Placement of implantable cardioverter-defibrillators in paediatric and congenital heart defect patients: a pipeline for model generation and simulation prediction of optimal configurations.
Rantner, Lukas J; Vadakkumpadan, Fijoy; Spevak, Philip J; Crosson, Jane E; Trayanova, Natalia A.
Afiliación
  • Rantner LJ; N. A. Trayanova: Johns Hopkins University, 3400 N Charles St., 216 Hackerman Hall, Baltimore, MD 21218, USA. ntrayanova@jhu.edu.
J Physiol ; 591(17): 4321-34, 2013 Sep 01.
Article en En | MEDLINE | ID: mdl-23798492
There is currently no reliable way of predicting the optimal implantable cardioverter-defibrillator (ICD) placement in paediatric and congenital heart defect (CHD) patients. This study aimed to: (1) develop a new image processing pipeline for constructing patient-specific heart-torso models from clinical magnetic resonance images (MRIs); (2) use the pipeline to determine the optimal ICD configuration in a paediatric tricuspid valve atresia patient; (3) establish whether the widely used criterion of shock-induced extracellular potential (Φe) gradients ≥5 V cm(-1) in ≥95% of ventricular volume predicts defibrillation success. A biophysically detailed heart-torso model was generated from patient MRIs. Because transvenous access was impossible, three subcutaneous and three epicardial lead placement sites were identified along with five ICD scan locations. Ventricular fibrillation was induced, and defibrillation shocks were applied from 11 ICD configurations to determine defibrillation thresholds (DFTs). Two configurations with epicardial leads resulted in the lowest DFTs overall and were thus considered optimal. Three configurations shared the lowest DFT among subcutaneous lead ICDs. The Φe gradient criterion was an inadequate predictor of defibrillation success, as defibrillation failed in numerous instances even when 100% of the myocardium experienced such gradients. In conclusion, we have developed a new image processing pipeline and applied it to a CHD patient to construct the first active heart-torso model from clinical MRIs.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Desfibriladores Implantables / Atresia Tricúspide / Implantación de Prótesis de Válvulas Cardíacas / Modelación Específica para el Paciente / Cardiopatías Congénitas / Modelos Cardiovasculares Tipo de estudio: Prognostic_studies / Risk_factors_studies Límite: Adolescent / Humans / Male Idioma: En Revista: J Physiol Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Desfibriladores Implantables / Atresia Tricúspide / Implantación de Prótesis de Válvulas Cardíacas / Modelación Específica para el Paciente / Cardiopatías Congénitas / Modelos Cardiovasculares Tipo de estudio: Prognostic_studies / Risk_factors_studies Límite: Adolescent / Humans / Male Idioma: En Revista: J Physiol Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos