The Clinical Performance of a New Chemiluminescent Immunoassay in Measuring Anti-ß2 Glycoprotein 1 and Anti-Cardiolipin Antibodies.
Zhou, Jiansuo; Hou, Xiuzhu; Zhang, Hua; Wang, Tiancheng; Cui, Liyan.
Afiliación
  • Zhou J; Department of Laboratory Medicine, Peking University Third Hospital, Beijing, China (mainland).
  • Hou X; Department of Laboratory Medicine, Peking University Third Hospital, Beijing, China (mainland).
  • Zhang H; Research Center of Clinical Epidemiology, Peking University Third Hospital, Beijing, China (mainland).
  • Wang T; Department of Laboratory Medicine, Peking University Third Hospital, Beijing, China (mainland).
  • Cui L; Department of Laboratory Medicine, Peking University Third Hospital, Beijing, China (mainland).
Med Sci Monit ; 24: 6816-6822, 2018 Sep 26.
Article en En | MEDLINE | ID: mdl-30256771
ABSTRACT
BACKGROUND Laboratory criterion is needed for the classification of antiphospholipid syndrome (APS), which contain anticardiolipin antibodies (aCL) and anti-ß2-glycoprotein 1 antibodies (aß2GP1). They are commonly identified by enzyme-linked immunosorbent assay (ELISA), but lack standardized kits, resulting in substantial variations in the antibody positivity between different laboratories. The emergence of chemiluminescence automated -BIO-FLASH may improve the situation. MATERIAL AND METHODS We selected 185 patients with APS, systemic lupus erythematosus (SLE), infertility, connective tissue disease (CTD), and other conditions in Peking University Third Hospital. We tested the aCL and aß2GP1 levels by EUROIMMUN ELISA and 105 patients had at least one positive result for aCL and aß2GP1, while the others had negative results. We retested them by chemiluminescence assay (CIA) and analyzed the result and compared the coincidence rate. The IgM levels were retested by AESKU ELISA. Data were analyzed using SPSS. RESULTS Our result suggested that CIA had good performance for IgG isotype of aCL and aß2GP1 in the coincidence rate. The positive coincidence rate of aCL IgM between CIA and EUROIMMUN ELISA was only 41.67%, but two ELISA kits showed good coincidence, CIA and AESKU ELISA had an obviously higher positive rate. CIA and AESKU had a higher coincidence than that of AESKU and EUROIMMUN in aß2GP1-IgM. CONCLUSIONS The new automated CIA BIO-FLASH is suitable for detecting aCL and aß2GP1 antibodies, especially IgG isotype, which may provide an alternative to time-consuming conventional ELISA method.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cardiolipinas / Beta 2 Glicoproteína I / Mediciones Luminiscentes Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Med Sci Monit Asunto de la revista: MEDICINA Año: 2018 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cardiolipinas / Beta 2 Glicoproteína I / Mediciones Luminiscentes Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Med Sci Monit Asunto de la revista: MEDICINA Año: 2018 Tipo del documento: Article