Cavitation phenomenon in mechanical prosthetic valves: Not only microbubbles.
Vriz, Olga; Arshi, Fatma; Ahmed, Mawada; Alhumaid, Mohammed; Galzerano, Domenico; Emmanual, Shisamma; Kinsara, Abdulhalim J; Aladmawi, Mohammed; Alamro, Bandar; Alshahid, Maie; Pedrizzettid, Gianni.
Afiliación
  • Vriz O; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
  • Arshi F; College of Medicine, Alfaisal University, Riyadh, Saudi Arabia.
  • Ahmed M; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
  • Alhumaid M; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
  • Galzerano D; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
  • Emmanual S; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
  • Kinsara AJ; College of Medicine, Alfaisal University, Riyadh, Saudi Arabia.
  • Aladmawi M; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
  • Alamro B; Ministry of National Guard Health Affair, COM-WR, King Abdullah International Research Center, King Saud Bin Abdulaziz University for Health Sciences, Jeddah, Saudi Arabia.
  • Alshahid M; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
  • Pedrizzettid G; Cardiac Centre, King Faisal Specialist Hospital and Research Center, Riyadh, Saudi Arabia.
Echocardiography ; 37(6): 876-882, 2020 06.
Article en En | MEDLINE | ID: mdl-32416011
ABSTRACT

INTRODUCTION:

Microbubbles (MBs) or cavitation is high-velocity, echo-bright findings present during the closing or opening of a mechanical valve (MVP). Cavitation bubble growth or gas emboli are less frequently described. We evaluated the hemodynamic parameters involved in the formation of gas emboli and the impact of gas emboli on requests for additional investigations. METHODS AND

RESULTS:

Transthoracic echocardiographic studies (TTE) of 57 patients (31 males, mean age 46.8 ± 13.8 years) with gas emboli were evaluated after heart valve replacement surgery. The majority (72%, n = 42) had a mitral or combined mitral/aortic MVP, with 28% (n = 16) an aortic MVP. The last TTE with and without gas emboli were considered for the same patient and the no emboli group was the control group (42 patients). The patient's blood pressure (BP) and heart rate (HR) were available for each TTE. Comparing the two TTEs, the systolic and diastolic BP, transmitral and aortic gradients, and left ventricular ejection fraction were similar but the HR (80.9 ± 18.7 vs 72.5 ± 13.9 bpm, P = .02) was significantly higher in the group with gas emboli. A TEE was performed 52 times in 27 patients, due to gas emboli, with one case positive for thrombus/vegetation. For 19 patients, a brain CT was requested. In two patients, the indication for the brain CT was gas emboli but the result was negative.

CONCLUSION:

Gas emboli are frequently present and associated to an increased HR. They can cause the misdiagnosis of endocarditis or thrombus formation with significant additional requests for diagnostic examinations.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Prótesis Valvulares Cardíacas / Microburbujas Límite: Adult / Humans / Male / Middle aged Idioma: En Revista: Echocardiography Asunto de la revista: CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Año: 2020 Tipo del documento: Article País de afiliación: Arabia Saudita

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Prótesis Valvulares Cardíacas / Microburbujas Límite: Adult / Humans / Male / Middle aged Idioma: En Revista: Echocardiography Asunto de la revista: CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Año: 2020 Tipo del documento: Article País de afiliación: Arabia Saudita