Mechanical wear analysis helps understand a mechanism of failure in retrieved magnetically controlled growing rods: a retrieval study.
Wei, Jack Z; Hothi, Harry S; Morganti, Holly; Bergiers, Sean; Dal Gal, Elisabetta; Likcani, Doris; Henckel, Johann; Hart, Alister J.
Afiliación
  • Wei JZ; Institute of Orthopaedics and Musculoskeletal Science, University College London, Brockley Hill, Stanmore, HA7 4LP, UK. zijian.wei.18@ucl.ac.uk.
  • Hothi HS; Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, HA7 4LP, UK.
  • Morganti H; Institute of Orthopaedics and Musculoskeletal Science, University College London, Brockley Hill, Stanmore, HA7 4LP, UK.
  • Bergiers S; Institute of Orthopaedics and Musculoskeletal Science, University College London, Brockley Hill, Stanmore, HA7 4LP, UK.
  • Dal Gal E; Institute of Orthopaedics and Musculoskeletal Science, University College London, Brockley Hill, Stanmore, HA7 4LP, UK.
  • Likcani D; Institute of Orthopaedics and Musculoskeletal Science, University College London, Brockley Hill, Stanmore, HA7 4LP, UK.
  • Henckel J; Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, HA7 4LP, UK.
  • Hart AJ; Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, HA7 4LP, UK.
BMC Musculoskelet Disord ; 21(1): 519, 2020 Aug 05.
Article en En | MEDLINE | ID: mdl-32758204
BACKGROUND: To assess the relationship between mechanical wear and the failure of the internal lengthening mechanism in retrieved MAGnetic Expansion Control (MAGEC) growing rods. METHODS: This study included 34 MAGEC rods retrieved from 20 patients. The state of the internal mechanism and mechanical wear were assessed in all the rods using plain radiographs and visual inspection. Metrology was then performed to assess the topography and mechanical wear of the telescopic bars, using a Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine. RESULTS: Plain radiographs showed evidence of a broken internal mechanism in 29% of retrieved rods. Single-side wear marks were found in 97% of retrieved rods. Material loss was found to significantly increase in rods with a damaged internal mechanism (p < 0.05) and rods with longer time in situ (r = 0.692, p < 0.05). CONCLUSIONS: We found an association between damage to the internal mechanism of the rods and (1) patterns of single-side longitudinal wear marks and (2) increased material loss. As the material loss was also found to increase over time of rod in situ, we emphasise the importance of early detection and revision of failed MAGEC rods in clinical practice.