Evaluation of the Onclarity HPV assay on the high-throughput COR system.
Expert Rev Mol Diagn
; 21(3): 333-342, 2021 03.
Article
en En
| MEDLINE
| ID: mdl-33849372
Background: Here we compare the performance of the high-throughput BD COR System (COR) to the Viper LT System (Viper) using the BD Onclarity HPV assay.Research Design and Methods: Remnant clinical specimens, contrived specimens in SurePath (BD) and PreservCyt (Hologic) media, and prospective clinical specimens in BD Cervical Brush Diluent (CBD) were tested. Outcomes included intra-laboratory agreement of Onclarity results on COR and inter-system agreement between COR and Viper.Results: Onclarity reproducibility on COR resulted in standard deviation and correlation of variation of Ct values ranging from 0.14 to 1.98 and 0.49% to 2.15%, respectively, for contrived specimens, and 0.9-3.08 and 2.89-9.21%, respectively, for clinical specimens. In the COR and Viper clinical agreement study, OPA for Onclarity ranged from 97.1%-98.9%, depending on the collection media type. PPA values for pooled, HPV(+) specimens at low positive (C95), and moderate positive (3XC95) target concentrations were ≥95.0% and 100%, respectively; PPA values associated with HPV 16, 18, 31, 45, 33/58, 52, 35/39/68, 51, and 56/59/66, individually, ranged from 93.8%-100%.Conclusions: Onclarity performance on COR is equivalent to Viper, and is accurate and reproducible for detection of all high-risk HPV genotypes, with a throughput of 330 results from a single 8-hour shift.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Neoplasias del Cuello Uterino
/
Infecciones por Papillomavirus
Tipo de estudio:
Diagnostic_studies
/
Observational_studies
/
Risk_factors_studies
Límite:
Female
/
Humans
Idioma:
En
Revista:
Expert Rev Mol Diagn
Asunto de la revista:
BIOLOGIA MOLECULAR
Año:
2021
Tipo del documento:
Article
País de afiliación:
Estados Unidos