Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 489
Filtrar
Más filtros

País/Región como asunto
Tipo del documento
Intervalo de año de publicación
1.
J Transl Med ; 22(1): 774, 2024 Aug 16.
Artículo en Inglés | MEDLINE | ID: mdl-39152426

RESUMEN

BACKGROUND: Colorectal cancer is the third most common tumour entity in the world and up to 50% of the patients develop liver metastases (CRLM) within five years. To improve and personalize therapeutic strategies, new diagnostic tools are urgently needed. For instance, biomechanical tumour properties measured by magnetic resonance elastography (MRE) could be implemented as such a diagnostic tool. We postulate that ex vivo MRE combined with histological and radiological evaluation of CRLM could provide biomechanics-based diagnostic markers for cell viability in tumours. METHODS: 34 CRLM specimens from patients who had undergone hepatic resection were studied using ex vivo MRE in a frequency range from 500 Hz to 5300 Hz with increments of 400 Hz. Single frequency evaluation of shear wave speed and wave penetration rate as proxies for stiffness and viscosity was performed, along with rheological model fitting based on the spring-pot model and powerlaw exponent α, ranging between 0 (complete solid behaviour) and 1 (complete fluid behaviour). For histological analysis, samples were stained with H&E and categorized according to the degree of regression. Quantitative histologic analysis was performed to analyse nucleus size, aspect ratio, and density. Radiological response was assessed according to RECIST-criteria. RESULTS: Five samples showed major response to chemotherapy, six samples partial response and 23 samples no response. For higher frequencies (> 2100 Hz), shear wave speed correlated significantly with the degree of regression (p ≤ 0.05) indicating stiffer properties with less viable tumour cells. Correspondingly, rheological analysis of α revealed more elastic-solid tissue properties at low cell viability and major response (α = 0.43 IQR 0.36, 0.47) than at higher cell viability and no response (α = 0.51 IQR 0.48, 0.55; p = 0.03). Quantitative histological analysis showed a decreased nuclear area and density as well as a higher nuclear aspect ratio in patients with major response to treatment compared to patients with no response (all p < 0.05). DISCUSSION: Our results suggest that MRE could be useful in the characterization of biomechanical property changes associated with cell viability in CRLM. In the future, MRE could be applied in clinical diagnosis to support individually tailored therapy plans for patients with CRLM.


Asunto(s)
Supervivencia Celular , Neoplasias Colorrectales , Diagnóstico por Imagen de Elasticidad , Elasticidad , Neoplasias Hepáticas , Humanos , Neoplasias Colorrectales/patología , Neoplasias Hepáticas/secundario , Neoplasias Hepáticas/patología , Neoplasias Hepáticas/diagnóstico por imagen , Masculino , Viscosidad , Femenino , Anciano , Persona de Mediana Edad , Anciano de 80 o más Años
2.
Magn Reson Med ; 2024 Oct 20.
Artículo en Inglés | MEDLINE | ID: mdl-39428691

RESUMEN

PURPOSES: To enhance the functional capability of MRI, this study aims to develop a novel MR elastography (MRE) sequence that achieves rapid acquisition without distortion artifacts. METHODS: A displacement-encoded stimulated echo (DENSE) with multiphase acquisition scheme was used to capture wave images. A center-out golden-angle stack-of-stars sampling pattern was introduced for improved SNR and data incoherence. A combination of Hadamard encoding and interleaved multislab acquisition schemes was used to increase the acquisition efficiency of MRE data with multiple directions and phase offsets. A generalized parallel-imaging and compressed-sensing method was further applied to accelerate the acquisition process. The imaging results of the proposed sequence were compared with those from six gradient echo (GRE)/EPI/DENSE-based MRE sequences via phantom and brain acquisitions. RESULTS: The proposed sequence achieved a 6-fold acceleration compared with GRE MRE. With the application of a conventional parallel-imaging and compressed-sensing algorithm, the scanning speed was further accelerated by 8-fold, matching the speed of EPI-based MRE. Phantom tests revealed small variances in stiffness measurements across the seven sequences (< 9.23%). The proposed sequence exhibited a higher contrast-to-noise ratio (1.38) than the two EPI-based sequences (0.61/0.76) and similar to GRE-based sequences (1.34/1.22/1.58). Brain imaging validated the effectiveness of the proposed sequence in accurate stiffness estimation and distortion artifact avoidance. CONCLUSION: A rapid DENSE-based MRE sequence with interleaved multislab acquisition and Hadamard encoding was developed at a speed matching EPI-based sequences, without compromising SNR or introducing distortion artifacts.

3.
Magn Reson Med ; 91(5): 1923-1935, 2024 May.
Artículo en Inglés | MEDLINE | ID: mdl-38098427

RESUMEN

PURPOSE: To demonstrate a novel MR elastography (MRE) technique, termed here wavelet MRE. With this technique, broadband motion sensitivity is achievable. Moreover, the true tissue displacement can be reconstructed with a simple inverse transform. METHODS: A wavelet MRE sequence was developed with motion-encoding gradients based on Haar wavelets. From the phase images' displacement was estimated using an inverse transform. Simulations were performed using a frequency sweep and a transient as ground-truth motions. A PVC phantom was scanned using wavelet MRE and standard MRE with both transient (one and 10 cycles of 90-Hz motion) and steady-state dual-frequency motion (30 and 60 Hz) for comparison. The technique was tested in a human brain, and motion trajectories were estimated for each voxel. RESULTS: In simulation, the displacement information estimated from wavelet MRE closely matched the true motion. In the phantom test, the MRE phase data generated from the displacement information derived from wavelet MRE agreed well with standard MRE data. Testing of wavelet MRE to assess transient motion waveforms in the brain was successful, and the tissue motion observed was consistent with a previous study. CONCLUSION: The uniform and broadband frequency response of wavelet MRE makes it a promising method for imaging transient, multifrequency motion, or motion with unknown frequency content. One potential application is measuring the response of brain tissue undergoing low-amplitude, transient vibrations as a model for the study of traumatic brain injury.


Asunto(s)
Diagnóstico por Imagen de Elasticidad , Imagen por Resonancia Magnética , Humanos , Imagen por Resonancia Magnética/métodos , Diagnóstico por Imagen de Elasticidad/métodos , Algoritmos , Encéfalo/diagnóstico por imagen , Fantasmas de Imagen , Sonido
4.
NMR Biomed ; 37(2): e5047, 2024 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-37813110

RESUMEN

Magnetic resonance elastography (MRE) is an accurate noninvasive diagnostic tool for assessing the stiffness of parenchymal organs, including the spleen. However, this measurement may be biased due to postprandial changes in splenic stiffness. The aim of the current study was to evaluate postprandial changes in spleen stiffness assessed by MRE in a large sample of healthy volunteers. This was a prospective institutional research ethics board-approved study. Healthy volunteers with no history of liver disease were recruited for an MRE test and blood draw from December 2018 to July 2019. Each participant underwent spleen MRE after at least 4 h of fasting and again 30 min after a 1000 kcal meal. Also, 14 randomly selected volunteers underwent additional MRE examinations at 1.5 and 2.5 h after food intake. The MRE data were acquired at 60 Hz using a 1.5-T MRI scanner. The spleen stiffness was assessed using a weighted mean of stiffness values from regions of interest manually drawn on three to five spleen slices. Spearman's rank correlation, Wilcoxon signed-rank, Friedman, and Mann-Whitney tests were used for statistical analysis. A total of 100 volunteers met the inclusion criteria and were eventually enrolled in this study (age 23 ± 2 years; 65 women). The mean spleen stiffness for the whole group increased by 7.9% (p < 0.001) from the mean ± SD value of 5.09 ± 0.63 (95% CI: 4.96-5.21) kPa in the fasting state to 5.47 ± 0.66 (95% CI 5.34-5.60) kPa 30 min after the meal and then gradually decreased. However, even 2 h 30 min after the meal, the spleen stiffness was higher than in the fasting state. This difference was statistically significant at p less than 0.001. It was concluded that meal intake results in a statistically significant elevation of spleen stiffness that persists for 2.5 h. This finding supports the recommendation for routine fasting for more than 2.5 h prior to assessing MRE-based spleen stiffness.


Asunto(s)
Diagnóstico por Imagen de Elasticidad , Bazo , Humanos , Femenino , Adulto Joven , Adulto , Bazo/diagnóstico por imagen , Diagnóstico por Imagen de Elasticidad/métodos , Estudios Prospectivos , Reproducibilidad de los Resultados , Imagen por Resonancia Magnética/métodos
5.
Acta Neuropathol ; 147(1): 8, 2024 01 04.
Artículo en Inglés | MEDLINE | ID: mdl-38175305

RESUMEN

Multiple sclerosis (MS) is a chronic neuroinflammatory disease that involves both white and gray matter. Although gray matter damage is a major contributor to disability in MS patients, conventional clinical magnetic resonance imaging (MRI) fails to accurately detect gray matter pathology and establish a clear correlation with clinical symptoms. Using magnetic resonance elastography (MRE), we previously reported global brain softening in MS and experimental autoimmune encephalomyelitis (EAE). However, it needs to be established if changes of the spatiotemporal patterns of brain tissue mechanics constitute a marker of neuroinflammation. Here, we use advanced multifrequency MRE with tomoelastography postprocessing to investigate longitudinal and regional inflammation-induced tissue changes in EAE and in a small group of MS patients. Surprisingly, we found reversible softening in synchrony with the EAE disease course predominantly in the cortex of the mouse brain. This cortical softening was associated neither with a shift of tissue water compartments as quantified by T2-mapping and diffusion-weighted MRI, nor with leukocyte infiltration as seen by histopathology. Instead, cortical softening correlated with transient structural remodeling of perineuronal nets (PNNs), which involved abnormal chondroitin sulfate expression and microgliosis. These mechanisms also appear to be critical in humans with MS, where tomoelastography for the first time demonstrated marked cortical softening. Taken together, our study shows that neuroinflammation (i) critically affects the integrity of PNNs in cortical brain tissue, in a reversible process that correlates with disease disability in EAE, (ii) reduces the mechanical integrity of brain tissue rather than leading to water accumulation, and (iii) shows similar spatial patterns in humans and mice. These results raise the prospect of leveraging MRE and quantitative MRI for MS staging and monitoring treatment in affected patients.


Asunto(s)
Diagnóstico por Imagen de Elasticidad , Encefalomielitis Autoinmune Experimental , Esclerosis Múltiple , Humanos , Animales , Ratones , Enfermedades Neuroinflamatorias , Imagen por Resonancia Magnética , Imagen de Difusión por Resonancia Magnética , Encefalomielitis Autoinmune Experimental/diagnóstico por imagen , Agua
6.
J Magn Reson Imaging ; 59(2): 688-698, 2024 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-37194646

RESUMEN

BACKGROUND: Pancreatic ductal adenocarcinoma (PDAC) stromal disposition is thought to influence chemotherapy efficacy and increase tissue stiffness, which could be quantified noninvasively via MR elastography (MRE). Current methods cause position-based errors in pancreas location over time, hampering accuracy. It would be beneficial to have a single breath-hold acquisition. PURPOSE: To develop and test a single breath-hold three-dimensional MRE technique utilizing prospective undersampling and a compressed sensing reconstruction (CS-MRE). STUDY TYPE: Prospective. POPULATION: A total of 30 healthy volunteers (HV) (31 ± 9 years; 33% male) and five patients with PDAC (69 ± 5 years; 80% male). FIELD STRENGTH/SEQUENCE: 3-T, GRE Ristretto MRE. ASSESSMENT: First, optimization of multi breath-hold MRE was done in 10 HV using four combinations of vibration frequency, number of measured wave-phase offsets, and TE and looking at MRE quality measures in the pancreas head. Second, viscoelastic parameters delineated in the pancreas head or tumor of CS-MRE were compared against (I) 2D and (II) 3D four breath-hold acquisitions in HV (N = 20) and PDAC patients. Intrasession repeatability was assessed for CS-MRE in a subgroup of healthy volunteers (N = 15). STATISTICAL TESTS: Tests include repeated measures analysis of variance (ANOVA), Bland-Altman analysis, and coefficients of variation (CoVs). A P-value <.05 was considered statistically significant. RESULTS: Optimization of the four breath-hold acquisitions resulted in 40 Hz vibration frequency, five wave-phases, and echo time (TE) = 6.9 msec as the preferred method (4BH-MRE). CS-MRE quantitative results did not differ from 4BH-MRE. Shear wave speed (SWS) and phase angle differed significantly between HV and PDAC patients using 4BH-MRE or CS-MRE. The limits of agreement for SWS were [-0.09, 0.10] m/second and the within-subject CoV was 4.8% for CS-MRE. DATA CONCLUSION: CS-MRE might allow a single breath-hold MRE acquisition with comparable SWS and phase angle as 4BH-MRE, and it may still enable to differentiate between HV and PDAC. LEVEL OF EVIDENCE: 2 Technical Efficacy Stage: 2.


Asunto(s)
Diagnóstico por Imagen de Elasticidad , Neoplasias Pancreáticas , Humanos , Masculino , Femenino , Estudios Prospectivos , Diagnóstico por Imagen de Elasticidad/métodos , Reproducibilidad de los Resultados , Contencion de la Respiración , Páncreas/diagnóstico por imagen , Neoplasias Pancreáticas/diagnóstico por imagen , Imagen por Resonancia Magnética/métodos
7.
J Magn Reson Imaging ; 2024 Oct 25.
Artículo en Inglés | MEDLINE | ID: mdl-39449547

RESUMEN

BACKGROUND: Cardiac magnetic resonance elastography (MRE) shows promise in assessing the mechanofunctional properties of the heart but faces clinical challenges, mainly synchronization with cardiac cycle, breathing, and external harmonic stimulation. PURPOSE: To determine the reproducibility of in vivo cardiac multifrequency MRE (MMRE) for assessing diastolic left ventricular (LV) stiffness and viscosity. STUDY TYPE: Prospective. SUBJECTS: This single-center study included a total of 28 participants (mean age, 56.6 ± 23.0 years; 16 male) consisting of randomly selected healthy participants (mean age, 44.6 ± 20.1 years; 9 male) and patients with aortic stenosis (mean age, 78.3 ± 3.8 years; 7 male). FIELD STRENGTH/SEQUENCE: 3 T, 3D multifrequency MRE with a single-shot spin-echo planar imaging sequence. ASSESSMENT: Each participant underwent two cardiac MMRE examinations on the same day. Full 3D wave fields were acquired in diastole at frequencies of 80, 90, and 100 Hz during a total of three breath-holds. Shear wave speed (SWS) and penetration rate (PR) were reconstructed as a surrogate for tissue stiffness and inverse viscous loss. Epicardial and endocardial ROIs were manually drawn by two independent readers to segment the LV myocardium. STATISTICAL TESTS: Shapiro-Wilk test, Bland-Altman analysis and intraclass correlation coefficient (ICC). P-value <0.05 were considered statistically significant. RESULTS: Bland-Altman analyses and intraclass correlation coefficients (ICC = 0.96 for myocardial stiffness and ICC = 0.93 for viscosity) indicated near-perfect test-retest repeatability among examinations on the same day. The mean SWS for scan and re-scan diastolic LV myocardium were 2.42 ± 0.24 m/s and 2.39 ± 0.23 m/s; the mean PR were 1.24 ± 0.17 m/s and 1.22 ± 0.14 m/s. Inter-reader variability showed good to excellent agreement for myocardial stiffness (ICC = 0.92) and viscosity (ICC = 0.85). DATA CONCLUSION: Cardiac MMRE is a promising and reproducible method for noninvasive assessment of diastolic LV stiffness and viscosity. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY: 1.

8.
J Magn Reson Imaging ; 59(5): 1582-1592, 2024 May.
Artículo en Inglés | MEDLINE | ID: mdl-37485870

RESUMEN

BACKGROUND: Pancreatic ductal adenocarcinoma (PDAC) stromal viscoelasticity can be measured using MR elastography (MRE). Bowel preparation regimens could affect MRE quality and knowledge on repeatability is crucial for clinical implementation. PURPOSE: To assess effects of four bowel preparation regimens on MRE quality and to evaluate repeatability and differentiate patients from healthy controls. STUDY TYPE: Prospective. POPULATION: 15 controls (41 ± 16 years; 47% female), 16 PDAC patients (one excluded, 66 ± 12 years; 40% female) with 15 age-/sex-matched controls (65 ± 11 years; 40% female). Final sample size was 25 controls and 15 PDAC. FIELD STRENGTH/SEQUENCE: 3-T, spin-echo echo-planar-imaging, turbo spin-echo, and fast field echo gradient-echo. ASSESSMENT: Four different regimens were used: fasting; scopolaminebutyl; drinking 0.5 L water; combination of 0.5 L water and scopolaminebutyl. MRE signal-to-noise ratio (SNR) was compared between all regimens. MRE repeatability (test-retest) and differences in shear wave speed (SWS) and phase angle (ϕ) were assessed in PDAC and controls. Regions-of-interest were defined for tumor, nontumorous (n = 8) tissue in PDAC, and whole pancreas in controls. Two radiologists delineated tumors twice for evaluation of intraobserver and interobserver variability. STATISTICAL TESTS: Repeated measures analysis of variance, coefficients of variation (CoVs), Bland-Altman analysis, (un)paired t-test, Mann-Whitney U-test, and Wilcoxon signed-rank test. P-value<0.05 was considered statistically significant. RESULTS: Preparation regimens did not significantly influence MRE-SNR. Therefore, the least burdensome preparation (fasting only) was continued. CoVs for tumor SWS were: intrasession (12.8%) and intersession (21.7%), and intraobserver (7.9%) and interobserver (10.3%) comparisons. For controls, CoVs were intrasession (4.6%) and intersession (6.4%). Average SWS for tumor, nontumor, and healthy tissue were: 1.74 ± 0.58, 1.38 ± 0.27, and 1.18 ± 0.16 m/sec (ϕ: 1.02 ± 0.17, 0.91 ± 0.07, and 0.85 ± 0.08 rad), respectively. Significant differences were found between all groups, except for ϕ between healthy-nontumor (P = 0.094). DATA CONCLUSION: The proposed bowel preparation regimens may not influence MRE quality. MRE may be able to differentiate between healthy tissue-tumor and tumor-nontumor. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY STAGE: 2.


Asunto(s)
Diagnóstico por Imagen de Elasticidad , Neoplasias Pancreáticas , Humanos , Femenino , Persona de Mediana Edad , Anciano , Masculino , Imagen por Resonancia Magnética/métodos , Diagnóstico por Imagen de Elasticidad/métodos , Estudios Prospectivos , Páncreas/diagnóstico por imagen , Neoplasias Pancreáticas/diagnóstico por imagen , Reproducibilidad de los Resultados , Agua
9.
J Magn Reson Imaging ; 2024 Oct 24.
Artículo en Inglés | MEDLINE | ID: mdl-39446078

RESUMEN

BACKGROUND: Intrinsic activation MR elastography (iMRE) uses cardiovascular pulsations to assess tissue viscoelastic properties. Applying it to focal liver lesions extends its capabilities. PURPOSE: To assess the viscoelastic parameters of focal liver lesions measured by iMRE and compare its diagnostic performance with extrinsic MRE (eMRE) for differentiating malignant and benign lesions. STUDY TYPE: Prospective. POPULATION: A total of 55 participants underwent MRI with research MRE sequences; 32 participants with 17 malignant and 15 benign lesions underwent both iMRE and eMRE. FIELD STRENGTH/SEQUENCE: iMRE at ~1 Hz heart rate used a 3 T scanner with a modified four-dimensional (4D)-quantitative flow gradient-echo phase contrast and low-velocity encoding cardiac-triggered technique. eMRE employed a gradient-echo sequence at 30, 40, and 60 Hz. ASSESSMENT: Liver displacements were measured using 4D-phase contrast and reconstructed via a nonlinear inversion algorithm to determine shear stiffness (SS) and damping ratio (DR). iMRE parameters were normalized to the corresponding values from the spleen. Lesions were manually segmented, and image quality was reviewed. STATISTICAL TESTS: Kruskal-Wallis, Mann-Whitney, Dunn's test, and areas under receiver operating characteristic curves (AUC) were assessed. RESULTS: SS was significantly higher in malignant than benign lesions with iMRE at 1 Hz (3.69 ± 1.31 vs. 1.63 ± 0.45) and eMRE at 30 Hz (3.76 ± 1.12 vs. 2.60 ± 1.26 kPa), 40 Hz (3.76 ± 1.12 vs. 2.60 ± 1.26 kPa), and 60 Hz (7.32 ± 2.87 vs. 2.48 ± 1.12 kPa). DR was also significantly higher in malignant than benign lesions at 40 Hz (0.36 ± 0.11 vs. 0.21 ± 0.01) and 60 Hz (0.89 ± 0.86 vs. 0.22 ± 0.09). The AUC were 0.86 for iMRE SS, 0.87-0.98 for eMRE SS, 0.47 for iMRE DR, and 0.62-0.86 for eMRE DR. DATA CONCLUSION: Cardiac-activated iMRE can characterize liver lesions and differentiate malignant from benign lesions through normalized SS maps. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY: Stage 2.

10.
J Magn Reson Imaging ; 2024 Aug 21.
Artículo en Inglés | MEDLINE | ID: mdl-39165139

RESUMEN

BACKGROUND: Magnetic resonance elastography (MRE) can quantify tissue biomechanics noninvasively, including pathological hepatic states like metabolic dysfunction-associated steatohepatitis. PURPOSE: To compare the performance of 2D/3D-MRE using the gravitational (GT) transducer concept with the current commercial acoustic (AC) solution utilizing a 2D-MRE approach. Additionally, quality index markers (QIs) were proposed to identify image pixels with sufficient quality for reliably estimating tissue biomechanics. STUDY TYPE: Prospective. POPULATION: One hundred seventy participants with suspected or confirmed liver disease (median age, 57 years [interquartile range (IQR), 46-65]; 66 females), and 11 healthy volunteers (median age, 31 years [IQR, 27-34]; 5 females). FIELD STRENGTH/SEQUENCE: Participants were scanned twice at 1.5 T and 60 Hz vibration frequency: first, using AC-MRE (2D-MRE, spin-echo EPI sequence, 11 seconds breath-hold), and second, using GT-MRE (2D- and 3D-MRE, gradient-echo sequence, 14 seconds breath-hold). ASSESSMENT: Image analysis was performed by four independent radiologists and one biomedical engineer. Additionally, superimposed analytic plane shear waves of known wavelength and attenuation at fixed shear modulus were used to propose pertinent QIs. STATISTICAL TESTS: Spearman's correlation coefficient (r) was applied to assess the correlation between modalities. Interreader reproducibility was evaluated using Bland-Altman bias and reproducibility coefficients. P-values <0.05 were considered statistically significant. RESULTS: Liver stiffness quantified via GT-2D/3D correlated well with AC-2D (r ≥ 0.89 [95% CI: 0.85-0.92]) and histopathological grading (r ≥ 0.84 [95% CI: 0.72-0.91]), demonstrating excellent agreement in Bland-Altman plots and between readers (κ ≥ 0.86 [95% CI: 0.81-0.91]). However, GT-2D showed a bias in overestimating stiffness compared to GT-3D. Proposed QIs enabled the identification of pixels deviating beyond 10% from true stiffness based on a combination of total wave amplitude, temporal sinusoidal nonlinearity, and wave signal-to-noise ratio for GT-3D. CONCLUSION: GT-MRE represents an alternative to AC-MRE for noninvasive liver tissue characterization. Both GT-2D and 3D approaches correlated strongly with the established commercial approach, offering advanced capabilities in abdominal imaging compared to AC-MRE. EVIDENCE LEVEL: 1 TECHNICAL EFFICACY: Stage 2.

11.
J Magn Reson Imaging ; 2024 Mar 06.
Artículo en Inglés | MEDLINE | ID: mdl-38449389

RESUMEN

BACKGROUND: Different MR elastography (MRE) systems may produce different stiffness measurements, making direct comparison difficult in multi-center investigations. PURPOSE: To assess the repeatability and reproducibility of liver stiffness measured by three typical MRE systems. STUDY TYPE: Prospective. POPULATION/PHANTOMS: Thirty volunteers without liver disease history (20 males, aged 21-28)/5 gel phantoms. FIELD STRENGTH/SEQUENCE: 3.0 T United Imaging Healthcare (UIH), 1.5 T Siemens Healthcare, 3.0 T General Electric Healthcare (GE)/Echo planar imaging-based MRE sequence. ASSESSMENT: Wave images of volunteers and phantoms were acquired by three MRE systems. Tissue stiffness was evaluated by two observers, while phantom stiffness was assessed automatically by code. The reproducibility across three MRE systems was quantified based on the mean stiffness of each volunteer and phantom. STATISTICAL TESTS: Intraclass correlation coefficients (ICC), coefficients of variation (CV), and Bland-Altman analyses were used to assess the interobserver reproducibility, the interscan repeatability, and the intersystem reproducibility. Paired t-tests were performed to assess the interobserver and interscan variation. Friedman tests with Dunn's multiple comparison correction were performed to assess the intersystem variation. P values less than 0.05 indicated significant difference. RESULTS: The reproducibility of stiffness measured by the two observers demonstrated consistency with ICC > 0.92, CV < 4.32%, Mean bias < 2.23%, and P > 0.06. The repeatability of measurements obtained using the electromagnetic system for the liver revealed ICC > 0.96, CV < 3.86%, Mean bias < 0.19%, P > 0.90. When considering the range of reproducibility across the three systems for liver evaluations, results ranged with ICCs from 0.70 to 0.87, CVs from 6.46% to 10.99%, and Mean biases between 1.89% and 6.30%. Phantom studies showed similar results. The values of measured stiffness differed across all three systems significantly. DATA CONCLUSION: Liver stiffness values measured from different MRE systems can be different, but the measurements across the three MRE systems produced consistent results with excellent reproducibility. EVIDENCE LEVEL: 1 TECHNICAL EFFICACY: Stage 2.

12.
J Magn Reson Imaging ; 59(1): 97-107, 2024 01.
Artículo en Inglés | MEDLINE | ID: mdl-37158252

RESUMEN

BACKGROUND: Nonalcoholic fatty liver disease (NAFLD) is increasing worldwide and is a growing cause of liver cirrhosis and cancer. The performance of the magnetic resonance elastography (MRE) visco-elastic parameters in diagnosing progressive forms of NAFLD, including nonalcoholic steatohepatitis (NASH) and substantial fibrosis (F ≥ 2), needs to be clarified. PURPOSE: To assess the value of three-dimensional MRE visco-elastic parameters as markers of NASH and substantial fibrosis in mice with NAFLD. STUDY TYPE: Prospective. ANIMAL MODEL: Two mouse models of NAFLD were induced by feeding with high fat diet or high fat, choline-deficient, amino acid-defined diet. FIELD STRENGTH/SEQUENCE: 7T/multi-slice multi-echo spin-echo MRE at 400 Hz with motion encoding in the three spatial directions. ASSESSMENT: Hepatic storage and loss moduli were calculated. Histological analysis was based on the NASH Clinical Research Network criteria. STATISTICAL TESTS: Mann-Whitney, Kruskal-Wallis tests, Spearman rank correlations and multiple regressions were used. Diagnostic performance was assessed with areas under the receiver operating characteristic curves (AUCs). P value <0.05 was considered significant. RESULTS: Among the 59 mice with NAFLD, 21 had NASH and 20 had substantial fibrosis (including 8 mice without and 12 mice with NASH). The storage and loss moduli had similar moderate accuracy for diagnosing NASH with AUCs of 0.67 and 0.66, respectively. For diagnosing substantial fibrosis, the AUC of the storage modulus was 0.73 and the AUC of the loss modulus was 0.81, indicating good diagnostic performance. Using Spearman correlations, histological fibrosis, inflammation and steatosis, but not ballooning, were significantly correlated with the visco-elastic parameters. Using multiple regression, fibrosis was the only histological feature independently associated with the visco-elastic parameters. CONCLUSION: MRE in mice with NAFLD suggests that the storage and loss moduli have good diagnostic performance for detecting progressive NAFLD defined as substantial fibrosis rather than NASH. EVIDENCE LEVEL: 1 TECHNICAL EFFICACY STAGE: 2.


Asunto(s)
Diagnóstico por Imagen de Elasticidad , Enfermedad del Hígado Graso no Alcohólico , Animales , Ratones , Enfermedad del Hígado Graso no Alcohólico/diagnóstico por imagen , Diagnóstico por Imagen de Elasticidad/métodos , Estudios Prospectivos , Biopsia , Hígado/diagnóstico por imagen , Hígado/patología , Cirrosis Hepática/diagnóstico por imagen , Cirrosis Hepática/etiología , Fibrosis
13.
J Magn Reson Imaging ; 59(1): 82-96, 2024 01.
Artículo en Inglés | MEDLINE | ID: mdl-37084171

RESUMEN

Neurodegenerative diseases such as Alzheimer's disease (AD) and Parkinson's disease (PD) present a major health burden to society. Changes in brain structure and cognition are generally only observed at the late stage of the disease. Although advanced magnetic resonance imaging (MRI) techniques such as diffusion imaging may allow identification of biomarkers at earlier stages of neurodegeneration, early diagnosis is still challenging. Magnetic resonance elastography (MRE) is a noninvasive MRI technique for studying the mechanical properties of tissues by measuring the wave propagation induced in the tissues using a purpose-built actuator. Here, we present a systematic review of preclinical and clinical studies in which MRE has been applied to study neurodegenerative diseases. Actuator systems for data acquisition, inversion algorithms for data analysis, and sample demographics are described and tissue stiffness measures obtained for the whole brain and internal structures are summarized. A total of six animal studies and eight human studies have been published. The animal studies refer to 123 experimental animals (68 AD and 55 PD) and 121 wild-type animals, while the human studies refer to 142 patients with neurodegenerative disease (including 56 AD and 17 PD) and 166 controls. The animal studies are consistent in the reporting of decreased stiffness of the hippocampal region in AD mice. However, in terms of disease progression, although consistent decreases in either storage modulus or shear modulus magnitude are reported for whole brain, there is variation in the results reported for the hippocampal region. The clinical studies are consistent in reports of a significant decrease in either whole brain storage modulus or shear modulus magnitude, in both AD and PD and with different brain structures affected in different neurodegenerative diseases. MRE studies of neurodegenerative diseases are still in their infancy, and in future it will be interesting to investigate potential relationships between brain mechanical properties and clinical measures, which may help elucidate the mechanisms underlying onset and progression of neurodegenerative diseases. EVIDENCE LEVEL: 1. TECHNICAL EFFICACY: Stage 2.


Asunto(s)
Enfermedad de Alzheimer , Diagnóstico por Imagen de Elasticidad , Enfermedades Neurodegenerativas , Enfermedad de Parkinson , Humanos , Ratones , Animales , Enfermedades Neurodegenerativas/diagnóstico por imagen , Diagnóstico por Imagen de Elasticidad/métodos , Imagen por Resonancia Magnética/métodos , Encéfalo/diagnóstico por imagen , Enfermedad de Alzheimer/diagnóstico por imagen
14.
Liver Int ; 44(7): 1700-1714, 2024 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-38558221

RESUMEN

BACKGROUND & AIMS: The predictors of progression from steatosis to more advanced stages of metabolic dysfunction-associated steatotic liver disease (MASLD) remain unclear. We evaluated the association between the quantity of hepatic steatosis and longitudinal changes in liver stiffness measurements (LSMs) using magnetic resonance elastography (MRE) in patients with MASLD. METHODS: We retrospectively analysed patients with MASLD who underwent at least two serial MRE and magnetic resonance imaging-based proton density fat fraction (MRI-PDFF) examinations at least 1 year apart. Fine-Gray competitive proportional hazard regression was used to identify LSM progression and regression factors. RESULTS: A total of 471 patients were enrolled. Factors linked to LSM progression were steatosis grade 3 (MRI-PDFF ≥17.1%, adjusted hazard ratio [aHR] 2.597; 95% confidence interval [CI] 1.483-4.547) and albumin-bilirubin grade 2 or 3 (aHR 2.790; 95% CI 1.284-6.091), while the only factor linked to LSM regression was % decrease rate of MRI-PDFF ≥5% (aHR 2.781; 95% CI 1.584-4.883). Steatosis grade 3 correlated with a higher incidence rate of LSM progression than steatosis grade 1 (MRI-PDFF <11.3%) in patients with LSM stage 0 (<2.5 kilopascal [kPa]), and a % annual decrease rate of MRI-PDFF ≥5% correlated with a higher incidence rate of LSM regression than that of MRI-PDFF >-5% and <5% in patients with LSM stage 1 or 2-4 (≥2.5 kPa). CONCLUSIONS: Severe hepatic steatosis was linked to significant LSM progression in patients with MASLD and low LSM (<2.5 kPa).


Asunto(s)
Progresión de la Enfermedad , Diagnóstico por Imagen de Elasticidad , Hígado Graso , Hígado , Imagen por Resonancia Magnética , Humanos , Masculino , Femenino , Estudios Retrospectivos , Persona de Mediana Edad , Hígado/patología , Hígado/diagnóstico por imagen , Hígado Graso/diagnóstico por imagen , Hígado Graso/patología , Anciano , Adulto , Índice de Severidad de la Enfermedad , Modelos de Riesgos Proporcionales , Enfermedad del Hígado Graso no Alcohólico/diagnóstico por imagen , Enfermedad del Hígado Graso no Alcohólico/complicaciones , Enfermedad del Hígado Graso no Alcohólico/patología
15.
Eur Radiol ; 34(11): 6980-6991, 2024 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-38767658

RESUMEN

OBJECTIVES: To investigate associations between tissue diffusion, stiffness, and different tumor microenvironment features in resected hepatocellular carcinoma (HCC). METHODS: Seventy-two patients were prospectively included for preoperative magnetic resonance (MR) diffusion-weighted imaging and MR elastography examination. The mean apparent diffusion coefficient (ADC) and stiffness value were measured on the central three slices of the tumor and peri-tumor area. Cell density, tumor-stroma ratio (TSR), lymphocyte-rich HCC (LR-HCC), and CD8 + T cell infiltration were estimated in resected tumors. The interobserver agreement of MRI measurements and subjective pathological evaluation was assessed. Variables influencing ADC and stiffness were screened with univariate analyses, and then identified with multivariable linear regression. The potential relationship between explored imaging biomarkers and histopathological features was assessed with linear regression after adjustment for other influencing factors. RESULTS: Seventy-two patients (male/female: 59/13, mean age: 56 ± 10.2 years) were included for analysis. Inter-reader agreement was good or excellent regarding MRI measurements and histopathological evaluation. No correlation between tumor ADC and tumor stiffness was found. Multivariable linear regression confirmed that cell density was the only factor associated with tumor ADC (Estimate = -0.03, p = 0.006), and tumor-stroma ratio was the only factor associated with tumor stiffness (Estimate = -0.18, p = 0.03). After adjustment for fibrosis stage (Estimate = 0.43, p < 0.001) and age (Estimate = 0.04, p < 0.001) in the multivariate linear regression, intra-tumoral CD8 + T cell infiltration remained a significant factor associated with peri-tumor stiffness (Estimate = 0.63, p = 0.02). CONCLUSIONS: Tumor ADC surpasses tumor stiffness as a biomarker of cellularity. Tumor stiffness is associated with tumor-stroma ratio and peri-tumor stiffness might be an imaging biomarker of intra-tumoral immune microenvironment. CLINICAL RELEVANCE STATEMENT: Tissue stiffness could potentially serve as an imaging biomarker of the intra-tumoral immune microenvironment of hepatocellular carcinoma and aid in patient selection for immunotherapy. KEY POINTS: Apparent diffusion coefficient reflects cellularity of hepatocellular carcinoma. Tumor stiffness reflects tumor-stroma ratio of hepatocellular carcinoma and is associated with tumor-infiltrating lymphocytes. Tumor and peri-tumor stiffness might serve as imaging biomarkers of intra-tumoral immune microenvironment.


Asunto(s)
Carcinoma Hepatocelular , Imagen de Difusión por Resonancia Magnética , Diagnóstico por Imagen de Elasticidad , Neoplasias Hepáticas , Microambiente Tumoral , Humanos , Carcinoma Hepatocelular/diagnóstico por imagen , Carcinoma Hepatocelular/patología , Masculino , Femenino , Neoplasias Hepáticas/diagnóstico por imagen , Neoplasias Hepáticas/patología , Persona de Mediana Edad , Diagnóstico por Imagen de Elasticidad/métodos , Imagen de Difusión por Resonancia Magnética/métodos , Estudios Prospectivos , Anciano , Adulto
16.
Eur Radiol ; 2024 Oct 23.
Artículo en Inglés | MEDLINE | ID: mdl-39438331

RESUMEN

OBJECTIVES: To inform clinical monitoring of children and young adults with metabolic dysfunction-associated steatotic liver disease (MASLD) by characterizing the real-world natural history of MASLD and identifying baseline predictors of liver disease progression. MATERIALS AND METHODS: This retrospective study included consecutive patients ages < 23 years with MASLD who underwent serial MR elastography (MRE) and/or MR fat fraction (FF) examinations between 09/2009 and 11/2022. Outcomes of MASLD were defined based on maximum ratio values. A relative change ≥ 19% in liver stiffness measures (LSM) and an absolute change ≥ 5% for liver FF were considered clinically meaningful. Random intercept models characterized the yearly rate of change in LSM (kilopascals per year) and FF (percentage per year). RESULTS: One hundred twenty-one patients (87 males, mean age at baseline: 12 ± 3 [SD] years) underwent 297 MRE examinations. The mean interval between the first and last MRE was 34 (± 24) months (range: 1-120 months). Among the 114 patients with serial LSM, 33% (38/114) showed progression, 46% (53/114) remained stable, and 21% (23/114) showed regression. Among the 88 patients with serial FF measures, 57% (50/88) showed progression, 2% (2/88) remained stable, and 41% (36/88) showed regression. LSM progression was associated with Hispanic ethnicity, baseline BMI-for-age percentile, baseline mean liver FF, and GGT changes over time. Predictors for liver FF progression included ALT, AST, GGT, and LDL. CONCLUSION: In a real-world sample of children and young adults with MASLD who underwent serial liver MRI, a minority of patients demonstrated improvements in liver stiffness or FF over time. KEY POINTS: Question In children, there is scarce data regarding the natural history of MASLD. Findings In this retrospective study, most children and young adults with MASLD had either unchanged or worsening liver stiffness (n = 91/114, 79%) and liver fat (n = 52/88, 59%). Clinical relevance Our findings emphasize the need for optimized care in pediatric MASLD. The identified risk factors for the progression of liver fat and stiffness may help to identify children who require interventions beyond changes in lifestyle.

17.
Hepatol Res ; 54(9): 851-858, 2024 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-38349813

RESUMEN

AIM: This study aimed to establish the shear wave measurement (SWM) cut-off value for each fibrosis stage using magnetic resonance (MR) elastography values as a reference standard. METHODS: We prospectively analyzed 594 patients with chronic liver disease who underwent SWM and MR elastography. Correlation coefficients (were analyzed, and the diagnostic value was evaluated by the area under the receiver operating characteristic curve. Liver stiffness was categorized by MR elastography as F0 (<2.61 kPa), F1 (≥2.61 kPa, <2.97 kPa, any fibrosis), F2 (≥2.97 kPa, <3.62 kPa, significant fibrosis), F3 (≥3.62 kPa, <4.62 kPa, advanced fibrosis), or F4 (≥4.62 kPa, cirrhosis). RESULTS: The median SWM values increased significantly with increasing fibrosis stage (p < 0.001). The correlation coefficient between SWM and MR elastography values was 0.793 (95% confidence interval 0.761-0.821). The correlation coefficients between SWM and MR elastography values significantly decreased with increasing body mass index and skin-capsular distance; skin-capsular distance values were associated with significant differences in sensitivity, specificity, accuracy, or positive predictive value, whereas body mass index values were not. The best cut-off values for any fibrosis, significant fibrosis, advanced fibrosis, and cirrhosis were 6.18, 7.09, 8.05, and 10.89 kPa, respectively. CONCLUSIONS: This multicenter study in a large number of patients established SWM cut-off values for different degrees of fibrosis in chronic liver diseases using MR elastography as a reference standard. It is expected that these cut-off values will be applied to liver diseases in the future.

18.
Artículo en Inglés | MEDLINE | ID: mdl-39291624

RESUMEN

BACKGROUND: Metabolic dysfunction-associated steatotic liver disease (MASLD) is the most common liver disease in children. Liver biopsy is considered the gold standard for diagnosis. Magnetic resonance elastography (MRE) and transient elastography (TE) are imaging modalities that can monitor fibrosis and steatosis noninvasively. More studies are needed to identify whether imaging modalities can provide accurate and reproducible data. We hypothesize that MRE provides reliable data similar to that of TE when compared to liver biopsy in children with MASLD/metabolic dysfunction-associated steatohepatitis. METHODS: We conducted a retrospective chart review of children with liver biopsy-proven MASLD at Children's Hospital Los Angeles between September 2017 and January 2023, investigating and comparing the predictive accuracy of MRE and TE in the detection of high-grade fibrosis on liver biopsy. RESULTS: Seventy-seven patients were reviewed, all of whom had undergone liver biopsy, MRE and TE for evaluation of MASLD. Fibrosis was identified in 90% of liver biopsies. The area under the receiver operating characteristic curves (AUROC) of MRE and TE for detection of high-grade fibrosis were 0.817 and 0.750, respectively, and not significantly different (p = 0.4785). CONCLUSION: We demonstrate that MRE and TE did not accurately predict high-grade fibrosis on liver biopsy. Between the two noninvasive imaging modalities, the correlation of identifying high-grade fibrosis was not statistically different; however, the AUROC for MRE was slightly superior to that of TE. Studies with larger cohorts will be required to validate these findings.

19.
Neuroradiology ; 2024 Oct 21.
Artículo en Inglés | MEDLINE | ID: mdl-39432071

RESUMEN

INTRODUCTION: Prognoses for pediatric brain tumors are suboptimal, as even in low-grade tumors, management techniques can lead to damage in the developing brain. Therefore, advanced neuroimaging methods are critical for developing optimal management plans and improving patient care. Magnetic resonance elastography (MRE) has allowed for the characterization of adult gliomas by their mechanical properties, which are uniquely sensitive to the complex interplay of cellularity, vasculature, and interstitium. However, pediatric tumors differ in behavior and cytoarchitecture, and their mechanical properties have never been assessed. METHODS: Here, we conduct the first study of pediatric brain tumor mechanical properties by using MRE to measure tissue stiffness and damping ratio in low grade gliomas (LGGs). We additionally measure the mechanical properties of non-neoplastic focal abnormal signal intensities (FASIs) in children with neurofibromatosis type 1 (NF1). RESULTS: 23 patients age 4-17 years who had MR imaging results consistent with a primary LGG or with NF1 were included in this study. We found that pediatric gliomas are on an average 10.9% softer (p = 0.010) with a 17.3% lower (p = 0.009) viscosity than reference tissue. Softness of tumors appeared consistent across tumor subtypes and unrelated to tumor size or contrast-enhancement. In NF1 we found that, unlike gliomas, FASIs are stiffer, though not significantly, than reference tissue by an average of 10.4% and have a 16.7% lower damping ratio. CONCLUSIONS: Measuring tumor mechanical properties patterning and heterogeneity has potential to aid in prediction of biological behavior and inform management strategies for pediatric patients.

20.
Acta Radiol ; 65(5): 414-421, 2024 May.
Artículo en Inglés | MEDLINE | ID: mdl-38342993

RESUMEN

BACKGROUND: Current liver magnetic resonance elastography (MRE) scans often require adjustments to driver amplitude to produce acceptable images. This could lead to time wastage and the potential loss of an opportunity to capture a high-quality image. PURPOSE: To construct a linear regression model of individualized driver amplitude to improve liver MRE image quality. MATERIAL AND METHODS: Data from 95 liver MRE scans of 61 participants, including abdominal missing volume ratio (AMVR), breath-holding status, the distance from the passive driver on the skin surface to the liver edge (Dd-l), body mass index (BMI), and lateral deflection of the passive driver with respect to the human sagittal plane (Angle α), were continuously collected. The Spearman correlation analysis and lasso regression were conducted to screen the independent variables. Multiple linear regression equations were developed to determine the optimal amplitude prediction model. RESULTS: The optimal formula for linear regression models: driver amplitude (%) = -16.80 + 78.59 × AMVR - 11.12 × breath-holding (end of expiration = 1, end of inspiration = 0) + 3.16 × Dd-l + 1.94 × BMI + 0.34 × angle α, with the model passing the F test (F = 22.455, P <0.001) and R2 value of 0.558. CONCLUSION: The individualized amplitude prediction model based on AMVR, breath-holding status, Dd-l, BMI, and angle α is a valuable tool in liver MRE examination.


Asunto(s)
Diagnóstico por Imagen de Elasticidad , Hígado , Imagen por Resonancia Magnética , Humanos , Diagnóstico por Imagen de Elasticidad/métodos , Masculino , Femenino , Modelos Lineales , Hígado/diagnóstico por imagen , Persona de Mediana Edad , Adulto , Imagen por Resonancia Magnética/métodos , Anciano , Contencion de la Respiración , Adulto Joven
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA