Cutaneous bioengineering instrumentation standardization: the Tissue Viability Imager.
Nilsson, Gert E; Zhai, Hongbo; Chan, Heidi P; Farahmand, Sara; Maibach, Howard I.
Afiliación
  • Nilsson GE; Department of Biomedical Engineering, Linköping University, Linköping, Sweden. gerni@glocalnet.net
Skin Res Technol ; 15(1): 6-13, 2009 Feb.
Article en En | MEDLINE | ID: mdl-19152572
ABSTRACT

BACKGROUND:

Tissue Viability Imaging (TiVi) is a new bioengineering technology intended for remote two-dimensional mapping of skin red blood cell concentration (RBC(conc)). Before use in the laboratory, work-site and dermatology clinic, critical performance parameters of this emerging technology require careful evaluation.

OBJECTIVE:

To assess short- and long-term stability, image uniformity, distance and image size dependence, ambient light and curvature influence in a production batch of Tissue Viability Imagers.

METHODS:

Four Tissue Viability Imagers from the same production batch were evaluated at two laboratories (one industrial and one dermatological) with respect to critical parameter performance.

RESULTS:

The average systematic drift in sensitivity over time was 0.27% and <1.02% for all four units tested. Difference in sensitivity between units was limited to 4.1% and was due to offset rather than gain deviation. Spatial variation in image uniformity was below 3.08% and 1.93% in the corners and centre of an individual image, respectively. This spatial variation could be further reduced to 0.25% and 0.13%, respectively by image normalization. Ambient light from a 40 W bulb or a 11 W fluorescent light source at a distance of 50-60 cm above the object, reduced the recorded values by about 10%, while the camera to object distance and image size had no detectable influence on sensitivity. Curved objects, such as human forearm, demonstrated an edge effect limited to below 10%.

CONCLUSION:

The critical TiVi performance parameters evaluated proved stable in relation to expected variations in skin RBC(conc) over time. Calibration by way of a two-point method may reduce differences in sensitivity between instruments to further facilitate inter-laboratory comparison of results.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Piel / Fenómenos Fisiológicos de la Piel / Análisis Espectral / Supervivencia Tisular / Interpretación de Imagen Asistida por Computador / Dermoscopía / Recuento de Eritrocitos Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: Skin Res Technol Asunto de la revista: DERMATOLOGIA Año: 2009 Tipo del documento: Article País de afiliación: Suecia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Piel / Fenómenos Fisiológicos de la Piel / Análisis Espectral / Supervivencia Tisular / Interpretación de Imagen Asistida por Computador / Dermoscopía / Recuento de Eritrocitos Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans País/Región como asunto: Europa Idioma: En Revista: Skin Res Technol Asunto de la revista: DERMATOLOGIA Año: 2009 Tipo del documento: Article País de afiliación: Suecia