Your browser doesn't support javascript.
loading
Semi-automatic carotid intraplaque hemorrhage detection and quantification on Magnetization-Prepared Rapid Acquisition Gradient-Echo (MP-RAGE) with optimized threshold selection.
Liu, Jin; Balu, Niranjan; Hippe, Daniel S; Ferguson, Marina S; Martinez-Malo, Vanesa; DeMarco, J Kevin; Zhu, David C; Ota, Hideki; Sun, Jie; Xu, Dongxiang; Kerwin, William S; Hatsukami, Thomas S; Yuan, Chun.
Afiliación
  • Liu J; University of Washington, Seattle, WA, USA.
  • Balu N; University of Washington, Seattle, WA, USA.
  • Hippe DS; University of Washington, Seattle, WA, USA.
  • Ferguson MS; University of Washington, Seattle, WA, USA.
  • Martinez-Malo V; University of Washington, Seattle, WA, USA.
  • DeMarco JK; Walter Reed National Military Medical Center, Bethesda, MD, USA.
  • Zhu DC; Michigan State University, East Lansing, MI, USA.
  • Ota H; Tohoku University, Miyagi, Japan.
  • Sun J; University of Washington, Seattle, WA, USA.
  • Xu D; University of Washington, Seattle, WA, USA.
  • Kerwin WS; University of Washington, Seattle, WA, USA.
  • Hatsukami TS; University of Washington, Seattle, WA, USA.
  • Yuan C; University of Washington, Seattle, WA, USA. cyuan@uw.edu.
J Cardiovasc Magn Reson ; 18(1): 41, 2016 07 16.
Article en En | MEDLINE | ID: mdl-27430263
ABSTRACT

BACKGROUND:

Intraplaque hemorrhage (IPH) is associated with atherosclerosis progression and subsequent cardiovascular events. We sought to develop a semi-automatic method with an optimized threshold for carotid IPH detection and quantification on MP-RAGE images using matched histology as the gold standard.

METHODS:

Fourteen patients scheduled for carotid endarterectomy underwent 3D MP-RAGE cardiovascular magnetic resonance (CMR) preoperatively. Presence and area of IPH were recorded using histology. Presence and area of IPH were also recorded on CMR based on intensity thresholding using three references for intensity normalization the sternocleidomastoid muscle (SCM), the adjacent muscle and the automatically generated local median value. The optimized intensity thresholds were obtained by maximizing the Youden's index for IPH detection. Using leave-one-out cross validation, the sensitivity and specificity for IPH detection based on our proposed semi-automatic method and the agreement with histology on IPH area quantification were evaluated.

RESULTS:

The optimized intensity thresholds for IPH detection were 1.0 times the SCM intensity, 1.6 times the adjacent muscle intensity and 2.2 times the median intensity. Using the semi-automatic method with the optimized intensity threshold, the following IPH detection and quantification performance was obtained sensitivities up to 59, 68 and 80 %; specificities up to 85, 74 and 79 %; Pearson's correlation coefficients (IPH area measurement) up to 0.76, 0.93 and 0.90, respectively, using SCM, the adjacent muscle and the local median value for intensity normalization, after heavily calcified and small IPH were excluded.

CONCLUSIONS:

A semi-automatic method with good performance on IPH detection and quantification can be obtained in MP-RAGE CMR, using an optimized intensity threshold comparing to the adjacent muscle. The automatically generated reference of local median value provides comparable performance and may be particularly useful for developing automatic classifiers. Use of the SCM intensity as reference is not recommended without coil sensitivity correction when surface coils are used.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Interpretación de Imagen Asistida por Computador / Arterias Carótidas / Enfermedades de las Arterias Carótidas / Placa Aterosclerótica / Hemorragia Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Cardiovasc Magn Reson Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Año: 2016 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Interpretación de Imagen Asistida por Computador / Arterias Carótidas / Enfermedades de las Arterias Carótidas / Placa Aterosclerótica / Hemorragia Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Cardiovasc Magn Reson Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Año: 2016 Tipo del documento: Article País de afiliación: Estados Unidos