Your browser doesn't support javascript.
loading
Intertechnique agreement and interstudy reproducibility of strain and diastolic strain rate at 1.5 and 3 Tesla: a comparison of feature-tracking and tagging in patients with aortic stenosis.
Singh, Anvesha; Steadman, Christopher D; Khan, Jamal N; Horsfield, Mark A; Bekele, Soliana; Nazir, Sheraz A; Kanagala, Prathap; Masca, Nicholas G D; Clarysse, Patrick; McCann, Gerry P.
Afiliación
  • Singh A; Department of Cardiovascular Sciences, University of Leicester, Glenfield Hospital, Leicester, United Kingdom; NIHR Leicester Cardiovascular Biomedical Research Unit, Glenfield Hospital, Leicester, United Kingdom.
J Magn Reson Imaging ; 41(4): 1129-37, 2015 Apr.
Article en En | MEDLINE | ID: mdl-24700404
ABSTRACT

PURPOSE:

To determine the interstudy reproducibility of myocardial strain and peak early-diastolic strain rate (PEDSR) measurement on cardiovascular magnetic resonance imaging (MRI) assessed with feature tracking (FT) and tagging, in patients with aortic stenosis (AS). MATERIALS AND

METHODS:

Cardiac MRI was performed twice (1-14 days apart) in 18 patients (8 at 1.5 Tesla [T], 10 at 3T) with moderate-severe AS. Circumferential peak systolic strain (PSS) and PEDSR were measured in all patients. Longitudinal PSS and PEDSR were assessed using FT in all patients, and tagging in the 3T sub-group.

RESULTS:

PSS was higher with FT than tagging (21.0 ± 1.9% versus 17.0 ± 3.4% at 1.5T, 21.4 ± 4.0% versus 17.7 ± 3.0% at 3T, P < 0.05), as was PEDSR (1.3 ± 0.3 s(-1) versus 1.0 ± 0.3 s(-1) , P = 0.10 at 1.5T and 1.3 ± 0.4 s(-1) versus 0.8 ± 0.3 s(-1) , P < 0.05 at 3T). The reproducibility of PSS was excellent with FT (coefficient of variation [CoV] 9-10%) and good with tagging at 1.5T (13-19%). Reproducibility of circumferential PEDSR was best at 1.5T when only basal/mid slices were included (CoV 12%), but moderate to poor at 3T (29-35%). Reproducibility of longitudinal strain was good with FT (10-16%) but moderate for PEDSR (∼30%).

CONCLUSION:

In patients with AS, FT consistently produces higher values compared with tagging. The interstudy reproducibility of PSS is excellent with FT and good with tagging. The reproducibility of circumferential PEDSR at 1.5T is good when only basal and mid slices are used.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Algoritmos / Reconocimiento de Normas Patrones Automatizadas / Interpretación de Imagen Asistida por Computador / Disfunción Ventricular Izquierda / Imagen por Resonancia Cinemagnética Tipo de estudio: Diagnostic_studies / Etiology_studies Límite: Aged / Female / Humans / Male Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2015 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Algoritmos / Reconocimiento de Normas Patrones Automatizadas / Interpretación de Imagen Asistida por Computador / Disfunción Ventricular Izquierda / Imagen por Resonancia Cinemagnética Tipo de estudio: Diagnostic_studies / Etiology_studies Límite: Aged / Female / Humans / Male Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2015 Tipo del documento: Article País de afiliación: Reino Unido