Setup accuracy for prone and supine whole breast irradiation.
Strahlenther Onkol
; 192(4): 254-9, 2016 Apr.
Article
en En
| MEDLINE
| ID: mdl-26864048
PURPOSE: To evaluate cone-beam computed tomography (CBCT) based setup accuracy and margins for prone and supine whole breast irradiation (WBI). METHODS: Setup accuracy was evaluated on 3559 CBCT scans of 242 patients treated with WBI and uncertainty margins were calculated using the van Herk formula. Uni- and multivariate analysis on individual margins was performed for age, body mass index (BMI) and cup size. RESULTS: The population-based margin in vertical (VE), lateral (LA) and longitudinal (LO) directions was 10.4/9.4/9.4 mm for the 103 supine and 10.5/22.4/13.7 mm for the 139 prone treated patients, being significantly (p < 0.01) different for the LA and LO directions. Multivariate analysis identified a significant (p < 0.05) correlation between BMI and the LO margin in supine position and the VE/LA margin in prone position. CONCLUSION: In this series, setup accuracy is significantly worse in prone compared to supine position for the LA and LO directions. However, without proper image-guidance, uncertainty margins of about 1 cm are also necessary for supine WBI. For patients with a higher BMI, larger margins are required.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Planificación de la Radioterapia Asistida por Computador
/
Neoplasias de la Mama
/
Posición Supina
/
Posición Prona
/
Tomografía Computarizada de Haz Cónico
/
Posicionamiento del Paciente
/
Errores de Configuración en Radioterapia
Tipo de estudio:
Etiology_studies
/
Incidence_studies
/
Observational_studies
/
Prognostic_studies
/
Risk_factors_studies
Límite:
Adult
/
Aged
/
Female
/
Humans
/
Middle aged
Idioma:
En
Revista:
Strahlenther Onkol
Asunto de la revista:
NEOPLASIAS
/
RADIOTERAPIA
Año:
2016
Tipo del documento:
Article
País de afiliación:
Bélgica