , Tórax
3.
Eur J Nucl Med Mol Imaging ; 48(6): 1726-1735, 2021 06.
Artículo en Inglés | MEDLINE | ID: mdl-33388972

RESUMEN

PURPOSE: In this paper, we aimed to evaluate the positron emission tomography (PET) performance of, to the best of our knowledge, the third commercially available whole-body integrated PET/magnetic resonance (MR) system. METHODS: The PET system performance was measured following the NEMA standards with and without simultaneous MR operation. PET spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses and random corrections, image quality, and time-of-flight (TOF) resolution were quantitatively evaluated. Clinical scans were acquired at the PET/MR system and compared with images acquired at a PET/CT with the same digital detector technology. RESULTS: Measurement results of essential PET performance were reported in the form of MR idle (MR pulsing). The axial, radial, and tangential spatial resolutions were measured as 2.72 mm (2.73 mm), 2.86 mm (2.85 mm), and 2.81 mm (2.82 mm) FWHM, respectively, at 1 cm radial offset. The NECR peak was measured as 129.2 kcps (129.5 kcps) at 14.7 kBq mL-1 (14.2 kBq mL-1). The scatter fraction at NECR peak was 37.9% (36.5%), and the maximum slice error below NECR was 4.1% (4.5%). Contrast recovery coefficients ranged from 51.8% (52.3%) for 10 mm hot sphere to 87.3% (87.2%) for 37 mm cold sphere. TOF resolution at 5.3 kBq mL-1 was measured at 535 ps (540 ps). With point source, TOF was measured to be 474 ps (485 ps). Clinical scans revealed similar image quality from the PET/MR and the comparative PET/CT system. CONCLUSION: The PET performance of the newly introduced integrated PET/MR system is not significantly affected by the simultaneous operation of an MR sequence (2-point DIXON sequence). Measurement results demonstrate comparable performance with other state-of-the-art PET/MR systems. The clinical benefits of high spatial resolution and long axial coverage remain to be further evaluated in specific clinical imaging applications.


Asunto(s)
Tomografía Computarizada por Tomografía de Emisión de Positrones , Tomografía de Emisión de Positrones , Humanos , Espectroscopía de Resonancia Magnética , Fantasmas de Imagen , Estándares de Referencia
4.
Z Med Phys ; 2024 Feb 09.
Artículo en Inglés | MEDLINE | ID: mdl-38341373

RESUMEN

PURPOSE: The Monte Carlo method is an effective tool to simulate and verify PET systems. Furthermore, it can help in the design and optimization of new medical imaging devices and algorithms. In this framework, the goal of this work is to verify the GATE toolkit performance when applied to simulate two Siemens Healthineers PET scanners: a standard axial field-of-view Biograph Vision scanner and the new long axial field-of-view Biograph Vision Quadra scanner. METHODS: The simulation toolkit GATE is based on GEANT4, comprising its main functionalities and a set of domain-specific features in the field of medical physics. To accomplish our purpose, the guidelines described in the NEMA NU 2-2018 protocol are reproduced. Then the simulated results are compared to experimental data available in the literature for both PET scanners. The assessment of the models includes different studies of sensitivity, count rate performances, spatial resolution and image quality. These tests are intended to evaluate the image quality of PET devices. RESULTS: In the spatial resolution test, relative errors lower than 8% are obtained between the experiments and GATE models. The sensitivity is 17.2 cps/kBq (Vision) and 175.9 cps/kBq (Quadra), representing relative differences with the experiment of 6% and 0.3%, respectively. Deviations from peak NECR are less than 9%. In the Image Quality test, the contrast recovery coefficient for hot spheres, with 8 iterations and 5 subsets, ranges between 57-83% for Vision and 54-86% for Quadra. These values represent a maximum deviation between the simulations and the experiments of 10% for the Quadra scanner. In the case of the Vision scanner, the highest difference is observed for the 10 mm sphere (∼38%) due to the higher contrast recovery of the experiment caused by the Gibbs artifact from the use of PSF reconstruction. CONCLUSIONS: The results of the simulations have provided evidence of a good agreement between the experimental data and the results obtained with GATE. Thus, this work supports the capability of this MC toolkit to accurately simulate the models of the Vision and Quadra scanners. This study has laid the basis for further research in this field and has identified several areas that could be explored.

5.
World J Nucl Med ; 23(1): 10-16, 2024 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-38595839

RESUMEN

Aim The objective of this study includes the NEMA (National Electrical Manufacturer Association) NU2-2018 performance evaluation of the uMIvista PET-CT (positron emission tomography-computed tomography) system. Methods The latest NEMA NU2-2018 guidelines have been followed for the evaluation of performance parameters of this PET-CT scanner: axial, tangential, and radial spatial resolution, sensitivity, counting losses, scatter, randomness, random and counting loss correction, image quality, time and energy resolution, image uniformity, and image registration alignment post installation of country first uMIvista PET-CT. Results The measured NEMA sensitivity of the uMIvista PET scanner was 12.053 cps/kBq. The spatial resolutions of the PET were measured as tangential, radial, and transaxial spatial resolutions at 10 mm, with 3.01 mm, 2.95 mm, and 2.93 mm, respectively; at 100 mm, with 3.17 mm, 3.42 mm, and 3.05 mm, respectively; and at 200 mm, with 3.65 mm, 4.54 mm, and 3.17 mm, respectively, at full-width half-maximum (FWHM); while at full-width tenths-maximum (FWTM), the values at 10 mm were 5.79 mm, 5.57 mm, and 5.69 mm, respectively, and at 100 mm were 5.59 mm, 5.96 mm, and 5.91 mm, respectively. The measured time-of-flight (TOF) timing resolution was 302.294 ps and the measured energy resolution was 11.76% with FWHM and FWTM. Conclusion The NEMA NU2-2018 performances of this TOF-integrated digital PET-CT system are extremely good in all parameters.

6.
Phys Med Biol ; 68(21)2023 10 26.
Artículo en Inglés | MEDLINE | ID: mdl-37802064

RESUMEN

Objective.Contrast recovery coefficient (CRC) is essential for image quality (IQ) assessment in positron emission tomography (PET), typically measured according to the National Electrical Manufacturers Association (NEMA) NU 2 standard. This study quantifies systematic uncertainties of the CRC measurement by a numerical investigation of the effects from scanner-independent parameters like voxel size, region-of-interest (ROI) misplacement, and sphere position on the underlying image grid.Approach.CRC measurements with 2D and 3D ROIs were performed on computer-generated images of a NEMA IQ-like phantom, using voxel sizes of 1-4 mm for sphere diameters of 5-40 mm-first in absence of noise and blurring, then with simulated spatial resolution and image noise with varying noise levels. The systematic uncertainties of the CRC measurement were quantified from above variations of scanner-independent parameters. Subsampled experimental images of a NEMA IQ phantom were additionally used to investigate the impact of ROI misplacement at different noise levels.Main results.In absence of noise and blurring, systematic uncertainties were up to 28.8% and 31.0% with 2D and 3D ROIs, respectively, for the 10 mm sphere, with the highest impact from ROI misplacement. In all cases, smaller spheres showed higher uncertainties with larger voxels. Contrary to prior assumptions, the use of 3D ROIs did not exhibit less susceptibility for parameter changes. Experimental and computer-generated images both demonstrated considerable variations on individual CRC measurements when background coefficient-of-variation exceeded 20%, despite negligible effects on mean CRC.Significance.This study underscores the effect of scanner-independent parameters on reliability, reproducibility, and comparability of CRC measurements. Our findings highlight the trade-off between the benefits of smaller voxel sizes and noise-induced CRC fluctuations, which is not considered in the current version of the NEMA IQ standards. The results furthermore warrant adjustments to the standard to accommodate the advances in sensitivity and spatial resolution of current-generation PET scanners.


Asunto(s)
Tomografía de Emisión de Positrones , Tomografía Computarizada por Rayos X , Reproducibilidad de los Resultados , Tomografía Computarizada por Rayos X/métodos , Tomografía de Emisión de Positrones/métodos , Estándares de Referencia , Fantasmas de Imagen , Procesamiento de Imagen Asistido por Computador
7.
EJNMMI Phys ; 9(1): 22, 2022 Mar 22.
Artículo en Inglés | MEDLINE | ID: mdl-35316846

RESUMEN

INTRODUCTION: A Vereos PET/CT device was adapted to be compatible with the experimentation in large animals within BSL-3 environment. The aim of this study was to investigate the impact of this modification on the performance according to NEMA NU2-2012 standard. METHODS: Spatial resolution, sensitivity, count rate performance, accuracies of corrections and image quality were assessed using the NEMA NU2-2012 standards before and after installation of a transparent poly-methyl methacrylate tube of 8 mm thickness, 680 mm diameter and 2800 mm long inside the tunnel of the system. In addition, CT performance tests were performed according to manufacturer standard procedure. RESULTS: Although the presence of the tube led to a slight decrease in sensitivity, performance measurements were in accordance with manufacturer preconisation ranges and comparable to previous performance published data. CONCLUSION: Modifications of Vereos PET/CT system allowing its use in BSL-3 conditions did not affect significantly its performance according to NEMA NU2-2012 standard. KEY POINTS: Question. Does a BSL-3 compatible modification alter Philips Vereos PET/CT performances according to NEMA NU2-2012 standards? Pertinent findings. Our Vereos PET/CT system was modified by a wall separating BSL-1 and BSL-3 sides and an 8 mm thickness PMMA tube inserted into the bore of the camera in order to extend the BSL-3 containment along the bed movement. The performances of our modified system according to NEMA NU2-2012 standards were not significantly impacted by the modifications and were in accordance with the values prescribed by the manufacturer. Implications for patients care. Our clinical PET/CT device was modified for human infectious diseases studies in Non-Human Primates. This unusual set up may then provide truly transposable data from preclinical studies into clinical application in infected patients.

8.
Med Phys ; 49(2): 1139-1152, 2022 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-34954831

RESUMEN

PURPOSE: The development of total-body PET scanners is of growing interest in the PET community. Investigation into the imaging properties of a hypothetical extended axial field-of-view (AFOV) GE Healthcare SiPM-based Discovery MI (DMI) system architecture has not yet been performed. In this work, we assessed its potential as a whole-body scanner using Monte Carlo simulations. The aim of this work was to (1) develop and validate a Monte Carlo model of a four-ring scanner and (2) extend its AFOV up to 2 m to evaluate performance gain through NEMA-based evaluation. METHODS: The DMI four-ring geometry and its pulse digitization scheme were modeled within the GATE Monte Carlo platform using published literature. The GATE scanner model was validated by comparing results against published NEMA performance measurements. Following the validation of the four-ring model, the model was extended to simulate 8-, 20-, 30-, and 40-ring systems. Spatial resolution, sensitivity, NECR, and scatter fraction were characterized with modified NEMA NU-2 2018 standards; however, the image quality measurements were not acquired due to computational limitations. Spatial resolutions were simulated for all scanner ring configurations using point sources to examine the effects of parallax errors. NEMA count rates were estimated using a standard 70 cm scatter phantom and an extended version of scatter phantom of length 200 cm with (1-800) MBq of 18 F for all scanners. Sensitivity was evaluated using NEMA methods with a 70 cm standard and a 200 cm long line source. RESULTS: The average FWHM of the radial/tangential/axial spatial resolution reconstructed with filtered back-projection at 1 and 10 cm from the scanner center were 3.94/4.10/4.41 mm and 5.29/4.89/5.90 mm for the four-ring scanner. Sensitivity was determined to be 14.86 cps/kBq at the center of the FOV for the four-ring scanner using a 70 cm line source. Sensitivity enhancement up to 21-fold and 60-fold were observed for 1 and 2 m AFOV scanners compared to four-ring scanner using a 200 cm long line source. Spatial resolution simulations in a 2 m AFOV scanner suggest a maximum degradation of ∼23.8% in the axial resolution compared to the four-ring scanner. However, the transverse resolution was found to be relatively constant when increasing the axial acceptance angle up to ±70°. The peak NECR was 212.92 kcps at 22.70 kBq/ml with a scatter fraction of 38.9% for a four-ring scanner with a 70 cm scatter phantom. Comparison of peak NECR using the 200 cm long scatter phantom relative to the four-ring scanner resulted in a NECR gain of 15 for the 20-ring and 28 for the 40-ring geometry. Spatial resolution, sensitivity, and scatter fraction showed an agreement within ∼7% compared with published measured values. CONCLUSIONS: The four-ring DMI scanner simulation was successfully validated against published NEMA measurements. Sensitivity and NECR performance of extended 1 and 2 m AFOV scanners based upon the DMI architecture were subsequently simulated. Increases in sensitivity and count-rate performance are consistent with prior simulation studies utilizing extensions of the Siemens mCT architecture and published NEMA measurements with the uEXPLORER system.


Asunto(s)
Tomografía de Emisión de Positrones , Simulación por Computador , Método de Montecarlo , Fantasmas de Imagen , Estándares de Referencia
9.
Front Physiol ; 11: 525575, 2020.
Artículo en Inglés | MEDLINE | ID: mdl-33041852

RESUMEN

NEMA characterization of PET systems is generally based on 18F because it is the most relevant radioisotope for the clinical use of PET. 18F has a half-life of 109.7 min and decays into stable 18O via ß+ emission with a probability of over 96% and a maximum positron energy of 0.633 MeV. Other commercially available PET radioisotopes, such as 82Rb and 68Ga have more complex decay schemes with a variety of prompt gammas, which can directly fall into the energy window and induce false coincidence detections by the PET scanner. METHODS: Aim of this work was three-fold: (A) Develop a GATE model of the GE Signa PET/MR to perform realistic and relevant Monte Carlo simulations (B) Validate this model with published sensitivity and Noise Equivalent Count Rate (NECR) data for 18F and 68Ga (C) Use the validated GATE-model to predict the system performance for other PET isotopes including 11C, 15O, 13N, 82Rb, and 68Ga and to evaluate the effect of a 3T magnetic field on the positron range. RESULTS: Simulated sensitivity and NECR tests performed with the GATE-model for different radioisotopes were in line with literature values. Simulated sensitivities for 18F and 68Ga were 21.2 and 19.0/kBq, respectively, for the center position and 21.1 and 19.0 cps/kBq, respectively, for the 10 cm off-center position compared to the corresponding measured values of 21.8 and 20.0 cps/kBq for the center position and 21.1 and 19.6 cps/kBq for the 10 cm off-center position. In terms of NECR, the simulated peak NECR was 216.8 kcps at 17.40 kBq/ml for 18F and 207.1 kcps at 20.10 kBq/ml for 68Ga compared to the measured peak NECR of 216.8 kcps at 18.60 kBq/ml and 205.6 kcps at 20.40 kBq/ml for18F and 68Ga, respectively. For 11C, 13N, and 15O, results confirmed a peak NECR similar to 18F with the effective activity concentration scaled by the inverse of the positron fraction. For 82Rb, and 68Ga, the peak NECR was lower than for 18F while the corresponding activity concentrations were higher. For the higher energy positron emitters, the positron range was confirmed to be tissue-dependent with a reduction of the positron range by a factor of 3 to 4 in the plane perpendicular to the magnetic field and an increased positron range along the direction of the magnetic field. CONCLUSION: Monte-Carlo simulations were used to predict sensitivity and NECR performance of GE Signa PET/MR for 18F, 15O, 13N, 11C, 82Rb, and 68Ga radioisotopes and were in line with literature data. Simulations confirmed that sensitivity and NECR were influenced by the particular decay scheme of each isotope. As expected, the positron range decreased in the direction perpendicular to the 3T magnetic field. However, this will be only partially improving the resolution properties of a clinical PET/MR system due to the limiting spatial resolution of the PET detector.

10.
EJNMMI Phys ; 7(1): 43, 2020 Jun 26.
Artículo en Inglés | MEDLINE | ID: mdl-32588139

RESUMEN

BACKGROUND: The aim of this study is to conduct physical performance evaluation on the uMI550 whole-body PET/CT system according to the NEMA NU2-2018 standard. METHODS: According to the NEMA NU2-2018, spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses and random corrections, image quality, and timing resolution were evaluated. Spatial resolution was measured by using a 22Na point source. System sensitivity was measured by inserting an 18F line-source in six concentric aluminum sleeves with varying diameters. Scatter fraction, count-rate performance, accuracy of count loss, and timing resolution were all calculated by analyzing dynamically acquired data of an 18F line-source inside a polyethylene cylinder in 20 cm diameter and 70 cm length. Image quality was assessed using a NEMA IEC body phantom with a 4:1 ratio of activity concentration of spheres to the warm background. Additionally, three patient studies were performed, with one brain scan and two whole-body scans, separately. The patient images were evaluated to get a visual first impression of uMI550. RESULTS: The tangential, radial, and axial spatial resolutions were measured as 2.91 mm, 2.98 mm, and 2.97 mm FWHM, respectively, at 1 cm radial offset. The total system sensitivity to line source at center was 10.24 cps/kBq. A NECR peak was measured as 124.4 kcps at 18.85 kBq/mL. The scatter fraction at NECR peak was 36.65%, and the maximum count-rate error at and below NEC peak was 1.55%. Contrast recovery coefficients were from 46.5 (10 mm) to 83.9% (37 mm). The timing resolution was measured as 372 ps at low count rate. CONCLUSION: NEMA NU-2 2018 testing was performed on the new SiPM-based uMI550 PET/CT system. The uMI550 shows a high-spatial resolution of less than 3 mm and a good timing resolution of 372 ps. It shows clinical significances on improving potentially diagnostic ability on small lesions.

11.
EJNMMI Phys ; 6(1): 11, 2019 Jul 04.
Artículo en Inglés | MEDLINE | ID: mdl-31273558

RESUMEN

BACKGROUND: Fully integrated PET/MR systems are being used frequently in clinical research and routine. National Electrical Manufacturers Association (NEMA) characterization of these systems is generally done with 18F which is clinically the most relevant PET isotope. However, other PET isotopes, such as 68Ga and 90Y, are gaining clinical importance as they are of specific interest for oncological applications and for follow-up of 90Y-based radionuclide therapy. These isotopes have a complex decay scheme with a variety of prompt gammas in coincidence. 68Ga and 90Y have higher positron energy and, because of the larger positron range, there may be interference with the magnetic field of the MR compared to 18F. Therefore, it is relevant to determine the performance of PET/MR for these clinically relevant and commercially available isotopes. METHODS: NEMA NU 2-2007 performance measurements were performed for characterizing the spatial resolution, sensitivity, image quality, and the accuracy of attenuation and scatter corrections for 18F, 68Ga, and 90Y. Scatter fraction and noise equivalent count rate (NECR) tests were performed using 18F and 68Ga. All phantom data were acquired on the GE Signa integrated PET/MR system, installed in UZ Leuven, Belgium. RESULTS: 18F, 68Ga, and 90Y NEMA performance tests resulted in substantially different system characteristics. In comparison with 18F, the spatial resolution is about 1 mm larger in the axial direction for 68Ga and no significative effect was found for 90Y. The impact of this lower resolution is also visible in the recovery coefficients of the smallest spheres of 68Ga in image quality measurements, where clearly lower values are obtained. For 90Y, the low number of counts leads to a large variability in the image quality measurements. The primary factor for the sensitivity change is the scale factor related to the positron emission fraction. There is also an impact on the peak NECR, which is lower for 68Ga than for 18F and appears at higher activities. CONCLUSIONS: The system performance of GE Signa integrated PET/MR was substantially different, in terms of NEMA spatial resolution, image quality, and NECR for 68Ga and 90Y compared to 18F. But these differences are compensated by the PET/MR scanner technologies and reconstructions methods.

12.
J Nucl Med ; 60(4): 561-567, 2019 04.
Artículo en Inglés | MEDLINE | ID: mdl-30361382

RESUMEN

The aim of this study was to evaluate the physical performance of the Vereos whole-body PET/CT system according to the National Electrical Manufacturers Association (NEMA) NU2-2012 standard and to compare it with other state-of-the-art PET/CT systems. Methods: Spatial resolution, sensitivity, count-rate performance, count rate accuracy, and image quality were assessed. Specifically, spatial resolution was measured using an 18F point-source. System sensitivity was calculated from acquisitions of an 18F line source inside aluminum tubes of varying thickness. Assessment of count rate performance and count rate accuracy was based on measurements of an 18F line source inside a 20-cm-diameter polyethylene cylinder. PET image quality was assessed using a NEMA IQ phantom. All measurements were performed according to the predefined and implemented NEMA NU2-2012 acquisition protocols at a clinical installation of a Vereos PET/CT system. Evaluation was performed using the software provided by the vendor. Results: The average (radial and tangential) transverse spatial resolution was 4.2, 4.5, and 5.5 mm in full width at half maximum for a 1-, 10-, and 20-cm radial offset, respectively, from the center of the field of view. Axial spatial resolution varied between 4.2 and 4.6 mm in full width at half maximum, depending on the radial source position. The average sensitivity was 5.2 cps/kBq. A peak noise-equivalent count (NEC) rate of 153.4 kcps was measured at an activity concentration of 54.9 kBq/mL. The scatter fraction at peak NEC rate was 33.9%, and the maximum count rate error at and below peak NEC rate was 6.8%. Contrast recovery coefficients varied from 54.3% (10-mm sphere) to 83.9% (22-mm sphere) for the hot spheres, and between 81.4% (28-mm sphere) and 87% (37-mm sphere) for the cold spheres for a given sphere-to-background ratio of 4:1. Conclusion: The overall performance characteristics of the Vereos PET/CT system are comparable to state-of-the-art whole-body PET/CT systems with the exception that the peak NEC rate occurs at a higher activity concentration, thus indicating the ability of the Vereos PET/CT system to cover a wider range of activities.


Asunto(s)
Tomografía Computarizada por Tomografía de Emisión de Positrones/normas , Sociedades/normas , Animales , Equipos y Suministros Eléctricos , Control de Calidad , Estándares de Referencia , Dispersión de Radiación , Relación Señal-Ruido
13.
Med Phys ; 46(7): 3025-3033, 2019 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-31069816

RESUMEN

The GE Discovery MI PET/CT system has a modular digital detector design allowing three, four, or five detector block rings that extend the axial field-of-view (FOV) from 15 to 25 cm in 5 cm increments. This study investigated the performance of the 5-ring system and compared it to 3- and 4-ring systems; the GE Discovery IQ system that uses conventional photomultiplier tubes; and the GE Signa PET/MR system that has a reduced transaxial FOV. METHODS: PET performance was evaluated at three different institutions. Spatial resolution, sensitivity, counting rate performance, accuracy, and image quality were measured in accordance with National Electrical Manufacturers Association NU 2-2012 standards. The mean energy resolution, mean timing resolution, and PET/CT subsystem alignment were also measured. Phantoms were used to determine the effects of varying acquisition time and reconstruction parameters on image quality. Retrospective patient scans were reconstructed with various scan durations to evaluate the impact on image quality. RESULTS: Results from all three institutions were similar. Radial/tangential/axial full width at half maximum spatial resolution measurements using the filtered back projection algorithm were 4.3/4.3/5.0 mm, 5.5/4.6/6.5 mm, and 7.4/5.0/6.6 mm at 1, 10, and 20 cm from the center of the FOV, respectively. Measured sensitivity at the center of the FOV (20.84 cps/kBq) was significantly higher than systems with reduced axial FOV. The peak noise-equivalent counting rate was 266.3 kcps at 20.8 kBq/ml, with a corresponding scatter fraction of 40.2%. The correction accuracy for count losses up to the peak noise-equivalent counting rate was 3.6%. For the 10-, 13-, 17-, 22-, 28-, and 37-mm spheres, contrast recoveries in the image quality phantom were measured to be 46.2%, 54.3%, 66.1%, 71.1%, 85.3%, and 89.3%, respectively. The mean energy and timing resolution were 9.55% and 381.7 ps, respectively. Phantom and patient images demonstrated excellent image quality, even at short acquisition times or low injected activity. CONCLUSION: Compared to other PET/CT models, the extended axial FOV improved the overall PET performance of the 5-ring GE Discovery MI scanner. This system offers the potential to reduce scan times or injected activities through increased sensitivity.


Asunto(s)
Tomografía Computarizada por Tomografía de Emisión de Positrones/normas , Humanos , Fantasmas de Imagen , Tomografía Computarizada por Tomografía de Emisión de Positrones/instrumentación , Estándares de Referencia
14.
Med Phys ; 45(1): 258-276, 2018 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-29091269

RESUMEN

PURPOSE: Quality control plays an increasingly important role in quantitative PET imaging and is typically performed using phantoms. The purpose of this work was to develop and validate a fully automated analysis method for two common PET/CT quality assurance phantoms: the NEMA NU-2 IQ and SNMMI/CTN oncology phantom. The algorithm was designed to only utilize the PET scan to enable the analysis of phantoms with thin-walled inserts. METHODS: We introduce a model-based method for automated analysis of phantoms with spherical inserts. Models are first constructed for each type of phantom to be analyzed. A robust insert detection algorithm uses the model to locate all inserts inside the phantom. First, candidates for inserts are detected using a scale-space detection approach. Second, candidates are given an initial label using a score-based optimization algorithm. Third, a robust model fitting step aligns the phantom model to the initial labeling and fixes incorrect labels. Finally, the detected insert locations are refined and measurements are taken for each insert and several background regions. In addition, an approach for automated selection of NEMA and CTN phantom models is presented. The method was evaluated on a diverse set of 15 NEMA and 20 CTN phantom PET/CT scans. NEMA phantoms were filled with radioactive tracer solution at 9.7:1 activity ratio over background, and CTN phantoms were filled with 4:1 and 2:1 activity ratio over background. For quantitative evaluation, an independent reference standard was generated by two experts using PET/CT scans of the phantoms. In addition, the automated approach was compared against manual analysis, which represents the current clinical standard approach, of the PET phantom scans by four experts. RESULTS: The automated analysis method successfully detected and measured all inserts in all test phantom scans. It is a deterministic algorithm (zero variability), and the insert detection RMS error (i.e., bias) was 0.97, 1.12, and 1.48 mm for phantom activity ratios 9.7:1, 4:1, and 2:1, respectively. For all phantoms and at all contrast ratios, the average RMS error was found to be significantly lower for the proposed automated method compared to the manual analysis of the phantom scans. The uptake measurements produced by the automated method showed high correlation with the independent reference standard (R2 ≥ 0.9987). In addition, the average computing time for the automated method was 30.6 s and was found to be significantly lower (P ≪ 0.001) compared to manual analysis (mean: 247.8 s). CONCLUSIONS: The proposed automated approach was found to have less error when measured against the independent reference than the manual approach. It can be easily adapted to other phantoms with spherical inserts. In addition, it eliminates inter- and intraoperator variability in PET phantom analysis and is significantly more time efficient, and therefore, represents a promising approach to facilitate and simplify PET standardization and harmonization efforts.


Asunto(s)
Algoritmos , Fluorodesoxiglucosa F18 , Reconocimiento de Normas Patrones Automatizadas/métodos , Fantasmas de Imagen , Tomografía Computarizada por Tomografía de Emisión de Positrones/instrumentación , Radiofármacos , Humanos
15.
J Nucl Med ; 58(7): 1155-1161, 2017 07.
Artículo en Inglés | MEDLINE | ID: mdl-28302761

RESUMEN

The aim of this study was to assess the physical performance of a new PET/CT system, the Discovery IQ with 5-ring detector blocks. Methods: Performance was measured using the National Electrical Manufacturers Association NU2-2012 methodology. Image quality was extended by accounting for different acquisition parameters (lesion-to-background ratios [8:1, 4:1, and 2:1] and acquisition times) and reconstruction algorithms (VUE-point HD [VPHD], VPHD with point-spread-function modeling [VPHD-S], and Q.Clear). Tomographic reconstruction was also assessed using a Jaszczak phantom. Additionally, 30 patient lesions were analyzed to account for differences in lesion volume and SUV quantification between reconstruction algorithms. Results: Spatial resolution ranged from 4.2 mm at 1 cm to 8.5 mm at 20 cm. Sensitivity measured at the center and at 10 cm was 22.8 and 20.4 kps/kBq, respectively. The noise-equivalent counting rate peak was 124 kcps at 9.1 kBq/cm3 The scatter fraction was 36.2%. The accuracy of correction for count losses and randoms was 3.9%. In the image quality test, contrast recovery for VPHD, VPHD-S, and Q.Clear ranged from 18%, 18%, and 13%, respectively (hot contrast; 10-mm sphere diameter; ratio, 2:1), to 68%, 67%, and 81%, respectively (cold contrast; 37-mm sphere diameter; ratio, 8:1). Background variability ranged from 3.4%, 3.0%, and 2.1%, respectively (ratio, 2:1), to 5.5%, 4.8%, and 3.7%, respectively (ratio, 8:1). On Q.Clear reconstruction, the decrease in the penalty term (ß) increased the contrast recovery coefficients and background variability. With the Jaszczak phantom, image quality increased overall when a reconstruction algorithm modeling the point-spread function was used, and use of Q.Clear increased the signal-to-noise ratio. Lesions analyzed using VPHD-S and Q.Clear had an SUVmean of 6.5 ± 3 and 7 ± 3, respectively (P < 0.01), and an SUVmax of 11 ± 4.8 and 12 ± 4, respectively (P < 0.01). No significant difference in mean lesion volume was found between algorithms. Conclusion: Among the various Discovery bismuth germanium oxide-based PET/CT scanners, the IQ with 5-ring detector blocks has the highest overall performance, with improved sensitivity and counting rate performance. Q.Clear reconstruction improves the PET image quality, with higher recovery coefficients and lower background variability.


Asunto(s)
Neoplasias/diagnóstico por imagen , Tomografía Computarizada por Tomografía de Emisión de Positrones/instrumentación , Imagen de Cuerpo Entero/instrumentación , Diseño de Equipo , Análisis de Falla de Equipo , Humanos , Fantasmas de Imagen , Reproducibilidad de los Resultados , Sensibilidad y Especificidad , Relación Señal-Ruido