Evaluation of an over-the-counter glycated hemoglobin (A1C) test kit.
Chang, Anna; Frank, Joy; Knaebel, Jennifer; Fullam, Jeanellen; Pardo, Scott; Simmons, David A.
Afiliación
  • Chang A; John Muir Physician Network Clinical Research Center, Concord, California, USA.
J Diabetes Sci Technol ; 4(6): 1495-503, 2010 Nov 01.
Article en En | MEDLINE | ID: mdl-21129346
ABSTRACT

BACKGROUND:

Glycated hemoglobin (A1C) monitoring is an integral component of diabetes management. This study was conducted to evaluate the performance of the A1CNow® SELFCHECK device when used by lay users and health care professionals (HCPs) to measure A1C.

METHODS:

Subjects performed two A1CNow SELFCHECK finger-stick self-tests followed by a finger-stick test of the subject's blood by a HCP. The primary endpoint assessed accuracy of the subject and HCP A1CNow SELFCHECK readings. Secondary endpoints included precision, comprehension of instructional material (written material±DVD), and product satisfaction. For accuracy comparison, a venous blood sample was drawn from each subject and tested by laboratory (TOSOH) analysis. Subject comprehension of product instructional material was evaluated via first-time failure (FTF) rate as recorded by the HCP, and subject satisfaction was assessed through written survey.

RESULTS:

A total of 110 subjects with (n=93) and without (n=17) diabetes participated. Of 177 subject A1C values, 165 (93.2%) were within the acceptable range of ±13.5% of the laboratory reference value and considered accurate. Regression analysis showed good correlation of subject values to laboratory and HCP results (R2=0.93 for both). The average within-subject coefficient of variation was 4.57% (n=74). The FTF rates with and without instructional DVD were 11.3% (n=56) and 39.6% (n=54), respectively. Subjects with diabetes/prediabetes overwhelmingly indicated that they were "very" to "extremely" likely (93.5%) to discuss their home A1C results with their HCP.

CONCLUSIONS:

Lay users found the A1CNow SELFCHECK easy to use, and both lay users and HCPs were able to measure A1C accurately.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estado Prediabético / Juego de Reactivos para Diagnóstico / Hemoglobina Glucada / Sistemas de Atención de Punto / Diabetes Mellitus Tipo 1 / Diabetes Mellitus Tipo 2 Tipo de estudio: Clinical_trials / Prognostic_studies / Qualitative_research Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: J Diabetes Sci Technol Asunto de la revista: ENDOCRINOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estado Prediabético / Juego de Reactivos para Diagnóstico / Hemoglobina Glucada / Sistemas de Atención de Punto / Diabetes Mellitus Tipo 1 / Diabetes Mellitus Tipo 2 Tipo de estudio: Clinical_trials / Prognostic_studies / Qualitative_research Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: J Diabetes Sci Technol Asunto de la revista: ENDOCRINOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos