Radiat Prot Dosimetry ; 165(1-4): 299-303, 2015 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-25877533

RESUMEN

The assessment of eye lens doses for workers during interventional radiology (IR) procedures was performed using a new eye lens dosemeter. In parallel, the results of routine individual monitoring were analysed and compared with the results obtained from measurements with a new eye lens dosemeter. The eye lens doses were assessed using Hp(3) measured at the level of the eyes and were compared with Hp(10) measured with the whole-body dosemeter above the lead collar. The information about use of protective measures, the number of performed interventional procedures per month and their fluoroscopy time was also collected. The assessment of doses to the lens of the eye was done for 50 IR workers at 9 Lithuanian hospitals for the period of 2012-2013. If the use of lead glasses is not taken into account, the estimated maximum annual dose equivalent to the lens of the eye was 82 mSv.


Asunto(s)
Cristalino/efectos de la radiación , Exposición Profesional/prevención & control , Exposición a la Radiación/prevención & control , Radiología Intervencionista/métodos , Calibración , Dispositivos de Protección de los Ojos , Fluoroscopía/métodos , Humanos , Plomo , Lituania , Cuerpo Médico , Enfermeras y Enfermeros , Traumatismos Ocupacionales/prevención & control , Médicos , Ropa de Protección , Equipos de Seguridad , Radiometría