[Determination of trace cobalt in human urine by graphite furnace atomic absorption spectrometr].
Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi
; 34(5): 384-6, 2016 May 20.
Article
en Zh
| MEDLINE
| ID: mdl-27514430
ABSTRACT
OBJECTIVE:
To establish a method to determine cobalt in human urine by graphite furnace atomic absorption spectrometry.METHODS:
Urine with 2% nitric acid diluted two-fold, to quantify the curve, graphite furnace atomic absorption spectrometric detection.RESULTS:
Co was linear within 2.5~40.0 ng/ml with r>0.999. Spike experiment showed that Co received good recovery rate, which was 90.8%~94.8%. Intra-assay precisions were 3.2%~5.1% for Co, inter-assay precisions were 4.4%~5.2% for Co.CONCLUSION:
The method by using graphite furnace atomic absorption spectrometr to determine urine Co was fast, accurate and with low matrix effect. It could meet the requirement in GBZ/T 210.5-2008.