Your browser doesn't support javascript.
loading
Method for the Determination of Iodide in Dried Blood Spots from Newborns by High Performance Liquid Chromatography Tandem Mass Spectrometry.
Kim, Un-Jung; Kannan, Kurunthachalam.
Afiliación
  • Kim UJ; Wadsworth Center, New York State Department of Health, and Department of Environmental Health Sciences, School of Public Health , State University of New York at Albany , Empire State Plaza, P.O. Box 509 , Albany , New York 12201-0509 , United States.
  • Kannan K; Wadsworth Center, New York State Department of Health, and Department of Environmental Health Sciences, School of Public Health , State University of New York at Albany , Empire State Plaza, P.O. Box 509 , Albany , New York 12201-0509 , United States.
Anal Chem ; 90(5): 3291-3298, 2018 03 06.
Article en En | MEDLINE | ID: mdl-29412637
ABSTRACT
Dried blood spots (DBS), collected for newborn screening programs in the United States, have been used to screen for congenital metabolic diseases in newborns for over 50 years. DBS provide an easy and inexpensive way to collect and store peripheral blood specimens and present an excellent resource for studies on the assessment of chemical exposures in newborns. In this study, a selective and sensitive method was developed for the analysis of iodide in DBS by high performance liquid chromatography electrospray tandem mass spectrometry. Accuracy, inter- and intraday precision, matrix effects, and detection limits of the method were determined. Further validation of the method was accomplished by concurrent analysis of whole blood and fortified blood spotted on a Whatman 903 filter card. A significant positive correlation was found between measured concentrations of iodide in venous whole blood and the same blood spotted as DBS. The method limit of detection was 0.15 ng/mL iodide. The method was further validated by the analysis of a whole blood sample certified for iodide levels (proficiency testing sample) by spotting on a filter card. Twenty DBS samples collected from newborns in New York State were analyzed to demonstrate the applicability of the method. The measured concentrations of iodide in whole blood of newborns from New York State ranged between method is applicable for the analysis of DBS collected for epidemiological studies that investigate the importance of iodide on the health of newborns.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cromatografía Líquida de Alta Presión / Espectrometría de Masas en Tándem / Pruebas con Sangre Seca / Yoduros Límite: Humans / Newborn Idioma: En Revista: Anal Chem Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cromatografía Líquida de Alta Presión / Espectrometría de Masas en Tándem / Pruebas con Sangre Seca / Yoduros Límite: Humans / Newborn Idioma: En Revista: Anal Chem Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos