Length Redundancy and Twist Improve the Biomechanical Properties of Polytetrafluoroethylene Bypass Grafts.
Lee, Regent; Stoddart, Miranda; Dyson, Igor; Cassimjee, Ismail; Handa, Ashok; Cheng, Christopher P.
Afiliación
  • Lee R; Nuffield Department of Surgical Sciences, John Radcliffe Hospital, University of Oxford, Oxford, Oxfordshire, UK. Electronic address: regent.lee@nds.ox.ac.uk.
  • Stoddart M; Nuffield Department of Surgical Sciences, John Radcliffe Hospital, University of Oxford, Oxford, Oxfordshire, UK.
  • Dyson I; Department of Engineering Science, University of Oxford, Oxford, Oxfordshire, UK.
  • Cassimjee I; Nuffield Department of Surgical Sciences, John Radcliffe Hospital, University of Oxford, Oxford, Oxfordshire, UK.
  • Handa A; Nuffield Department of Surgical Sciences, John Radcliffe Hospital, University of Oxford, Oxford, Oxfordshire, UK.
  • Cheng CP; Nuffield Department of Surgical Sciences, John Radcliffe Hospital, University of Oxford, Oxford, Oxfordshire, UK; Department of Vascular Surgery, Stanford University, Palo Alto, CA. Electronic address: cpc@stanford.edu.
Ann Vasc Surg ; 61: 410-415, 2019 Nov.
Article en En | MEDLINE | ID: mdl-31319171
ABSTRACT

BACKGROUND:

The iliofemoropopliteal artery significantly changes path length during normal hip and knee flexion. Prosthetic bypass grafts, such as polytetrafluoroethylene (PTFE) grafts, are relatively stiff and thus can subject graft anastomoses to high tension when the path length increases. The aim of this study was to examine the influence of length redundancy and twist on the biomechanical properties of PTFE bypass grafts.

METHODS:

Unreinforced and ring-reinforced PTFE grafts were loaded in an axial mechanical testing machine to measure the tensile and compressive axial forces with varying levels of length redundancy and axial twist.

RESULTS:

Adding 5-15% length redundancy to a graft decreases the force to cause 5% extension by > 90% without substantially increasing shortening forces. Adding 4.5°/cm of axial twist imparts a corkscrew shape to the graft without increasing extension or shortening forces in the presence of length redundancy. Ring-reinforced PTFE grafts require more length redundancy to experience these reductions in forces especially in the presence of axial twist.

CONCLUSIONS:

A modest amount of length redundancy and twist (i.e., a cork-screw condition) confers improved biomechanical properties in a PTFE graft, especially in ring-reinforced grafts. This should be taken into consideration when fashioning an arterial bypass graft in the iliofemoropopliteal segment.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Politetrafluoroetileno / Diseño de Prótesis / Prótesis Vascular / Implantación de Prótesis Vascular / Extremidad Inferior / Hemodinámica Idioma: En Revista: Ann Vasc Surg Asunto de la revista: ANGIOLOGIA Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Politetrafluoroetileno / Diseño de Prótesis / Prótesis Vascular / Implantación de Prótesis Vascular / Extremidad Inferior / Hemodinámica Idioma: En Revista: Ann Vasc Surg Asunto de la revista: ANGIOLOGIA Año: 2019 Tipo del documento: Article