1.
Lab Chip ; 10(12): 1618-22, 2010 Jun 21.
Artículo en Inglés | MEDLINE | ID: mdl-20383403

RESUMEN

Although advanced fluid handling using elastomeric valves is useful for a variety of lab-on-a-chip procedures, their operation has traditionally relied on external laboratory infrastructure (such as gas tanks, computers, and ground electricity). This dependence has held back the use of elastomeric microvalves for point-of-care settings. Here, we demonstrate that microfabricated microvalves, via liquid-filled control channels, can be actuated using only a handheld instrument powered by a 9 V battery. This setup can achieve on-off fluid control with fast response times, coordinated switching of multiple valves, and operation of a biological assay. In the future, this technique may enable the widely used elastomeric microvalves (made by multilayer soft lithography) to be increasingly adopted for portable sensors and lab-on-a-chip systems.


Asunto(s)
Elastómeros , Suministros de Energía Eléctrica , Técnicas Analíticas Microfluídicas/instrumentación , Sistemas de Atención de Punto , Computadores , Diseño de Equipo , Membranas Artificiales