Your browser doesn't support javascript.
loading
Structural valve deterioration in a starr-edwards mitral caged-disk valve prosthesis.
Aoyagi, Shigeaki; Tayama, Kei-Ichiro; Okazaki, Teiji; Shintani, Yusuke; Kono, Michitaka; Wada, Kumiko; Kosuga, Ken-Ichi; Mori, Ryusuke; Tanaka, Hiroyuki.
Afiliación
  • Aoyagi S; Department of Cardiovascular Surgery, Munakata Suikokai General Hospital, Fukutsu, Japan. shigeakiaoyagi@gmail.com
Circ J ; 77(1): 105-8, 2013.
Article en En | MEDLINE | ID: mdl-23001071
ABSTRACT

BACKGROUND:

The durability of the Starr-Edwards (SE) mitral caged-disk valve, model 6520, is not clearly known, and structural valve deterioration in the SE disk valve is very rare. METHODS AND

RESULTS:

Replacement of the SE mitral disk valve was performed in 7 patients 23-40 years after implantation. Macroscopic examination of the removed disk valves showed no structural abnormalities in 3 patients, in whom the disk valves were removed at <26 years after implantation. Localized disk wear was found at the sites where the disk abutted the struts of the cage, in disk valves excised >36 years after implantation in 4 patients. Disk fracture, a longitudinal split in the disk along its circumference at the site of incorporation of the titanium ring, was detected in the valves removed 36 and 40 years after implantation, respectively, and many cracks were also observed on the outflow aspect of the disk removed 40 years after implantation.

CONCLUSIONS:

Disk fracture and localized disk wear were found in the SE mitral disk valves implanted >36 years previously. The present results suggest that SE mitral caged-disk valves implanted >20 years previously should be carefully followed up, and that those implanted >30 years previously should be electively replaced with modern prosthetic valves
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Falla de Prótesis / Prótesis Valvulares Cardíacas / Válvula Mitral Tipo de estudio: Prognostic_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Circ J Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2013 Tipo del documento: Article País de afiliación: Japón
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Falla de Prótesis / Prótesis Valvulares Cardíacas / Válvula Mitral Tipo de estudio: Prognostic_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Circ J Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2013 Tipo del documento: Article País de afiliación: Japón