Your browser doesn't support javascript.
loading
A Pilot Study of Augmented Reality Technology Applied to the Acetabular Cup Placement During Total Hip Arthroplasty.
Ogawa, Hiroyuki; Hasegawa, Seiichirou; Tsukada, Sachiyuki; Matsubara, Masaaki.
Afiliación
  • Ogawa H; Department of Orthopaedic Surgery, Nissan Tamagawa Hospital, Tokyo, Japan; Department of Orthopaedic Surgery, The Fraternity Memorial Hospital, Tokyo, Japan.
  • Hasegawa S; Department of Orthopaedic Surgery, The Fraternity Memorial Hospital, Tokyo, Japan.
  • Tsukada S; Department of Orthopaedic Surgery, Hokusuikai Kinen Hospital, Ibaraki, Japan.
  • Matsubara M; Department of Orthopaedic Surgery, Nissan Tamagawa Hospital, Tokyo, Japan.
J Arthroplasty ; 33(6): 1833-1837, 2018 06.
Article en En | MEDLINE | ID: mdl-29502961
BACKGROUND: We developed an acetabular cup placement device, the AR-HIP system, using augmented reality (AR). The AR-HIP system allows the surgeon to view an acetabular cup image superimposed in the surgical field through a smartphone. The smartphone also shows the placement angle of the acetabular cup. This preliminary study was performed to assess the accuracy of the AR-HIP system for acetabular cup placement during total hip arthroplasty (THA). METHODS: We prospectively measured the placement angles using both a goniometer and AR-HIP system in 56 hips of 54 patients undergoing primary THA. We randomly determined the order of intraoperative measurement using the 2 devices. At 3 months after THA, the placement angle of the acetabular cup was measured on computed tomography images. The primary outcome was the absolute value of the difference between intraoperative and postoperative computed tomography measurements. RESULTS: The measurement angle using AR-HIP was significantly more accurate in terms of radiographic anteversion than that using a goniometer (2.7° vs 6.8°, respectively; mean difference 4.1°; 95% confidence interval, 3.0-5.2; P < .0001). There was no statistically significant difference in terms of radiographic inclination (2.1° vs 2.6°; mean difference 0.5°; 95% confidence interval, -1.1 to 0.1; P = .13). CONCLUSION: In this pilot study, the AR-HIP system provided more accurate information regarding acetabular cup placement angle than the conventional method. Further studies are required to confirm the utility of the AR-HIP system as a navigation tool.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Acetábulo Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Asunto de la revista: ORTOPEDIA Año: 2018 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Acetábulo Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Asunto de la revista: ORTOPEDIA Año: 2018 Tipo del documento: Article País de afiliación: Japón