Acute changes in aortic wall mechanical properties after stent placement in rabbits.
Vernhet, H; Juan, J M; Demaria, R; Oliva-Lauraire, M C; Senac, J P; Dauzat, M.
Afiliación
  • Vernhet H; Laboratory of Cardiovascular Physiology, Montpellier I University, Faculty of Medicine, Nimes, France.
J Vasc Interv Radiol ; 11(5): 634-8, 2000 May.
Article en En | MEDLINE | ID: mdl-10834497
ABSTRACT

PURPOSE:

To evaluate mechanical property changes after endovascular stent placement in small-diameter arteries. MATERIALS AND

METHODS:

Self-expanding stents (Wallstent) were placed in the infrarenal aorta of five New Zealand White rabbits via a surgical right femoral approach. Blood pressure changes (deltaP) were monitored in the aorta. Blood flow velocity was measured with a 20-MHz, pulsed Doppler probe (n = 4) to calculate the pulsatility index. Aortic diameter (dA) and diameter changes (delta(d)) were measured with a 20-MHz probe in echo-tracking mode. Diameter compliance (Cd) and distensibility coefficient (DC) were calculated as Cd = 2(delta)d/(delta)P and DC = 2delta(d)/delta(P)/dA.

RESULTS:

Aortic diameter increased from 3.360 +/- 0.4033 mm to 4.020 +/- 0.3033 mm after stent placement at the stent level only. Compliance decreased from 77.644 +/- 24.306 mm kPa(-1) to 31.150 +/-8.245 x 10(-3) mm kPa(-1) at the stent level, and was then significantly lower than upstream (98.500 +/- 53.196 mm kPa(-1)) and down-stream (59.047 +/- 13.833 mm kPa(-1)). There was no significant change in pulsatility index.

CONCLUSIONS:

Endovascular stent placement produces a significant decrease in arterial wall compliance of the rabbit abdominal aorta.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Aorta Abdominal / Prótesis Vascular / Stents Tipo de estudio: Evaluation_studies Límite: Animals Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 2000 Tipo del documento: Article País de afiliación: Francia
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Aorta Abdominal / Prótesis Vascular / Stents Tipo de estudio: Evaluation_studies Límite: Animals Idioma: En Revista: J Vasc Interv Radiol Asunto de la revista: ANGIOLOGIA / RADIOLOGIA Año: 2000 Tipo del documento: Article País de afiliación: Francia