Performance evaluation of the Bio-Rad Laboratories GS HIV Combo Ag/Ab EIA, a 4th generation HIV assay for the simultaneous detection of HIV p24 antigen and antibodies to HIV-1 (groups M and O) and HIV-2 in human serum or plasma.
Bentsen, Christopher; McLaughlin, Lisa; Mitchell, Elizabeth; Ferrera, Carol; Liska, Sally; Myers, Robert; Peel, Sheila; Swenson, Paul; Gadelle, Stephane; Shriver, M Kathleen.
Afiliación
  • Bentsen C; Bio-Rad Laboratories, 6565 185th Avenue NE, Redmond, WA 98052, USA. christopher bentsen@bio-rad.com
J Clin Virol ; 52 Suppl 1: S57-61, 2011 Dec.
Article en En | MEDLINE | ID: mdl-21995929
ABSTRACT

BACKGROUND:

A multi-center study was conducted to evaluate the Bio-Rad GS HIV Combo Ag/Ab EIA, a 4th generation HIV-1/HIV-2 assay for the simultaneous detection of HIV p24 antigen and antibodies to HIV-1 (groups M and O) and HIV-2 in human serum or plasma in adult and pediatric populations.

OBJECTIVES:

The objectives of the study were to assess assay performance for the detection of acute HIV infections; sensitivity in known HIV positive samples; percent agreement with HIV status; specificity in low and high risk individuals of unknown HIV status; and to compare assay performance to a 3rd generation HIV assay. STUDY

DESIGN:

The evaluation included testing 9150 samples at four U.S. clinical trial sites, using three kit lots. Unlinked samples were from routine testing, repositories or purchased from vendors.

RESULTS:

GS HIV Combo Ag/Ab EIA detection in samples from individuals in two separate populations with acute HIV infection was 95.2% (20/21) and 86.4% (38/44). Sensitivity was 100% (1603/1603) in known antibody positive [HIV-1 Groups M and O, and HIV-2] samples. HIV p24 antigen detection was 100% (53/53) in HIV-1 culture supernatants. HIV-1 seroconversion panel detection improved by a range of 0-20 days compared to a 3rd generation HIV test. Specificity was 99.9% (5989/5996) in low risk, 99.9% (959/960) in high risk and 100% (100/100) in pediatric populations.

CONCLUSION:

The GS HIV Combo Ag/Ab EIA significantly reduced the diagnostic window when compared to the 3rd generation screening assay, enabling earlier diagnosis of HIV infection. The performance parameters of the Bio-Rad GS HIV Combo Ag/Ab EIA are well suited for use in HIV diagnostic settings.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Anticuerpos Anti-VIH / Infecciones por VIH / Técnicas para Inmunoenzimas / VIH-1 / VIH-2 / Proteína p24 del Núcleo del VIH Tipo de estudio: Clinical_trials / Diagnostic_studies / Etiology_studies / Evaluation_studies / Risk_factors_studies Límite: Humans País/Región como asunto: America do norte Idioma: En Revista: J Clin Virol Asunto de la revista: VIROLOGIA Año: 2011 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Anticuerpos Anti-VIH / Infecciones por VIH / Técnicas para Inmunoenzimas / VIH-1 / VIH-2 / Proteína p24 del Núcleo del VIH Tipo de estudio: Clinical_trials / Diagnostic_studies / Etiology_studies / Evaluation_studies / Risk_factors_studies Límite: Humans País/Región como asunto: America do norte Idioma: En Revista: J Clin Virol Asunto de la revista: VIROLOGIA Año: 2011 Tipo del documento: Article