Fetal radiation dose in three common CT examinations during pregnancy - Monte Carlo study.
Kelaranta, A; Mäkelä, T; Kaasalainen, T; Kortesniemi, M.
Afiliación
  • Kelaranta A; HUS Medical Imaging Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, FI-00290 Helsinki, Finland. Electronic address: anna.kelaranta@alumni.helsinki.fi.
  • Mäkelä T; HUS Medical Imaging Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, FI-00290 Helsinki, Finland; Department of Physics, University of Helsinki, P.O. Box 64, Helsinki FI-00014, Finland.
  • Kaasalainen T; HUS Medical Imaging Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, FI-00290 Helsinki, Finland.
  • Kortesniemi M; HUS Medical Imaging Center, University of Helsinki and Helsinki University Hospital, P.O. Box 340, FI-00290 Helsinki, Finland; Department of Physics, University of Helsinki, P.O. Box 64, Helsinki FI-00014, Finland.
Phys Med ; 43: 199-206, 2017 Nov.
Article en En | MEDLINE | ID: mdl-28941740
ABSTRACT

PURPOSE:

To determine fetal doses in different stages of pregnancy in three common computed tomography (CT) examinations pulmonary CT angiography, abdomino-pelvic and trauma scan with Monte Carlo (MC) simulations.

METHODS:

An adult female anthropomorphic phantom was scanned with a 64-slice CT using pulmonary angiography, abdomino-pelvic and trauma CT scan protocols. Three different sized gelatin boluses placed on the phantom's abdomen simulated different stages of pregnancy. Intrauterine dose was used as a surrogate to a dose absorbed to the fetus. MC simulations were performed to estimate uterine doses. The simulation dose levels were calibrated with volumetric CT dose index (CTDIvol) measurements and MC simulations in a cylindrical CTDI body phantom and compared with ten point doses measured with metal-oxide-semiconductor field-effect-transistor dosimeters. Intrauterine volumes and uterine walls were segmented and the respective dose volume histograms were calculated.

RESULTS:

The mean intrauterine doses in different stages of pregnancy varied from 0.04 to 1.04mGy, from 4.8 to 5.8mGy, and from 9.8 to 12.6mGy in the CT scans for pulmonary angiography, abdomino-pelvic and trauma CT scans, respectively. MC simulations showed good correlation with the MOSFET measurement at the measured locations.

CONCLUSIONS:

The three studied examinations provided highly varying fetal doses increasing from sub-mGy level in pulmonary CT angiography to notably higher levels in abdomino-pelvic and trauma scans where the fetus is in the primary exposure range. Volumetric dose distribution offered by MC simulations in an appropriate anthropomorphic phantom provides a comprehensive dose assessment when applied in adjunct to point-dose measurements.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Dosis de Radiación / Tomografía Computarizada por Rayos X / Método de Montecarlo / Feto Tipo de estudio: Health_economic_evaluation Límite: Female / Humans / Pregnancy Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Dosis de Radiación / Tomografía Computarizada por Rayos X / Método de Montecarlo / Feto Tipo de estudio: Health_economic_evaluation Límite: Female / Humans / Pregnancy Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2017 Tipo del documento: Article