A fast MR-thermometry method for quantitative assessment of temperature increase near an implanted wire.
Delcey, Marylène; Bour, Pierre; Ozenne, Valéry; Ben Hassen, Wadie; Quesson, Bruno.
Afiliación
  • Delcey M; IHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, Pessac-Bordeaux, France.
  • Bour P; Centre de recherche Cardio-Thoracique de Bordeaux, U1045, Univ. Bordeaux, Bordeaux, France.
  • Ozenne V; INSERM, Centre de recherche Cardio-Thoracique de Bordeaux, U1045, Bordeaux, France.
  • Ben Hassen W; Siemens Healthcare, Saint-Denis, France.
  • Quesson B; IHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, Pessac-Bordeaux, France.
PLoS One ; 16(5): e0250636, 2021.
Article en En | MEDLINE | ID: mdl-33983935
ABSTRACT

PURPOSE:

To propose a MR-thermometry method and associated data processing technique to predict the maximal RF-induced temperature increase near an implanted wire for any other MRI sequence.

METHODS:

A dynamic single shot echo planar imaging sequence was implemented that interleaves acquisition of several slices every second and an energy deposition module with adjustable parameters. Temperature images were processed in real time and compared to invasive fiber-optic measurements to assess accuracy of the method. The standard deviation of temperature was measured in gel and in vivo in the human brain of a volunteer. Temperature increases were measured for different RF exposure levels in a phantom containing an inserted wire and then a MR-conditional pacemaker lead. These calibration data set were fitted to a semi-empirical model allowing estimation of temperature increase of other acquisition sequences.

RESULTS:

The precision of the measurement obtained after filtering with a 1.6x1.6 mm2 in plane resolution was 0.2°C in gel, as well as in the human brain. A high correspondence was observed with invasive temperature measurements during RF-induced heating (0.5°C RMSE for a 11.5°C temperature increase). Temperature rises of 32.4°C and 6.5°C were reached at the tip of a wire and of a pacemaker lead, respectively. After successful fitting of temperature curves of the calibration data set, temperature rise predicted by the model was in good agreement (around 5% difference) with measured temperature by a fiber optic probe, for three other MRI sequences.

CONCLUSION:

This method proposes a rapid and reliable quantification of the temperature rise near an implanted wire. Calibration data set and resulting fitting coefficients can be used to estimate temperature increase for any MRI sequence as function of its power and duration.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Prótesis e Implantes / Temperatura / Imagen por Resonancia Magnética / Termometría Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2021 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Prótesis e Implantes / Temperatura / Imagen por Resonancia Magnética / Termometría Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2021 Tipo del documento: Article País de afiliación: Francia