Optimization and evaluation of Luminex performance with supernatants of antigen-stimulated peripheral blood mononuclear cells.
Surenaud, Mathieu; Manier, Céline; Richert, Laura; Thiébaut, Rodolphe; Levy, Yves; Hue, Sophie; Lacabaratz, Christine.
Afiliación
  • Surenaud M; INSERM, U955, Equipe 16, Créteil, F-94010, France.
  • Manier C; Université Paris Est, Faculté de médecine, Créteil, F-94010, France.
  • Richert L; Vaccine Research Institute (VRI), Créteil, F-94010, France.
  • Thiébaut R; INSERM, U955, Equipe 16, Créteil, F-94010, France.
  • Levy Y; Université Paris Est, Faculté de médecine, Créteil, F-94010, France.
  • Hue S; Vaccine Research Institute (VRI), Créteil, F-94010, France.
  • Lacabaratz C; Vaccine Research Institute (VRI), Créteil, F-94010, France.
BMC Immunol ; 17(1): 44, 2016 11 11.
Article en En | MEDLINE | ID: mdl-27835944
BACKGROUND: The Luminex bead-based multiplex assay is useful for quantifying immune mediators such as cytokines and chemokines. Cross-comparisons of reagents for this technique from different suppliers have already been performed using serum or plasma but rarely with supernatants collected from antigen-stimulated peripheral blood mononuclear cells (PBMC). Here, we first describe an optimization protocol for cell culture including quantity of cells and culture duration to obtain reproducible cytokine and chemokine quantifications. Then, we compared three different Luminex kit suppliers. RESULTS: Intraclass correlation coefficients (ICCs) for a 2-days stimulation protocol were >0.8 for IFNγ and Perforin. The specific concentration was maximal after two or five days of stimulation, depending on the analyte, using 0.5 million PBMC per well, a cell quantity that gave the same level of specific cytokine secretion as 1.0 million. In the second part of the study, Luminex kits from Millipore showed a better working range than Bio-Rad and Ozyme ones. For tuberculin purified protein derivative (PPD)-stimulated samples, the overall mean pooled coefficients of variation (CVs) for all donors and all cytokines was 17.2 % for Bio-Rad, 19.4 % for Millipore and 26.7 % for Ozyme. Although the different kits gave cytokine concentrations that were generally compatible, there were discrepancies for particular cytokines. Finally, evaluation of precision and reproducibility of a 15-plex Millipore kit using a "home-made" internal control showed a mean intra-assay CV <13 % and an inter-assay CV <18 % for each cytokine concentration. CONCLUSIONS: A protocol with a single round of stimulation but with two time points gave the best results for assaying different cytokines. Millipore kits appear to be slightly more sensitive than those from Bio-Rad and Ozyme. However, we conclude that the panel of analytes that need to be quantified should be the main determinant of kit selection. Using an internal control we demonstrated that a 15-plex magnetic Milliplex kit displayed good precision and reproducibility. Our findings should help optimize assays for evaluating immune responses during the course of disease or infection, or in response to vaccine or therapy.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Leucocitos Mononucleares / Separación Inmunomagnética / Microesferas Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: BMC Immunol Asunto de la revista: ALERGIA E IMUNOLOGIA Año: 2016 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Leucocitos Mononucleares / Separación Inmunomagnética / Microesferas Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: BMC Immunol Asunto de la revista: ALERGIA E IMUNOLOGIA Año: 2016 Tipo del documento: Article País de afiliación: Francia