Initial clinical experience of N13-ammonia myocardial perfusion PET/CT using a compact superconducting production system.
Pieper, Justin; Patel, Vaiibhav N; Escolero, Sylvia; Nelson, Jacob R; Poitrasson-Rivière, Alexis; Shreves, Christopher K; Freiburger, Nick; Hubers, David; Rothley, Jill; Corbett, James R; Oliverio, Joseph; Ficaro, Edward P; Weinberg, Richard L; Murthy, Venkatesh L.
Afiliación
  • Pieper J; Department of Internal Medicine, University of Michigan, 1500 E Medical Center Drive, Ann Arbor, MI, 48109, USA. jpieper@med.umich.edu.
  • Patel VN; Division of Cardiovascular Medicine, Department of Medicine, University of Michigan, Ann Arbor, USA.
  • Escolero S; University of Michigan Medical School, Ann Arbor, MI, USA.
  • Nelson JR; University of Michigan Medical School, Ann Arbor, MI, USA.
  • Poitrasson-Rivière A; INVIA, Ann Arbor, MI, USA.
  • Shreves CK; Ionetix Corporation, San Francisco, CA, USA.
  • Freiburger N; Ionetix Corporation, San Francisco, CA, USA.
  • Hubers D; Division of Nuclear Medicine, Department of Radiology, University of Michigan, Ann Arbor, MI, USA.
  • Rothley J; Division of Nuclear Medicine, Department of Radiology, University of Michigan, Ann Arbor, MI, USA.
  • Corbett JR; Division of Cardiovascular Medicine, Department of Medicine, University of Michigan, Ann Arbor, USA.
  • Oliverio J; Division of Nuclear Medicine, Department of Radiology, University of Michigan, Ann Arbor, MI, USA.
  • Ficaro EP; Ionetix Corporation, San Francisco, CA, USA.
  • Weinberg RL; INVIA, Ann Arbor, MI, USA.
  • Murthy VL; Division of Cardiovascular Medicine, Department of Medicine, University of Michigan, Ann Arbor, USA.
J Nucl Cardiol ; 28(1): 295-299, 2021 02.
Article en En | MEDLINE | ID: mdl-31515758
ABSTRACT

BACKGROUND:

Although N13-ammonia has favorable properties among FDA approved radiotracers, complexity of implementation has limited its use. We describe the initial patient experience of N13-ammonia PET imaging using a compact N13-ammonia production system.

METHODS:

N13 was produced using the ION-12SC, a 12MeV, 10uA superconducting minimally shielded cyclotron, and reduced to N13-ammonia in an automated multi-use purification unit. Patients were power injected with 9.3 ± 1.1 mCi (344.1 ± 40.7 MBq) of N13-ammonia for rest imaging, and 18.8 ± 0.9 mCi (695.6 ± 33.3 MBq) of N13-ammonia was injected at peak hyperemia for stress testing. Images were interpreted for relative perfusion, left ventricular volumes/function, blood flow quantification, and scored for image quality.

RESULTS:

In total 97 patients underwent 98 N13-ammonia PET scans (32 rest only/65 rest-stress/1 stress only). Image quality was 91.8% good or excellent. None were poor/non-diagnostic. Study durations were acceptable. Tracer related radiation dosimetry to patients was 0.7 ± 0.1 mSv (rest only), and 2.1 ± 0.1 mSv (rest-stress).

CONCLUSION:

Clinical N13-ammonia production by the Ionetix ION-12SC delivers high quality myocardial PET perfusion images in a rapid protocol.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Imagen de Perfusión Miocárdica / Tomografía Computarizada por Tomografía de Emisión de Positrones / Amoníaco / Radioisótopos de Nitrógeno Tipo de estudio: Guideline Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Imagen de Perfusión Miocárdica / Tomografía Computarizada por Tomografía de Emisión de Positrones / Amoníaco / Radioisótopos de Nitrógeno Tipo de estudio: Guideline Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Estados Unidos