Magnetic-controlled capsule endoscopy vs. gastroscopy for gastric diseases: a two-center self-controlled comparative trial.
Zou, Wen-Bin; Hou, Xiao-Hua; Xin, Lei; Liu, Jun; Bo, Lu-Min; Yu, Guan-Yu; Liao, Zhuan; Li, Zhao-Shen.
Afiliación
  • Zou WB; Digestive Endoscopy Center, Department of Gastroenterology, Shanghai Changhai Hospital, Second Military Medical University, Shanghai, China.
  • Hou XH; Department of Gastroenterology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
  • Xin L; Digestive Endoscopy Center, Department of Gastroenterology, Shanghai Changhai Hospital, Second Military Medical University, Shanghai, China.
  • Liu J; Department of Gastroenterology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
  • Bo LM; Digestive Endoscopy Center, Department of Gastroenterology, Shanghai Changhai Hospital, Second Military Medical University, Shanghai, China.
  • Yu GY; Digestive Endoscopy Center, Department of Gastroenterology, Shanghai Changhai Hospital, Second Military Medical University, Shanghai, China.
  • Liao Z; Digestive Endoscopy Center, Department of Gastroenterology, Shanghai Changhai Hospital, Second Military Medical University, Shanghai, China.
  • Li ZS; Digestive Endoscopy Center, Department of Gastroenterology, Shanghai Changhai Hospital, Second Military Medical University, Shanghai, China.
Endoscopy ; 47(6): 525-8, 2015 Jun.
Article en En | MEDLINE | ID: mdl-25590177
ABSTRACT
BACKGROUND AND STUDY

AIMS:

We developed a novel magnetic-controlled capsule endoscopy (MCE) system for use in the human stomach. The aim of the current study was to compare the diagnostic accuracy of MCE with that of standard gastroscopy for gastric diseases. PATIENTS AND

METHODS:

A total of 68 patients were enrolled in this self-controlled trial. Patients were evaluated by both MCE and gastroscopy. Gastroscopy was performed 4 ­â€Š24 hours after completion of the MCE examination.

RESULTS:

The positive percent agreement between MCE and gastroscopy was 96.0 %, and the negative percent agreement was 77.8 %. The overall agreement was 91.2 % with a kappa value of 0.765 (P < 0.001). A total of 68 pathological findings were detected, of which 53 were identified by both methods. The MCE and standard gastroscopy missed seven and eight findings, respectively.

CONCLUSIONS:

MCE showed a diagnostic accuracy similar to that of standard gastroscopy. These results suggest that MCE is a promising alternative to gastroscopy for noninvasive screening of gastric diseases.Clinical trial registration number NCT01903629.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Gastropatías / Gastroscopía / Endoscopía Capsular / Magnetismo Tipo de estudio: Clinical_trials / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Endoscopy Año: 2015 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Gastropatías / Gastroscopía / Endoscopía Capsular / Magnetismo Tipo de estudio: Clinical_trials / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Endoscopy Año: 2015 Tipo del documento: Article País de afiliación: China