Your browser doesn't support javascript.
loading
Premature Trigger of ERI in Medtronic EnRhythm Devices.
Middeldorp, Melissa E; Mahajan, Rajiv; Elliott, Adrian D; Pathak, Rajeev K; Twomey, Darragh; Wilson, Lauren; Stolcman, Simon; Munawar, Dian A; Kumar, Sharath; Lau, Dennis H; Sanders, Prashanthan.
Afiliación
  • Middeldorp ME; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Mahajan R; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Elliott AD; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Pathak RK; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Twomey D; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Wilson L; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Stolcman S; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Munawar DA; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Kumar S; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Lau DH; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Sanders P; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
Pacing Clin Electrophysiol ; 40(6): 624-628, 2017 Jun.
Article en En | MEDLINE | ID: mdl-28294359
ABSTRACT

BACKGROUND:

Medical technology has made significant advances over the last few decades with smaller and more dynamic pacemakers. However, technical failures leading to premature replacement is a cause of concern. We present a series of Medtronic EnRhythm devices that reached premature elective replacement indicator (ERI).

METHODS:

The database of Centre of Heart Rhythm Disorders was searched for EnRhythm device implantation from 2006 to 2011. Battery depletion <8.5 years was considered premature considering the projected average longevity to be 8.5-10.5 years. An unexpected premature ERI was defined when it was reached within 3 months of last normal check. Device follow-up was conducted every 3 months after advisory.

RESULTS:

A total of 88 EnRhythm pacemakers were implanted. Over a median follow-up of 6.2 years (range 0.3-9.2), 39 (44.3%) EnRhythm devices reached premature ERI. In 11 (28%), ERI was not recognized and patients were being investigated for other causes of unsteadiness or dyspnea prior to device check. Notably, three (7%) patients had premature ERI < 3.5 years. Ten (25.6%) had sudden and unexpected premature ERI. While asynchronous pacing was observed, there were no cases of absence of pacing.

CONCLUSIONS:

The rate of premature ERI for EnRhythm devices was 44.3%, significantly higher than reported by the manufacturer. Of concern, a sizeable proportion occurred unexpectedly, warranting more frequent reviews and empirical replacement in some patients. With the experience of the EnRhythm, appropriate monitoring strategies are recommended for future advisories.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Marcapaso Artificial / Suministros de Energía Eléctrica / Análisis de Falla de Equipo / Remoción de Dispositivos / Falla de Equipo / Seguridad de Equipos / Recall de Suministro Médico Límite: Aged / Female / Humans / Male País/Región como asunto: Oceania Idioma: En Revista: Pacing Clin Electrophysiol Año: 2017 Tipo del documento: Article País de afiliación: Australia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Marcapaso Artificial / Suministros de Energía Eléctrica / Análisis de Falla de Equipo / Remoción de Dispositivos / Falla de Equipo / Seguridad de Equipos / Recall de Suministro Médico Límite: Aged / Female / Humans / Male País/Región como asunto: Oceania Idioma: En Revista: Pacing Clin Electrophysiol Año: 2017 Tipo del documento: Article País de afiliación: Australia