Bioresorbable vascular scaffolds versus everolimus-eluting stents: a biomechanical analysis of the ABSORB III Imaging substudy.
EuroIntervention ; 16(12): e989-e996, 2020 12 18.
Article en En | MEDLINE | ID: mdl-32091401
ABSTRACT

AIMS:

The Absorb bioresorbable vascular scaffold (BVS) has high rates of target lesion failure (TLF) at three years. Low wall shear stress (WSS) promotes several mechanisms related to device TLF. We investigated the impact of BVS compared to XIENCE V (XV) on coronary WSS after device deployment. METHODS AND

RESULTS:

In the prospective, randomised, controlled ABSORB III Imaging study (BVS [n=77] or XV [n=36]), computational fluid dynamics were performed on fused angiographic and intravascular ultrasound (IVUS) images of post-implanted vessels. Low WSS was defined as <1 Pa. There were no differences in demographics, clinical risks, angiographic reference vessel diameter or IVUS minimal lumen diameter between BVS and XV patients. A greater proportion of vessels treated with BVS compared to XV demonstrated low WSS across the whole device (BVS 17/77 [22%] vs XV 2/36 [6%], p<0.029). Compared to XV, BVS demonstrated lower median circumferential WSS (1.73 vs 2.21 Pa; p=0.036), outer curvature WSS (p=0.026), and inner curvature WSS (p=0.038). Similarly, BVS had lower proximal third WSS (p=0.024), middle third WSS (p=0.047) and distal third WSS (p=0.028) when compared to XV. In a univariable logistic regression analysis, patients who received BVS were 4.8 times more likely to demonstrate low WSS across the scaffold/stent when compared to XV patients. Importantly, in a multivariable linear regression model, hypertension (beta 0.186, p=0.023), lower contrast frame count velocity (beta -0.411, p<0.001), lower post-stent residual plaque burden (beta -0.338, p<0.001), lower % underexpanded frames (beta -0.170, p=0.033) and BVS deployment (beta 0.251, p=0.002) remained independently associated with a greater percentage of stented coronary vessel areas exposed to low WSS.

CONCLUSIONS:

In this randomised controlled study, the Absorb BVS was 4.8 times more likely than the XV metallic stent to demonstrate low WSS. BVS implantation, lower blood velocity and lower residual post-stent plaque burden were independently associated with greater area of low WSS.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Stents Liberadores de Fármacos Tipo de estudio: Clinical_trials / Observational_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: EuroIntervention Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Stents Liberadores de Fármacos Tipo de estudio: Clinical_trials / Observational_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: EuroIntervention Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos