J Immunol Methods ; 425: 88-96, 2015 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-26130368

RESUMEN

Biotherapeutics can elicit immune responses, which can alter the exposure, safety, and efficacy of the therapeutics. A well-designed and robust bioanalytical method is critical for the detection and characterization of relevant anti-drug antibody (ADA) and the success of an immunogenicity study. As a fundamental criterion in immunogenicity testing, assay cut points need to be statistically established with a risk-based approach to reduce subjectivity. This manuscript describes the development of a validated, web-based, multi-tier customized assay statistical tool (CAST) for assessing cut points of ADA assays. The tool provides an intuitive web interface that allows users to import experimental data generated from a standardized experimental design, select the assay factors, run the standardized analysis algorithms, and generate tables, figures, and listings (TFL). It allows bioanalytical scientists to perform complex statistical analysis at a click of the button to produce reliable assay parameters in support of immunogenicity studies.


Asunto(s)
Anticuerpos/inmunología , Formación de Anticuerpos/inmunología , Automatización/métodos , Automatización/normas , Inmunoensayo/métodos , Inmunoensayo/normas , Algoritmos , Modelos Estadísticos , Estándares de Referencia , Programas Informáticos