RESUMEN
We present the performance analysis and specifications of a portable auto-phoropter system that can be employed for fast refractive assessment of a large population. A customized Shack-Hartmann wavefront sensor is developed to accurately measure the defocus and astigmatism of the eye within ±10D and ±6D, respectively. Three fluidic lenses are designed to correct the vision in real time. A digital Snellen chart is integrated into the system to validate the accuracy of the measurement and the correction by means of achieving 20/20 vision. The refractive error of eight subjects (16 eyes) has been measured objectively (without patient's feedback) using the proposed system and the results are compared with their clinical prescription through the Bland-Altman method. It is shown that the auto-phoropter takes less than 8 s to measure and correct the eye refractive error with an accuracy of ±0.25D.