Your browser doesn't support javascript.
loading
Cyclosporine monitoring by fluorescence polarization immunoassay.
Wang, P P; Simpson, E; Meucci, V; Morrison, M; Lunetta, S; Zajac, M; Boeckx, R.
Afiliación
  • Wang PP; Abbott Laboratories, Abbott Park, IL 60064.
Clin Biochem ; 24(1): 55-8, 1991 Feb.
Article en En | MEDLINE | ID: mdl-2060133
The monitoring of cyclosporine levels in whole blood and serum has become a routine procedure for the post-transplant management of immunosuppression. We have developed fluorescence polarization immunoassays for cyclosporine in whole blood and in serum using a monoclonal antibody. These assays are fast (20 determinations in less than 22 min), sensitive (25 micrograms/L for whole blood samples and 10 micrograms/L for serum samples), and precise (CV less than 7% in both assays). Cross-reactivities with AM1 (Metabolite 17) and AM4N (Metabolite 21) are less than 8.5 and 2.5%, respectively, for both assays.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Trasplante de Riñón / Ciclosporinas Límite: Humans Idioma: En Revista: Clin Biochem Año: 1991 Tipo del documento: Article
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Trasplante de Riñón / Ciclosporinas Límite: Humans Idioma: En Revista: Clin Biochem Año: 1991 Tipo del documento: Article