Accuracy and user performance evaluation of a blood glucose monitoring system which wirelessly transmits results to compatible insulin pumps.
Caswell, Michael; Brown, Daniel; Frank, Joy; Wallace, Jane F; Pardo, Scott.
Afiliación
  • Caswell M; Consumer Product Testing Company, Inc, Fairfield, NJ, USA.
  • Brown D; Ascensia Diabetes Care, Valhalla, NY, USA.
  • Frank J; Consumer Product Testing Company, Inc, Fairfield, NJ, USA.
  • Wallace JF; Ascensia Diabetes Care, Valhalla, NY, USA.
  • Pardo S; Ascensia Diabetes Care, Valhalla, NY, USA.
Curr Med Res Opin ; 36(5): 757-764, 2020 05.
Article en En | MEDLINE | ID: mdl-32096654
ABSTRACT

Objective:

To perform evaluations of the CONTOUR PLUS LINK 2.4 blood glucose monitoring system (BGMS) assessed according to ISO 151972013 criteria.

Methods:

Clinical trial registered at ClinicalTrials.gov (NCT01824355). In a laboratory study (Study 1), capillary fingertip blood samples from 100 subjects were evaluated in duplicate, using three test strip lots. In a clinical trial (Study 2), 113 adults with diabetes were enrolled, and BGMS results and Yellow Springs Instruments (YSI) analyzer reference measurements were compared for subject- and trial staff-obtained fingertip blood and subject-obtained palm blood. Subjects completed a questionnaire to evaluate BGMS ease of use.

Results:

In Study 1, 100% of combined results (all test strip lots) fulfilled ISO 151972013 section 6.3 criteria. In Study 2, 97.7% of subject-obtained fingertip results and 100% of trial staff-obtained fingertip results met ISO 151972013 section 8 criteria. Additionally, 93.8% of palm results were within ± 15 mg/dL of mean YSI measurements for glucose concentrations <100 mg/dL or ± 15% for glucose concentrations ≥100 mg/dL. Most subjects found the BGMS easy to use. There were three non-serious, non-device related adverse events.

Conclusion:

The BGMS exceeded minimum ISO 151972013-specified accuracy criteria in the laboratory and in the hands of lay users with diabetes.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Sistemas de Infusión de Insulina / Automonitorización de la Glucosa Sanguínea / Diabetes Mellitus Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Curr Med Res Opin Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Sistemas de Infusión de Insulina / Automonitorización de la Glucosa Sanguínea / Diabetes Mellitus Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Curr Med Res Opin Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos