Your browser doesn't support javascript.
loading
Velocity fluctuation reduction in vector Doppler ultrasound using a hybrid single/dual-beam algorithm.
Article en En | MEDLINE | ID: mdl-12578140
In order to reduce the fluctuations in the velocity magnitude estimate, we propose a modification to the standard algorithm for reconstructing the (two component) vector velocity from the measured Doppler shifts in two directions. This uses the standard dual-beam algorithm, combined with temporal smoothing, to find only the velocity angle, then uses the single-beam algorithm to estimate the velocity magnitude. We present initial data showing the significant reduction in velocity estimate fluctuation that this hybrid method achieves compared to the standard algorithm.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Vasos Sanguíneos / Algoritmos / Aumento de la Imagen / Ultrasonografía Doppler Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: IEEE Trans Ultrason Ferroelectr Freq Control Asunto de la revista: MEDICINA NUCLEAR Año: 2003 Tipo del documento: Article
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Vasos Sanguíneos / Algoritmos / Aumento de la Imagen / Ultrasonografía Doppler Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: IEEE Trans Ultrason Ferroelectr Freq Control Asunto de la revista: MEDICINA NUCLEAR Año: 2003 Tipo del documento: Article