Repeat patient testing-based quality control shows promise for use in veterinary biochemistry testing.
Flatland, Bente; Freeman, Kathleen P.
Afiliación
  • Flatland B; Department of Biomedical and Diagnostic Sciences, College of Veterinary Medicine A205, University of Tennessee, Knoxville, TN, USA.
  • Freeman KP; Department of Biomedical and Diagnostic Sciences, College of Veterinary Medicine A205, University of Tennessee, Knoxville, TN, USA.
Vet Clin Pathol ; 49(4): 590-606, 2020 Dec.
Article en En | MEDLINE | ID: mdl-33314202
ABSTRACT

BACKGROUND:

Repeat patient testing-based quality control (RPT-QC) is a form of statistical QC and an alternative to commercial quality control materials (QCM).

OBJECTIVE:

This study investigated the suitability of canine heparinized plasma for use in RPT-QC and assessed the predicted performance of RPT-QC for the detection of analytical error in chemistry testing.

METHODS:

The stability of canine plasma for RPT-QC was investigated via storage at two temperatures for three or six time points. Storage data were analyzed using repeated measures ANOVA and by comparing results for stored specimens to baseline data using predetermined criteria. To generate RPT-QC limit-setting and -validation data, leftover plasma was prospectively measured. Once control limits were established, these were challenged by measuring specimens for which the repeat aliquot had been manipulated to mimic analytical error. Finally, the predicted performance of RPT-QC and QCM-QC with four control rules was investigated using Westgard's EZ Rules 3.

RESULTS:

Refrigerated storage of canine plasma for 7 days allowed mild changes facilitating RPT-QC. RPT-QC limits for 12 of 17 common measurands were validated. Validated limits successfully flagged differences from manipulated specimen pairs as "error." The predicted performance of RPT-QC for analytical error detection (represented by smallest achievable allowable total error, given a probability of error detection ≥ 85% and a probability of false rejection ≤ 5%) for four common control rules is comparable to that of QCM-QC.

CONCLUSIONS:

This study provides evidence that RPT-QC using canine heparinized plasma refrigerated for 7 days can be used with simple control rules and low numbers of control materials, suggesting RPT-QC is applicable to both reference and in-clinic laboratory settings.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Servicios de Laboratorio Clínico / Laboratorios Tipo de estudio: Prognostic_studies Límite: Animals Idioma: En Revista: Vet Clin Pathol Asunto de la revista: MEDICINA VETERINARIA / PATOLOGIA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Servicios de Laboratorio Clínico / Laboratorios Tipo de estudio: Prognostic_studies Límite: Animals Idioma: En Revista: Vet Clin Pathol Asunto de la revista: MEDICINA VETERINARIA / PATOLOGIA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos