A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws.
Wichelhaus, Alice; Emmerich, Judith; Mittlmeier, Thomas.
Afiliación
  • Wichelhaus A; Department of Trauma, Hand and Reconstructive Surgery, University Faculty of Medicine Rostock, Schillingallee 35, 18057 Rostock, Germany.
  • Emmerich J; Department of Trauma, Hand and Reconstructive Surgery, University Faculty of Medicine Rostock, Schillingallee 35, 18057 Rostock, Germany.
  • Mittlmeier T; Department of Trauma, Hand and Reconstructive Surgery, University Faculty of Medicine Rostock, Schillingallee 35, 18057 Rostock, Germany.
Case Rep Orthop ; 2016: 7049130, 2016.
Article en En | MEDLINE | ID: mdl-27800201
ABSTRACT
This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws. During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at 6 weeks after implantation. The supposed osseous integration attributed to magnesium-based screws could not be reproduced in this particular case. Thus, it can be reasoned that the use of magnesium-based screws for partial wrist arthrodesis cannot be encouraged, at least not in dual use.

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Case Rep Orthop Año: 2016 Tipo del documento: Article País de afiliación: Alemania