Comparison of diffusion-weighted MRI acquisition techniques for normal pancreas at 3.0 Tesla.
Diagn Interv Radiol
; 20(5): 368-73, 2014.
Article
en En
| MEDLINE
| ID: mdl-25010365
ABSTRACT
PURPOSE:
We aimed to optimize diffusion-weighted imaging (DWI) acquisitions for normal pancreas at 3.0 Tesla. MATERIALS ANDMETHODS:
Thirty healthy volunteers were examined using four DWI acquisition techniques with b values of 0 and 600 s/mm2 at 3.0 Tesla, including breath-hold DWI, respiratory-triggered DWI, respiratory-triggered DWI with inversion recovery (IR), and free-breathing DWI with IR. Artifacts, signal-to-noise ratio (SNR) and apparent diffusion coefficient (ADC) of normal pancreas were statistically evaluated among different DWI acquisitions.RESULTS:
Statistical differences were noticed in artifacts, SNR, and ADC values of normal pancreas among different DWI acquisitions by ANOVA (P <0.001). Normal pancreas imaging had the lowest artifact in respiratory-triggered DWI with IR, the highest SNR in respiratory-triggered DWI, and the highest ADC value in free-breathing DWI with IR. The head, body, and tail of normal pancreas had statistically different ADC values on each DWI acquisition by ANOVA (P < 0.05).CONCLUSION:
The highest image quality for normal pancreas was obtained using respiratory-triggered DWI with IR. Normal pancreas displayed inhomogeneous ADC values along the head, body, and tail structures.
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Páncreas
/
Imagen de Difusión por Resonancia Magnética
Límite:
Adult
/
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Diagn Interv Radiol
Asunto de la revista:
DIAGNOSTICO POR IMAGEM
/
RADIOLOGIA
Año:
2014
Tipo del documento:
Article
País de afiliación:
China