Your browser doesn't support javascript.
loading
Precision of the iDXA for visceral adipose tissue measurement in severely obese patients.
Carver, Tamara E; Court, Olivier; Christou, Nicolas V; Reid, Ryan E R; Andersen, Ross E.
Afiliación
  • Carver TE; 1Department of Kinesiology and Physical Education, McGill University, Montreal, QC, CANADA; and 2Bariatric Surgery, McGill University Health Center, Montreal, QC, CANADA.
Med Sci Sports Exerc ; 46(7): 1462-5, 2014 Jul.
Article en En | MEDLINE | ID: mdl-24389525
ABSTRACT
UNLABELLED A new measurement tool, the automated software CoreScan, for the GE Lunar iDXA, has been validated for measuring visceral adipose tissue (VAT) against computed tomography in normal-weight populations. However, no study has evaluated the precision of CoreScan in measuring VAT among severely obese patients.

PURPOSE:

The purpose of the study was to evaluate the precision of CoreScan for VAT measurements in severely obese adults (body mass index > 40 kg·m(-2)).

METHODS:

A total of 55 obese participants with a mean age of 46 ± 11 yr, body mass index of 49 ± 6 kg·m(-2), and body mass of 137.3 ± 21.3 kg took part in this study. Two consecutive iDXA scans with repositioning of the total body were conducted for each participant. The coefficient of variation, the root-mean-square averages of SD of repeated measurements, the corresponding 95% least significant change, and intraclass correlations were calculated.

RESULTS:

Precision error was 8.77% (percent coefficient of variation), with a root-mean-square SD of 0.294 kg and an intraclass correlation of 0.96. Bland-Altman plots demonstrated a mean precision bias of -0.08 ± 0.41 kg, giving a coefficient of repeatability of 0.82 kg and a bias range of -0.890 to 0.725 kg.

CONCLUSIONS:

When interpreting VAT results with the iDXA in severely obese populations, clinicians should be aware of the precision error for this important clinical parameter.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Obesidad Mórbida / Absorciometría de Fotón / Grasa Intraabdominal Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Med Sci Sports Exerc Año: 2014 Tipo del documento: Article País de afiliación: Canadá

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Obesidad Mórbida / Absorciometría de Fotón / Grasa Intraabdominal Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Med Sci Sports Exerc Año: 2014 Tipo del documento: Article País de afiliación: Canadá