The detection of oesophageal varices using a novel, disposable, probe-based transnasal endoscope: a prospective diagnostic pilot study.
Sami, Sarmed S; Ragunath, Krish; Wilkes, Emilie A; James, Martin; Mansilla-Vivar, Rodrigo; Ortiz-Fernández-Sordo, Jacobo; White, Jonathan; Khanna, Amardeep; Coletta, Marina; Samuel, Sunil; Aithal, Guruprasad P; Guha, Indra Neil.
Afiliación
  • Sami SS; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Ragunath K; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Wilkes EA; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • James M; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Mansilla-Vivar R; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Ortiz-Fernández-Sordo J; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • White J; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Khanna A; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Coletta M; Gastroenterology and Endoscopy Unit, Fondazione IRCCS Ca' Granda Ospedale Maggiore Policlinico, Department of Pathophysiology and Transplantation, Università degli Studi di Milano, Milan, Italy.
  • Samuel S; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Aithal GP; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK.
  • Guha IN; NIHR Nottingham Digestive Diseases Biomedical Research Unit, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, UK. neil.guha@nottingham.ac.uk.
Liver Int ; 36(11): 1639-1648, 2016 11.
Article en En | MEDLINE | ID: mdl-27125510
ABSTRACT
BACKGROUND &

AIMS:

Screening for oesophageal varices (OV) using conventional oesophagogastroduodenoscopy (C-OGD) is invasive and requires costly monitoring, recovery, and decontamination facilities. We aimed to evaluate the technical feasibility, acceptability and accuracy of a novel, portable and disposable office-based transnasal endoscope (EG Scan™ ) compared to C-OGD as the reference standard.

METHODS:

This was a prospective cohort study. Consecutive adult patients with cirrhosis were invited to participate. All subjects underwent the two procedures on the same day performed by two endoscopists in a blinded design. Patients completed preference and validated tolerability (10-point visual analogue scale (VAS)) questionnaires on day 0 and day 14 post procedures.

RESULTS:

Forty-five of 50 patients (90%) completed both interventions. Mean age was 59 years and OV prevalence was 49%. Patients reported higher preference (percentage) and better experience (mean VAS) with EG Scan compared to C-OGD on day 0 (76.5% vs. 23.5%, P < 0.001; 7.8 vs. 6.8, P = 0.058, respectively) and day 14 (77.8% vs. 22.2%, P < 0.001; 7.0 vs. 5.5, P = 0.0013 respectively). Sensitivity and specificity of the EG Scan for the diagnosis of any size OV were 0.82 (95% confidence interval (CI) 0.60-0.95), and 0.78 (95% CI 0.56-0.93) respectively. Corresponding values for the diagnosis of clinically significant (medium/large) OV were 0.92 (95% CI 0.62-1.0), 0.97 (95% CI 0.84-1.0) respectively. No serious adverse events occurred.

CONCLUSIONS:

EG Scan accuracy was higher for the diagnosis of medium/large OV compared to any size OV. Patients' preference and overall experience of the EG Scan was favourable compared to C-OGD 14 days after procedures.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Várices Esofágicas y Gástricas / Esofagoscopía / Equipos Desechables Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Liver Int Asunto de la revista: GASTROENTEROLOGIA Año: 2016 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Várices Esofágicas y Gástricas / Esofagoscopía / Equipos Desechables Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Liver Int Asunto de la revista: GASTROENTEROLOGIA Año: 2016 Tipo del documento: Article País de afiliación: Reino Unido