Gunshot damage to monoplace hyperbaric chamber acrylic.
Weaver, Lindell K; Deru, Kayla; Foley, John.
Afiliación
  • Weaver LK; Division of Hyperbaric Medicine Intermountain Medical Center, Murray, Utah U.S.
  • Deru K; Intermountain LDS Hospital, Salt Lake City, Utah.
  • Foley J; Department of Medicine, University of Utah School of Medicine, Salt Lake City, Utah U.S.
Undersea Hyperb Med ; 47(2): 229-234, 2020.
Article en En | MEDLINE | ID: mdl-32574439
Introduction: Health care workers are vulnerable to workplace violence, including active shooter incidents. Little is known about how firearms could damage monoplace chamber acrylic and whether a breached pressurized chamber presents additional threat to the patient or bystanders. Methods: In a remote area where firearm discharge is permitted, we tested the durability of sections of monoplace hyperbaric chamber acrylic under various firearm discharges. Firearms were discharged at acrylic sections from a distance of 17 feet at 45 degrees and 10 degrees from perpendicular while wearing protective gear. Firearm calibers ranged from .22 caliber handgun to 5.56 mm AR-15 rifle. We also conducted similar testing on a monoplace hyperbaric chamber pressurized with >99% oxygen to a differential pressure of 14.7 psig (2.0 atmospheres absolute at sea level). Handguns were remotely fired at a distance of 12 feet from the chamber (30 degrees from perpendicular), while the rifles were fired at a distance of 60 feet from the chamber. Result: Higher-caliber handguns penetrated or fractured the acrylic sections only after multiple shots. The tested rifles caused full-thickness penetration and fracture with a single shot. However, the pressurized monoplace hyperbaric chamber required two shots from the AR-15 rifle, separated by approximately 60 mm, to penetrate the acrylic, resulting in rapid depressurization. The chamber otherwise remained intact, with no explosion or conflagration observed. Conclusion: An intact or pressurized chamber performs differently than stand-alone acrylic sections under firearms testing. In a worst-case active shooter scenario, the pressurized monoplace chamber tested posed no additional threat to bystanders beyond the significant risk of ricochet.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Armas de Fuego / Análisis de Falla de Equipo / Oxigenoterapia Hiperbárica Idioma: En Revista: Undersea Hyperb Med Asunto de la revista: FISIOLOGIA Año: 2020 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Armas de Fuego / Análisis de Falla de Equipo / Oxigenoterapia Hiperbárica Idioma: En Revista: Undersea Hyperb Med Asunto de la revista: FISIOLOGIA Año: 2020 Tipo del documento: Article