Compared vulnerabilities to small cardiac motions between different cameras used for myocardial perfusion imaging.
Salvadori, Julien; Petegnief, Yolande; Sabbah, Remi; Morel, Olivier; Boulahdour, Hatem; Karcher, Gilles; Marie, Pierre-Yves; Imbert, Laetitia.
Afiliación
  • Salvadori J; Institut de Cancérologie de Lorraine, Université de Lorraine, 54000, Nancy, France.
  • Petegnief Y; CHU-Besançon, Université de Franche-Comté, Service de Médecine Nucléaire, 25000, Besançon, France.
  • Sabbah R; CHU-Besançon, Université de Franche-Comté, Service de Médecine Nucléaire, 25000, Besançon, France.
  • Morel O; CHU-Besançon, Université de Franche-Comté, Service de Médecine Nucléaire, 25000, Besançon, France.
  • Boulahdour H; CHU-Besançon, Université de Franche-Comté, Service de Médecine Nucléaire, 25000, Besançon, France.
  • Karcher G; CHRU-Nancy, Université de Lorraine, Plateforme Nancyclotep, 54000, Nancy, France.
  • Marie PY; Médecine Nucléaire, Hôpital de Brabois, CHRU-Nancy, Université de Lorraine, Service de Médecine Nucléaire, 54000, Nancy, France.
  • Imbert L; CHRU-Nancy, Université de Lorraine, Plateforme Nancyclotep, 54000, Nancy, France.
J Nucl Cardiol ; 26(4): 1313-1322, 2019 08.
Article en En | MEDLINE | ID: mdl-29327252
ABSTRACT
This phantom-based study was aimed to determine whether cardiac CZT-cameras, which provide an enhanced spatial resolution and image contrast compared to Anger cameras, are similarly affected by small cardiac motions. Translations of a left ventricular (LV) insert at half-SPECT acquisitions through six possible orthogonal directions and with 5- or 10-mm amplitude were simulated on the Discovery NM-530c and DSPECT CZT-cameras and on an Anger Symbia T2 camera equipped with an astigmatic (IQ.SPECT) or conventional parallel-hole collimator (Conv.SPECT). SPECT images were initially reconstructed as currently recommended for clinical routine. The heterogeneity in recorded activity from the 17 LV segments gradually increased between baseline and motions simulated at 5- and 10-mm amplitudes with all cameras, although being higher for Anger- than CZT-cameras at each step and resulting in a higher mean number of artifactual abnormal segments (at 10-mm amplitude, Conv.SPECT 3.7; IQ.SPECT 1.8, Discovery 0.7, DSPECT 0). However, this vulnerability to motion was markedly (1) decreased for Conv.SPECT reconstructed without the recommended Resolution Recovery algorithm and (2) increased for DSPECT reconstructed without the recommended cardiac model. CZT-cameras and especially the DSPECT appear less vulnerable to small cardiac motions than Anger-cameras although these differences are strongly dependent on reconstruction parameters.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Imagen de Perfusión Miocárdica / Corazón / Movimiento (Física) Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Imagen de Perfusión Miocárdica / Corazón / Movimiento (Física) Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Francia