Your browser doesn't support javascript.
loading
Adverse Local Tissue Responses to Failed Temporomandibular Joint Implants.
Mercuri, Louis G; Urban, Robert M; Hall, Deborah J; Mathew, Mathew T.
Afiliación
  • Mercuri LG; Visiting Professor, Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL; Clinical Consultant, TMJ Concepts, Ventura, CA. Electronic address: lgm@tmjconcepts.com.
  • Urban RM; Associate Professor, Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL.
  • Hall DJ; Assistant Professor, Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL.
  • Mathew MT; Associate Professor, Department of Biomedical Sciences, University of Illinois College of Medicine, Rockford, IL.
J Oral Maxillofac Surg ; 75(10): 2076-2084, 2017 Oct.
Article en En | MEDLINE | ID: mdl-28449848
ABSTRACT

PURPOSE:

The purpose of this study was to determine whether failed alloplastic temporomandibular joint replacement (TMJR) devices can elicit the aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) reaction seen in some patients with metal-on-metal hip arthroplasties. MATERIALS AND

METHODS:

This study involved analysis of paraffin-embedded sections of peri-implant tissue from failed TMJ implant cases obtained from 3 independent sources. Hematoxylin and eosin staining, conventional and polarized light microscopy, back-scattered electron imaging, and energy-dispersive x-ray analysis were used. Immunohistochemical methods were used to identify T and B lymphocytes and macrophages.

RESULTS:

The total TMJR device specimens showed primary macrophage and lymphocytic responses similar to responses reported previously for failed total hip implants, including ALVAL. No chronic or acute inflammation was apparent in the failed hemiarthroplasty TMJR cases.

CONCLUSION:

In this limited preliminary study, the local tissue responses to the failed TMJR implants showed similar primary macrophage and lymphocyte responses to previously reported failed metal-on-metal and metal-on-polyethylene orthopedic total joint replacement devices. No such local inflammatory responses were seen with the failed TMJR hemiarthroplasty devices.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Articulación Temporomandibular / Vasculitis / Falla de Prótesis / Prótesis Articulares / Prótesis Mandibular Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Oral Maxillofac Surg Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Articulación Temporomandibular / Vasculitis / Falla de Prótesis / Prótesis Articulares / Prótesis Mandibular Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: J Oral Maxillofac Surg Año: 2017 Tipo del documento: Article