RESUMEN
BACKGROUND: In two interim analyses of this trial, patients with advanced heart failure who were treated with a fully magnetically levitated centrifugal-flow left ventricular assist device were less likely to have pump thrombosis or nondisabling stroke than were patients treated with a mechanical-bearing axial-flow left ventricular assist device. METHODS: We randomly assigned patients with advanced heart failure to receive either the centrifugal-flow pump or the axial-flow pump irrespective of the intended goal of use (bridge to transplantation or destination therapy). The composite primary end point was survival at 2 years free of disabling stroke or reoperation to replace or remove a malfunctioning device. The principal secondary end point was pump replacement at 2 years. RESULTS: This final analysis included 1028 enrolled patients: 516 in the centrifugal-flow pump group and 512 in the axial-flow pump group. In the analysis of the primary end point, 397 patients (76.9%) in the centrifugal-flow pump group, as compared with 332 (64.8%) in the axial-flow pump group, remained alive and free of disabling stroke or reoperation to replace or remove a malfunctioning device at 2 years (relative risk, 0.84; 95% confidence interval [CI], 0.78 to 0.91; P<0.001 for superiority). Pump replacement was less common in the centrifugal-flow pump group than in the axial-flow pump group (12 patients [2.3%] vs. 57 patients [11.3%]; relative risk, 0.21; 95% CI, 0.11 to 0.38; P<0.001). The numbers of events per patient-year for stroke of any severity, major bleeding, and gastrointestinal hemorrhage were lower in the centrifugal-flow pump group than in the axial-flow pump group. CONCLUSIONS: Among patients with advanced heart failure, a fully magnetically levitated centrifugal-flow left ventricular assist device was associated with less frequent need for pump replacement than an axial-flow device and was superior with respect to survival free of disabling stroke or reoperation to replace or remove a malfunctioning device. (Funded by Abbott; MOMENTUM 3 ClinicalTrials.gov number, NCT02224755.).
Asunto(s)
Insuficiencia Cardíaca/terapia , Corazón Auxiliar , Diseño de Prótesis , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Supervivencia sin Enfermedad , Femenino , Corazón Auxiliar/efectos adversos , Humanos , Análisis de Intención de Tratar , Estimación de Kaplan-Meier , Masculino , Persona de Mediana Edad , Falla de Prótesis , Reoperación/estadística & datos numéricos , Accidente Cerebrovascular/etiologíaRESUMEN
BACKGROUND: In an early analysis of this trial, use of a magnetically levitated centrifugal continuous-flow circulatory pump was found to improve clinical outcomes, as compared with a mechanical-bearing axial continuous-flow pump, at 6 months in patients with advanced heart failure. METHODS: In a randomized noninferiority and superiority trial, we compared the centrifugal-flow pump with the axial-flow pump in patients with advanced heart failure, irrespective of the intended goal of support (bridge to transplantation or destination therapy). The composite primary end point was survival at 2 years free of disabling stroke (with disabling stroke indicated by a modified Rankin score of >3; scores range from 0 to 6, with higher scores indicating more severe disability) or survival free of reoperation to replace or remove a malfunctioning device. The noninferiority margin for the risk difference (centrifugal-flow pump group minus axial-flow pump group) was -10 percentage points. RESULTS: Of 366 patients, 190 were assigned to the centrifugal-flow pump group and 176 to the axial-flow pump group. In the intention-to-treat population, the primary end point occurred in 151 patients (79.5%) in the centrifugal-flow pump group, as compared with 106 (60.2%) in the axial-flow pump group (absolute difference, 19.2 percentage points; 95% lower confidence boundary, 9.8 percentage points [P<0.001 for noninferiority]; hazard ratio, 0.46; 95% confidence interval [CI], 0.31 to 0.69 [P<0.001 for superiority]). Reoperation for pump malfunction was less frequent in the centrifugal-flow pump group than in the axial-flow pump group (3 patients [1.6%] vs. 30 patients [17.0%]; hazard ratio, 0.08; 95% CI, 0.03 to 0.27; P<0.001). The rates of death and disabling stroke were similar in the two groups, but the overall rate of stroke was lower in the centrifugal-flow pump group than in the axial-flow pump group (10.1% vs. 19.2%; hazard ratio, 0.47; 95% CI, 0.27 to 0.84, P=0.02). CONCLUSIONS: In patients with advanced heart failure, a fully magnetically levitated centrifugal-flow pump was superior to a mechanical-bearing axial-flow pump with regard to survival free of disabling stroke or reoperation to replace or remove a malfunctioning device. (Funded by Abbott; MOMENTUM 3 ClinicalTrials.gov number, NCT02224755 .).
Asunto(s)
Insuficiencia Cardíaca/terapia , Corazón Auxiliar , Diseño de Prótesis , Adulto , Anciano , Anciano de 80 o más Años , Presión Sanguínea , Femenino , Insuficiencia Cardíaca/mortalidad , Insuficiencia Cardíaca/fisiopatología , Corazón Auxiliar/efectos adversos , Humanos , Análisis de Intención de Tratar , Estimación de Kaplan-Meier , Masculino , Persona de Mediana Edad , Calidad de Vida , Reoperación/estadística & datos numéricos , Accidente Cerebrovascular/epidemiología , Accidente Cerebrovascular/etiología , Accidente Cerebrovascular/mortalidad , Trombosis/etiología , Resultado del Tratamiento , Prueba de PasoRESUMEN
AIMS: Intra-aortic balloon pump (IABP) devices are commonly used in patients with heart failure related cardiogenic shock (HF-CS), including those with out-of-proportion right ventricular (RV) dysfunction. Pulmonary artery pulsatility index (PAPi) is a haemodynamic surrogate for RV performance. We aimed to assess short-term haemodynamic changes in patients with HF-CS following IABP support stratified by baseline PAPi. METHODS AND RESULTS: This is a single-centre study of 67 consecutive patients with HF-CS who underwent IABP placement between 2020 and 2022. The primary aim was haemodynamic changes of specific variables on pulmonary artery catheter monitoring over 72 h following IABP placement. Secondary aims were clinically significant changes in diuretic regimens, changes in inotropes or vasopressors at 72 h following IABP, along with clinical outcomes. Prior to IABP placement, 57% of the total cohort (median age 59 years [48, 69], 31% female) had Society of Cardiovascular Angiography and Interventions Stage C HF-CS. Thirty-eight (56%) patients had a PAPi <2.0. Following 72 h of IABP support, the PAPi <2.0 group had an observed significant decrease in central venous pressure (CVP; 20 to 12 mmHg, P < 0.001) and mean pulmonary artery pressure (mPAP; 37.5 to 28.5 mmHg, P = 0.001), and an increase in PAPi (1 to 1.6, P = 0.001). No significant change in cardiac index (CI; 2 to 2.1 L/min/m2, P = 0.31) was observed. The PAPi ≥2.0 group (N = 29) had no observed significant change in CVP (10 to 8 mmHg, P = 0.47), or PAPi (2.6 to 2.8, P = 0.92), but there was a significant improvement in CI (1.9 to 2.5 L/min/m2, P = 0.004) along with reduction in mPA (37 to 29 mmHg, P = 0.03). The PAPi <2.0 group had a significant increase in diuretic requirement (52.6% vs. 20.7%, P = 0.01) and numerically greater addition of inotropes/vasopressors (47.3% vs. 34.4%, P = 0.07) compared with the PAPi ≥2.0 group at 72 h following IABP placement. Significantly more patients in the PAPi ≥2.0 group underwent left ventricular assist device (55.2% vs. 26.3%, P = 0.02), with no overall significant differences observed in escalation to veno-arterial extracorporeal membrane oxygenation, 30-day mortality, renal replacement therapy post-IABP, or rates of heart transplantation. CONCLUSIONS: IABP devices in those with HF-CS and low or abnormal PAPi may provide modest short-term haemodynamic benefits without significant improvement in CI, along with greater need for adjustment in medical therapeutics to achieve haemodynamic optimization.
RESUMEN
BACKGROUND: Contemporary ventricular assist device therapy results in a high rate of successful heart transplantation but is associated with bleeding, infections, and other complications. Further reductions in pump size, centrifugal design, and intrapericardial positioning may reduce complications and improve outcomes. METHODS AND RESULTS: We studied a small, intrapericardially positioned, continuous-flow centrifugal pump in patients requiring an implanted ventricular assist device as a bridge to heart transplantation. The course of investigational pump recipients was compared with that of patients implanted contemporaneously with commercially available devices. The primary outcome, success, was defined as survival on the originally implanted device, transplantation, or explantation for ventricular recovery at 180 days and was evaluated for both noninferiority and superiority. Secondary outcomes included a comparison of survival between groups and functional and quality-of-life outcomes and adverse events in the investigational device group. A total of 140 patients received the investigational pump, and 499 patients received a commercially available pump implanted contemporaneously. Success occurred in 90.7% of investigational pump patients and 90.1% of controls, establishing the noninferiority of the investigational pump (P<0.001; 15% noninferiority margin). At 6 months, median 6-minute walk distance improved by 128.5 m, and both disease-specific and global quality-of-life scores improved significantly. CONCLUSIONS: A small, intrapericardially positioned, continuous-flow, centrifugal pump was noninferior to contemporaneously implanted, commercially available ventricular assist devices. Functional capacity and quality of life improved markedly, and the adverse event profile was favorable. CLINICAL TRIAL REGISTRATION: URL: http://www.clinicaltrials.gov. Unique identifier: NCT00751972.
Asunto(s)
Trasplante de Corazón/tendencias , Corazón Auxiliar/tendencias , Pericardio , Listas de Espera , Adulto , Anciano , Femenino , Estudios de Seguimiento , Insuficiencia Cardíaca/mortalidad , Insuficiencia Cardíaca/fisiopatología , Insuficiencia Cardíaca/cirugía , Trasplante de Corazón/mortalidad , Humanos , Masculino , Persona de Mediana Edad , Pericardio/fisiopatología , Estudios Prospectivos , Encuestas y Cuestionarios , Tasa de Supervivencia/tendencias , Resultado del Tratamiento , Disfunción Ventricular Izquierda/mortalidad , Disfunción Ventricular Izquierda/fisiopatología , Disfunción Ventricular Izquierda/cirugía , Listas de Espera/mortalidadRESUMEN
BACKGROUND: Endomyocardial biopsy is the standard method of monitoring for rejection in recipients of a cardiac transplant. However, this procedure is uncomfortable, and there are risks associated with it. Gene-expression profiling of peripheral-blood specimens has been shown to correlate with the results of an endomyocardial biopsy. METHODS: We randomly assigned 602 patients who had undergone cardiac transplantation 6 months to 5 years previously to be monitored for rejection with the use of gene-expression profiling or with the use of routine endomyocardial biopsies, in addition to clinical and echocardiographic assessment of graft function. We performed a noninferiority comparison of the two approaches with respect to the composite primary outcome of rejection with hemodynamic compromise, graft dysfunction due to other causes, death, or retransplantation. RESULTS: During a median follow-up period of 19 months, patients who were monitored with gene-expression profiling and those who underwent routine biopsies had similar 2-year cumulative rates of the composite primary outcome (14.5% and 15.3%, respectively; hazard ratio with gene-expression profiling, 1.04; 95% confidence interval, 0.67 to 1.68). The 2-year rates of death from any cause were also similar in the two groups (6.3% and 5.5%, respectively; P=0.82). Patients who were monitored with the use of gene-expression profiling underwent fewer biopsies per person-year of follow-up than did patients who were monitored with the use of endomyocardial biopsies (0.5 vs. 3.0, P<0.001). CONCLUSIONS: Among selected patients who had received a cardiac transplant more than 6 months previously and who were at a low risk for rejection, a strategy of monitoring for rejection that involved gene-expression profiling, as compared with routine biopsies, was not associated with an increased risk of serious adverse outcomes and resulted in the performance of significantly fewer biopsies. (ClinicalTrials.gov number, NCT00351559.)
Asunto(s)
Biopsia , Perfilación de la Expresión Génica , Rechazo de Injerto/diagnóstico , Trasplante de Corazón , Adolescente , Adulto , Anciano , Biopsia/efectos adversos , Biopsia/estadística & datos numéricos , Intervalos de Confianza , Endocardio/patología , Femenino , Estudios de Seguimiento , Rechazo de Injerto/genética , Rechazo de Injerto/mortalidad , Trasplante de Corazón/mortalidad , Humanos , Terapia de Inmunosupresión/efectos adversos , Estimación de Kaplan-Meier , Masculino , Persona de Mediana Edad , Miocardio/patología , Reoperación , Tasa de Supervivencia , Adulto JovenRESUMEN
AIMS: Continuous-flow left ventricular assist devices (LVADs) as destination therapy (DT) are a recommended treatment by National Institute for Health and Care Excellence England for end-stage heart failure patients ineligible for cardiac transplantation. Despite the fact that DT is frequently used as an LVAD indication across other major European countries and the United States, with consistent improvements in quality-of-life and longevity, National Health Service (NHS) England does not currently fund DT, mainly due to concerns over cost-effectiveness. On the basis of the recently published ENDURANCE Supplemental Trial studying DT patients, we assessed for the first time the cost-effectiveness of DT LVADs compared with medical management (MM) in the NHS England. METHODS AND RESULTS: We developed a Markov multiple-state economic model using NHS cost data. LVAD survival and adverse event rates were derived from the ENDURANCE Supplemental Trial. MM survival was based on Seattle Heart Failure Model estimates in the absence of contemporary clinical trials for this population. Incremental cost-effectiveness ratios (ICERs) were calculated over a lifetime horizon. A discount rate of 3.5% per year was applied to costs and benefits. Deterministic ICER was £46 207 per quality-adjusted life year (QALY). Costs and utilities were £204 022 and 3.27 QALYs for the LVAD arm vs. £77 790 and 0.54 QALYs for the MM arm. Sensitivity analyses confirmed robustness of the primary analysis. CONCLUSIONS: The implantation of the HeartWare™ HVAD™ System in patients ineligible for cardiac transplantation as DT is a cost-effective therapy in the NHS England healthcare system under the end-of-life willingness-to-pay threshold of £50 000/QALY, which applies for VAD patients.
Asunto(s)
Insuficiencia Cardíaca , Trasplante de Corazón , Corazón Auxiliar , Análisis Costo-Beneficio , Insuficiencia Cardíaca/cirugía , Humanos , Medicina Estatal , Estados Unidos/epidemiologíaRESUMEN
Invasive pulmonary mucormycosis is a potentially fatal infection that can occur in immunosuppressed patients such as those who have undergone orthotopic heart transplant (OHT). High-dose intravenous antifungal agents, including amphotericin B, are generally accepted as the first-line medical treatment, with prompt surgical resection of lesions if feasible. The body of evidence guiding treatment decisions, however, is sparse, particularly regarding adjustment of immunosuppression during acute infection and long-term recovery. We present 2 cases of patients with pulmonary mucormycosis occurring within the first 6 months after OHT, both of whom successfully recovered after appropriate medical and surgical treatment, and we highlight differences in immunosuppression management strategies for this life-threatening condition.
Asunto(s)
Trasplante de Corazón , Mucormicosis , Anfotericina B , Antifúngicos/uso terapéutico , Trasplante de Corazón/efectos adversos , Humanos , Huésped Inmunocomprometido , Mucormicosis/diagnóstico , Mucormicosis/tratamiento farmacológico , Mucormicosis/etiologíaRESUMEN
There is limited data on the cost-effectiveness of continuous-flow left ventricular assist devices (LVAD) in the United States particularly for the bridge-to-transplant indication. Our objective is to study the cost-effectiveness of a small intrapericardial centrifugal LVAD compared with medical management (MM) and subsequent heart transplantation using the respective clinical trial data. We developed a Markov economic framework. Clinical inputs for the LVAD arm were based on prospective trials employing the HeartWare centrifugal-flow ventricular assist device system. To better assess survival in the MM arm, and in the absence of contemporary trials randomizing patients to LVAD and MM, estimates from the Seattle Heart Failure Model were used. Costs inputs were calculated based on Medicare claim analyses and when appropriate prior published literature. Time horizon was lifetime. Costs and benefits were appropriately discounted at 3% per year. The deterministic cost-effectiveness analyses resulted in $69,768 per Quality Adjusted Life Year and $56,538 per Life Year for the bridge-to-transplant indication and $102,587 per Quality Adjusted Life Year and $87,327 per Life Year for destination therapy. These outcomes signify a substantial improvement compared with prior studies and re-open the discussion around the cost-effectiveness of LVADs.
Asunto(s)
Insuficiencia Cardíaca/terapia , Corazón Auxiliar/economía , Análisis Costo-Beneficio , Femenino , Humanos , Masculino , Cadenas de Markov , Medicare , Estudios Prospectivos , Calidad de Vida , Años de Vida Ajustados por Calidad de Vida , Estados UnidosRESUMEN
BACKGROUND: Although a potentially important pathophysiologic factor in heart failure, the prevalence and predictors of anemia have not been well studied in unselected patients with heart failure. METHODS: The Study of Anemia in a Heart Failure Population (STAMINA-HFP) Registry prospectively studied the prevalence of anemia and the relationship of hemoglobin to health-related quality of life and outcomes among patients with heart failure. A random selection algorithm was used to reduce bias during enrollment of patients seen in specialty clinics or clinics of community cardiologists with experience in heart failure. In this initial report, data on prevalence and correlates of anemia were analyzed in 1,076 of the 1,082 registry patients who had clinical characteristics and hemoglobin determined by finger-stick at baseline. RESULTS: Overall (n = 1,082), the registry patients were 41% female and 73% white with a mean age (+/-SD) of 64 +/- 14 years (68 +/- 13 years in community and 57 +/- 14 years in specialty sites, P < .001). Among the 1,076 patients in the prevalence analysis, mean hemoglobin was 13.3 +/- 2.1 g/dL (median 13.2 g/dL); and anemia (defined by World Health Organization criteria) was present in 34%. Age identified patients at risk for anemia, with 40% of patients >70 years affected. CONCLUSIONS: Initial results from the STAMINA-HFP Registry suggest that anemia is a common comorbidity in unselected outpatients with heart failure. Given the strong association of anemia with adverse outcomes in heart failure, this study supports further investigation concerning the importance of anemia as a therapeutic target in this condition.
Asunto(s)
Anemia/epidemiología , Insuficiencia Cardíaca/complicaciones , Hemoglobinas/metabolismo , Sistema de Registros , Medición de Riesgo/métodos , Anciano , Anemia/sangre , Anemia/etiología , Femenino , Estudios de Seguimiento , Insuficiencia Cardíaca/sangre , Humanos , Incidencia , Masculino , Persona de Mediana Edad , Prevalencia , Pronóstico , Estudios Prospectivos , Calidad de Vida , Factores de RiesgoRESUMEN
BACKGROUND: In some patients with heart failure, beta-blockers can improve left ventricular (LV) function and reduce morbidity and mortality. We hypothesized that gadolinium-enhanced cardiovascular magnetic resonance imaging (CMR) can predict reversible myocardial dysfunction and remodeling in heart failure patients treated with beta-blockers. METHODS AND RESULTS: Forty-five patients with chronic heart failure underwent CMR. Contrast imaging using gadolinium was performed to obtain high-resolution spatial maps of myocardial scarring and viability. Cine imaging was performed to assess LV function and morphology and was repeated in 35 patients after 6 months of beta-blockade. Gadolinium CMR demonstrated scarring in 30 of 45 patients (67%). Scarring was found in 100% of patients with ischemic cardiomyopathy (28 of 28) but in only 12% with nonischemic cardiomyopathy (2 of 17). In the 35 patients who were maintained on beta-blockers and had a second study, there was an inverse relation between the extent of scarring at baseline and the likelihood of contractile improvement 6 months later (P<0.001). For instance, contractility improved in 56% (674 of 1207) of regions with no scarring but in only 3% with >75% scarring (8 of 232). Multivariate analysis showed that the amount of dysfunctional but viable myocardium by CMR was an independent predictor of the change in ejection fraction (P=0.01), mean wall motion score (P=0.0007), LV end-diastolic volume index (P=0.007), and LV end-systolic volume index (P< or =0.0001). CONCLUSIONS: For heart failure patients treated with beta-blockers, gadolinium-enhanced CMR predicts the response in LV function and remodeling.
Asunto(s)
Antagonistas Adrenérgicos beta/uso terapéutico , Gadolinio , Insuficiencia Cardíaca/diagnóstico , Imagen por Resonancia Magnética , Disfunción Ventricular Izquierda/diagnóstico , Remodelación Ventricular , Anciano , Cicatriz/diagnóstico , Insuficiencia Cardíaca/tratamiento farmacológico , Insuficiencia Cardíaca/fisiopatología , Compuestos Heterocíclicos , Humanos , Imagen por Resonancia Cinemagnética , Masculino , Persona de Mediana Edad , Análisis Multivariante , Contracción Miocárdica/efectos de los fármacos , Compuestos Organometálicos , Valor Predictivo de las Pruebas , Estudios Prospectivos , Recuperación de la Función/efectos de los fármacos , Volumen Sistólico/efectos de los fármacos , Resultado del Tratamiento , Disfunción Ventricular Izquierda/tratamiento farmacológico , Disfunción Ventricular Izquierda/fisiopatología , Remodelación Ventricular/efectos de los fármacosRESUMEN
BACKGROUND: The Ventricular Assist Device for the Treatment of Advanced Heart Failure (ADVANCE) Bridge to Transplant (BTT) trial was a multicenter, prospective trial of the HeartWare Ventricular Assist Device (HVAD). The performance of the HVAD in various demographic sub-groups was evaluated. METHODS: Baseline characteristics, adverse events, and survival were compared for men vs. women and whites vs. non-whites in the combined ADVANCE BTT and continued access protocol trial. Of 332 patients enrolled in these trials, 236 were men and 96 women, with 228 whites and 104 non-whites. RESULTS: At baseline, women had a smaller body surface area (1.8 ± 0.2 vs. 2.1 ± 0.3 m2, p < 0.0001), less hypertension (50.0% vs. 61.9%, p = 0.05), and less ischemic cardiomyopathy (15.6% vs. 45.3%, p < 0.0001). Differences in Kaplan-Meier survival were not significant at 180 days (men, 91.8%; women, 91.7%) and 1 year (men, 85.3%; women, 85.1%) despite adjustment for baseline differences. Men had a lower incidence of early right heart failure and renal and respiratory dysfunction, and a shorter length of stay. In the analysis by race, non-whites were younger than whites and had less ischemic heart failure, more hypertension, and lower creatinine levels at baseline. Non-whites had lower rates of arrhythmia, bleeding requiring rehospitalization, and device malfunctions than whites. Survival was high in non-whites and whites, at 94.1% vs. 90.4% at 180 days and 89.2% vs. 82.8% at 1 year, respectively, despite adjustment for baseline differences. CONCLUSIONS: Although heart failure etiology differed between men and women and between whites and non-whites, sex and race were not factors that affected survival in patients receiving the HVAD as BTT, which was high in all sub-groups.
Asunto(s)
Insuficiencia Cardíaca/mortalidad , Insuficiencia Cardíaca/terapia , Trasplante de Corazón , Corazón Auxiliar , Grupos Raciales , Factores Sexuales , Adulto , Creatinina/sangre , Femenino , Estudios de Seguimiento , Insuficiencia Cardíaca/etnología , Humanos , Hipertensión/epidemiología , Incidencia , Estimación de Kaplan-Meier , Tiempo de Internación , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Estudios Retrospectivos , Tasa de Supervivencia , Resultado del TratamientoRESUMEN
BACKGROUND: The basis for increased mortality after heart transplantation in African Americans and other non-Caucasian racial groups is poorly defined. We hypothesized that increased risk of adverse events is driven by biologic factors. To test this hypothesis in the Invasive Monitoring Attenuation through Gene Expression (IMAGE) study, we determined whether the event rate of the primary outcome of acute rejection, graft dysfunction, death, or retransplantation varied by race as a function of calcineurin inhibitor (CNI) levels and gene expression profile (GEP) scores. METHODS: We determined the event rate of the primary outcome, comparing racial groups, stratified by time after transplant. Logistic regression was used to compute the relative risk across racial groups, and linear modeling was used to measure the dependence of CNI levels and GEP score on race. RESULTS: In 580 patients monitored for a median of 19 months, the incidence of the primary end point was 18.3% in African Americans, 22.2% in other non-Caucasians, and 8.5% in Caucasians (p < 0.001). There were small but significant correlations of race and tacrolimus trough levels to the GEP score. Tacrolimus levels were similar among the races. Of patients receiving tacrolimus, other non-Caucasians had higher GEP scores than the other racial groups. African American recipients demonstrated a unique decrease in expression of the FLT3 gene in response to higher tacrolimus levels. CONCLUSIONS: African Americans and other non-Caucasian heart transplant recipients were 2.5-times to 3-times more likely than Caucasians to experience outcome events in the Invasive Monitoring Attenuation through Gene Expression study. The increased risk of adverse outcomes may be partly due to the biology of the alloimmune response, which is less effectively inhibited at similar tacrolimus levels in minority racial groups.
Asunto(s)
Perfilación de la Expresión Génica/métodos , Rechazo de Injerto/genética , Trasplante de Corazón , Grupos Raciales , Adolescente , Adulto , Anciano , Femenino , Rechazo de Injerto/etnología , Insuficiencia Cardíaca/cirugía , Humanos , Incidencia , Masculino , Persona de Mediana Edad , Estudios Retrospectivos , Tasa de Supervivencia/tendencias , Estados Unidos/epidemiología , Adulto JovenRESUMEN
Serum B-type natriuretic peptide (BNP) levels reflect myocardial strain and are known to be elevated in patients with heart failure. To determine if BNP levels are elevated in patients with aortic regurgitation, we measured BNP levels in patients with chronic asymptomatic aortic regurgitation and normal left ventricular systolic function.
Asunto(s)
Insuficiencia de la Válvula Aórtica/sangre , Péptido Natriurético Encefálico/sangre , Adulto , Anciano , Insuficiencia de la Válvula Aórtica/fisiopatología , Enfermedad Crónica , Humanos , Masculino , Persona de Mediana Edad , Función Ventricular Izquierda/fisiologíaRESUMEN
Beta blockers improve left ventricular (LV) ejection fraction but patient responses are heterogenous. We investigated the role of contractile reserve (CR) in predicting beta-blocker response in ischemic and nonischemic cardiomyopathy. Resting and low-dose dobutamine echocardiograms were recorded in 32 patients with heart failure (LV ejection fraction Asunto(s)
Antagonistas Adrenérgicos beta/uso terapéutico
, Cardiomiopatía Dilatada/fisiopatología
, Contracción Miocárdica/fisiología
, Isquemia Miocárdica/fisiopatología
, Disfunción Ventricular Izquierda/tratamiento farmacológico
, Disfunción Ventricular Izquierda/fisiopatología
, Adulto
, Anciano
, Anciano de 80 o más Años
, Inhibidores de la Enzima Convertidora de Angiotensina/uso terapéutico
, Cardiomiopatía Dilatada/complicaciones
, Cardiomiopatía Dilatada/tratamiento farmacológico
, Femenino
, Estudios de Seguimiento
, Humanos
, Masculino
, Persona de Mediana Edad
, Isquemia Miocárdica/complicaciones
, Isquemia Miocárdica/tratamiento farmacológico
, Valor Predictivo de las Pruebas
, Estudios Prospectivos
, Factores de Tiempo
, Disfunción Ventricular Izquierda/complicaciones
RESUMEN
BACKGROUND: Few studies have reported on hospital length of stay (LOS) after left ventricular assist device (LVAD) implantation. The purpose of this study was to determine pre- and peri-operative predictors of hospital LOS after LVAD implantation. METHODS: We analyzed adult primary continuous-flow LVAD patients implanted between June 23, 2006 and December 31, 2010 at 105 institutions from the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS). Retrospective analyses included measures of central tendency, frequencies, correlations and stepwise multivariable regression modeling (p ≤ 0.05). Independent variables included demographic characteristics, pre-implant clinical and behavioral variables and concomitant surgery. RESULTS: Characteristics of the patients (n = 2,200) included: mean age 54.6 ± 12.6 years; 79% male; 69% white; 57% INTERMACS Profile 1 or 2; 37% diabetic; 21% with history of coronary artery bypass graft (CABG); 7% with history of valve surgery; and 37% with concomitant surgery. Median hospital LOS (implant to discharge) was 20 days. Significant predictors of an increased hospital LOS included demographic characteristics (older age and non-white), pre-implant clinical variables (history of CABG or valve surgery, diabetes, ascites, INTERMACS Profiles 1 and 2, low albumin, high blood urea nitrogen, high right atrial pressure) and concomitant surgery, explaining 12% variance (F = 22.65, p<0.001). CONCLUSIONS: Demographic characteristics, pre-implant variables and concomitant surgery partially explained hospital LOS after continuous-flow LVAD implant. These variables have implications regarding selection of patients for mechanical circulatory support.
Asunto(s)
Insuficiencia Cardíaca/terapia , Corazón Auxiliar , Tiempo de Internación/estadística & datos numéricos , Periodo Perioperatorio , Periodo Preoperatorio , Sistema de Registros , Adulto , Anciano , Estudios de Cohortes , Femenino , Humanos , Masculino , Persona de Mediana Edad , Análisis Multivariante , Selección de Paciente , Estudios Retrospectivos , Factores de Tiempo , Resultado del TratamientoRESUMEN
BACKGROUND: Gene expression profiling test scores have primarily been used to identify heart transplant recipients who have a low probability of rejection at the time of surveillance testing. We hypothesized that the variability of gene expression profiling test scores within a patient may predict risk of future events of allograft dysfunction or death. METHOD: Patients from the IMAGE study with rejection surveillance gene expression profiling tests performed at 1- to 6-month intervals were selected for this cohort study. Gene expression profiling score variability was defined as the standard deviation of an individual's cumulative test scores. Gene expression profiling ordinal score (range, 0-39), threshold score (binary value=1 if ordinal score ≥ 34), and score variability were studied in multivariate Cox regression models to predict future clinical events. RESULTS: Race, age at time of transplantation, and time posttransplantation were significantly associated with future events in the univariate analysis. In the multivariate analyses, gene expression profiling score variability, but not ordinal scores or scores over threshold, was independently associated with future clinical events. The regression coefficient P values were <0.001, 0.46, and 0.773, for gene expression profiling variability, ordinal, and threshold scores, respectively. The hazard ratio for a 1 unit increase in variability was 1.76 (95% CI, 1.4-2.3). DISCUSSION: The variability of a heart recipient's gene expression profiling test scores over time may provide prognostic utility. This information is independent of the probability of acute cellular rejection at the time of testing that is rendered from a single ordinal gene-expression profiling test score.
Asunto(s)
Perfilación de la Expresión Génica , Pruebas Genéticas , Rechazo de Injerto/genética , Trasplante de Corazón/efectos adversos , Adulto , Anciano , Biopsia , Femenino , Perfilación de la Expresión Génica/métodos , Predisposición Genética a la Enfermedad , Pruebas Genéticas/métodos , Rechazo de Injerto/patología , Trasplante de Corazón/mortalidad , Humanos , Masculino , Persona de Mediana Edad , Análisis Multivariante , Valor Predictivo de las Pruebas , Modelos de Riesgos Proporcionales , Factores de Riesgo , Factores de Tiempo , Resultado del Tratamiento , Estados UnidosRESUMEN
Despite the widespread use of loop diuretics to treat acute decompensated heart failure (ADHF), robust data supporting their role and optimal dosing strategies are scarce. This analysis aimed to compare clinical outcomes of patients admitted with ADHF who received a diuretic dosing protocol with those who received the usual diuretic therapy. We performed an observational medical records review to compare the use of a nurse-driven diuretic dosing protocol with usual diuretic dosing for patients admitted with ADHF during a 1-year period. Using a propensity scoring model, comparisons were made between groups for total weight loss, length of stay (LOS), 30-day readmissions, in-hospital mortality, 30-day mortality, and acute kidney failure. Sixty-eight of the 596 patients admitted with ADHF during the study period received the diuretic protocol. Protocol use was associated with an additional 2.63-kg weight loss (P=.003) but a trend toward increased LOS compared with patients receiving usual care (P=.097). However, patients receiving the protocol had a significantly lower risk of 30-day readmission (odds ratio, 0.46, 95% confidence interval, 0.22-0.95). Protocol use was not associated with significant differences in kidney failure, inpatient mortality, or 30-day mortality. A diuretic dosing protocol for patients admitted with ADHF improves weight loss and may lower 30-day readmissions, at the cost of potentially increasing LOS.
Asunto(s)
Volumen Sanguíneo/efectos de los fármacos , Bumetanida/administración & dosificación , Furosemida/administración & dosificación , Insuficiencia Cardíaca , Enfermedad Aguda , Administración Intravenosa , Anciano , Distribución de Chi-Cuadrado , Protocolos Clínicos , Diuréticos/administración & dosificación , Relación Dosis-Respuesta a Droga , Monitoreo de Drogas , Femenino , Insuficiencia Cardíaca/tratamiento farmacológico , Insuficiencia Cardíaca/mortalidad , Insuficiencia Cardíaca/fisiopatología , Humanos , Tiempo de Internación/estadística & datos numéricos , Modelos Lineales , Masculino , Persona de Mediana Edad , Evaluación del Resultado de la Atención al Paciente , Readmisión del Paciente/estadística & datos numéricos , Estudios Retrospectivos , Análisis de Supervivencia , Resultado del Tratamiento , Estados Unidos/epidemiologíaRESUMEN
BACKGROUND: Heart transplantation requires substantial personal, financial, and psychosocial resources. Using an existing multisite data set, we examined predictors of mortality at 5 to 10 years after heart transplantation. METHODS: All 555 participants completed a self-report quality of life instrument. Of these patients, 55 (10%) died 5 to 10 years after heart transplantation. Statistical analyses included frequencies, means, Pearson correlation coefficients, and Cox proportional hazard modeling. RESULTS: Educational level and higher levels of social and economic satisfaction were predictive of improved survival. Conversely, married status, more cumulative infections, the presence of hematologic disorders, higher New York Heart Association (NYHA) class, and poor adherence to medical care predicted worse survival. CONCLUSIONS: Demographic, clinical, psychosocial, and behavioral factors were important predictors of long-term survival after heart transplantation. These findings have important implications for patient selection for heart transplantation, as well as for posttransplantation care.
Asunto(s)
Adaptación Psicológica , Trasplante de Corazón/mortalidad , Trasplante de Corazón/psicología , Satisfacción del Paciente , Calidad de Vida , Adulto , Factores de Edad , Femenino , Estudios de Seguimiento , Humanos , Masculino , Persona de Mediana Edad , Pronóstico , Estudios Prospectivos , Distribución por Sexo , Encuestas y Cuestionarios , Tasa de Supervivencia/tendencias , Factores de Tiempo , Estados Unidos/epidemiologíaRESUMEN
BACKGROUND: The HeartWare Ventricular Assist System (HeartWare Inc, Framingmam, MA) is a miniaturized implantable, centrifugal design, continuous-flow blood pump. The pivotal bridge to transplant and continued access protocols trials have enrolled patients with advanced heart failure in a bridge-to-transplant indication. METHODS: The primary outcome, success, was defined as survival on the originally implanted device, transplant, or explant for ventricular recovery at 180 days. Secondary outcomes included an evaluation of survival, functional and quality of life outcomes, and adverse events. RESULTS: A total of 332 patients in the pivotal bridge to transplant and continued access protocols trial have completed their 180-day primary end-point assessment. Survival in patients receiving the HeartWare pump was 91% at 180 days and 84% at 360 days. Quality of life scores improved significantly, and adverse event rates remain low. CONCLUSIONS: The use of the HeartWare pump as a bridge to transplant continues to demonstrate a high 180-day survival rate despite a low rate of transplant. Adverse event rates are similar or better than those observed in historical bridge-to-transplant trials, despite longer exposure times due to longer survival and lower transplant rates.
Asunto(s)
Insuficiencia Cardíaca/cirugía , Corazón Auxiliar , Protocolos Clínicos , Diseño de Equipo , Femenino , Trasplante de Corazón , Humanos , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Resultado del Tratamiento , Listas de EsperaRESUMEN
BACKGROUND: Tricuspid regurgitation is common in patients with advanced heart failure. The ideal operative strategy for managing tricuspid valve regurgitation (TR) in patients undergoing left ventricular assist device (LVAD) implantation is unclear. This study was designed to evaluate the effect on outcomes of concomitant tricuspid valve repair (TVR) for moderate to severe (3(+)/4(+)) TR at the time of LVAD implantation. METHODS: Patients with >3(+) TR who underwent LVAD implantation from 2005 to 2009 were retrospectively evaluated. Pre-, intra- and post-operative data, including hemodynamics, inotrope requirements and end-organ function parameters, were considered. Outcomes of patients receiving TVR were compared with those who did not receive TVR (NTVR). RESULTS: Seventy-two LVADs were implanted during the study period. Forty-two (58%) patients had ≥ 3(+) TR prior to LVAD implantation. Eight patients underwent TVR and 34 patients did not undergo TVR (NTVR). There were no significant differences in baseline characteristics or severity of TR between the two groups. The TVR group had a longer cardiopulmonary bypass time (p < 0.01) and required more blood products (p < 0.05). Higher post-operative creatinine and blood urea nitrogen (BUN) values were noted in the TVR group. One patient in the TVR group and 3 patients in the NTVR group required right-sided mechanical assistance (p = 0.6). There was no significant difference in short- or long-term mortality between the two groups. CONCLUSIONS: TVR for ≥ 3(+) TR prolonged operative time and showed similar outcomes compared with LVAD implantation alone. A benefit of performing TVR was not demonstrated. As such, TVR may not be necessary at the time of LVAD implantation.