Recovery from forward masking in cochlear implant listeners depends on stimulation mode, level, and electrode location.
J Acoust Soc Am
; 141(5): 3190, 2017 05.
Article
en En
| MEDLINE
| ID: mdl-28682084
Psychophysical recovery from forward masking was measured in adult cochlear implant users of CochlearTM and Advanced BionicsTM devices, in monopolar and in focused (bipolar and tripolar) stimulation modes, at four electrode sites across the arrays, and at two levels (loudness balanced across modes and electrodes). Results indicated a steeper psychophysical recovery from forward masking in monopolar over bipolar and tripolar modes, modified by differential effects of electrode and level. The interactions between factors varied somewhat across devices. It is speculated that psychophysical recovery from forward masking may be driven by different populations of neurons in the different modes, with a broader stimulation pattern resulting in a greater likelihood of response by healthier and/or faster-recovering neurons within the stimulated population. If a more rapid recovery from prior stimulation reflects responses of neurons not necessarily close to the activating site, the spectral pattern of the incoming acoustic signal may be distorted. These results have implications for speech processor implementations using different degrees of focusing of the electric field. The primary differences in the shape of the recovery function were observed in the earlier portion (between 2 and 45 ms) of recovery, which is significant in terms of the speech envelope.
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Enmascaramiento Perceptual
/
Percepción del Habla
/
Implantes Cocleares
/
Personas con Deficiencia Auditiva
/
Implantación Coclear
Límite:
Adolescent
/
Adult
/
Aged
/
Child
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
J Acoust Soc Am
Año:
2017
Tipo del documento:
Article
País de afiliación:
Estados Unidos