A dual-tuned 17 O/1 H head array for direct brain oximetry at 3 Tesla.
Lakshmanan, Karthik; Dehkharghani, Seena; Madelin, Guillaume; Brown, Ryan.
Afiliación
  • Lakshmanan K; Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, New York.
  • Dehkharghani S; Center for Advanced Imaging Innovation and Research (CAI2R), Department of Radiology, New York University School of Medicine, New York, New York.
  • Madelin G; Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, New York.
  • Brown R; Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, New York.
Magn Reson Med ; 83(4): 1512-1518, 2020 04.
Article en En | MEDLINE | ID: mdl-31593372
PURPOSE: To design and build a dual-tuned 17 O/1 H coil for direct brain oximetry at 3T. METHODS: A dual-tuned 17 O/1 H coil comprising 2 degenerate mode birdcage coils was constructed to facilitate high-sensitivity 17 O and 1 H imaging. In vivo 17 O brain images were acquired in a healthy volunteer using a fermat looped orthogonally encoded trajectories sequence, together with high-resolution structural brain 1 H images. RESULTS: Natural abundance 17 O images with a nominal resolution of 8 mm3 were acquired in under 20 minutes exhibiting clear delineation of the physiological 17 O distribution. One-millimeter isotropic 1 H structural brain images demonstrated excellent quality and anatomical detail using routine clinical imaging sequence parameters and parallel acceleration. CONCLUSION: A dual-tuned 17 O/1 H array was constructed to enable high-sensitivity 17 O and 1 H imaging under standard clinical 3 T scanning conditions.