Improving mandibular reconstruction by using topology optimization, patient specific design and additive manufacturing?-A biomechanical comparison against miniplates on human specimen.
Lang, Jan J; Bastian, Mirjam; Foehr, Peter; Seebach, Michael; Weitz, Jochen; von Deimling, Constantin; Schwaiger, Benedikt J; Micheler, Carina M; Wilhelm, Nikolas J; Grosse, Christian U; Kesting, Marco; Burgkart, Rainer.
Afiliación
  • Lang JJ; Department of Orthopedics and Sports Orthopedics, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.
  • Bastian M; Department of Mechanical Engineering, Chair of Non-destructive Testing, Technical University of Munich, Munich, Germany.
  • Foehr P; Department of Orthopedics and Sports Orthopedics, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.
  • Seebach M; Department of Orthopedics and Sports Orthopedics, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.
  • Weitz J; Department of Mechanical Engineering, Chair of Non-destructive Testing, Technical University of Munich, Munich, Germany.
  • von Deimling C; Department of Mechanical Engineering, Institute for Machine Tools and Industrial Management, Technical University of Munich, Munich, Germany.
  • Schwaiger BJ; Department of Oral and Maxillofacial Surgery, Josefinum, and Private Practice for Oral and Maxillofacial Surgery at Pferseepark, Augsburg, Germany.
  • Micheler CM; Department of Orthopedics and Sports Orthopedics, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.
  • Wilhelm NJ; Department of Mechanical Engineering, Chair of Applied Mechanics, Technical University of Munich, Munich, Germany.
  • Grosse CU; Department of Diagnostic and Interventional Neuroradiology, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.
  • Kesting M; Department of Orthopedics and Sports Orthopedics, Klinikum rechts der Isar, School of Medicine, Technical University of Munich, Munich, Germany.
  • Burgkart R; Department of Mechanical Engineering, Institute for Machine Tools and Industrial Management, Technical University of Munich, Munich, Germany.
PLoS One ; 16(6): e0253002, 2021.
Article en En | MEDLINE | ID: mdl-34101755
In this study, topology optimized, patient specific osteosynthesis plates (TOPOS-implants) are evaluated for the mandibular reconstruction using fibula segments. These shape optimized implants are compared to a standard treatment with miniplates (thickness: 1.0 mm, titanium grade 4) in biomechanical testing using human cadaveric specimen. Mandible and fibula of 21 body donors were used. Geometrical models were created based on automated segmentation of CT-scans of all specimens. All reconstructions, including cutting guides for osteotomy as well as TOPOS-implants, were planned using a custom-made software tool. The TOPOS-implants were produced by electron beam melting (thickness: 1.0 mm, titanium grade 5). The fibula-reconstructed mandibles were tested in static and dynamic testing in a multi-axial test system, which can adapt to the donor anatomy and apply side-specific loads. Static testing was used to confirm mechanical similarity between the reconstruction groups. Force-controlled dynamic testing was performed with a sinusoidal loading between 60 and 240 N (reconstructed side: 30% reduction to consider resected muscles) at 5 Hz for up to 5 · 105 cycles. There was a significant difference between the groups for dynamic testing: All TOPOS-implants stayed intact during all cycles, while miniplate failure occurred after 26.4% of the planned loading (1.32 · 105 ± 1.46 · 105 cycles). Bone fracture occurred in both groups (miniplates: n = 3, TOPOS-implants: n = 2). A correlation between bone failure and cortical bone thickness in mandible angle as well as the number of bicortical screws used was demonstrated. For both groups no screw failure was detected. In conclusion, the topology optimized, patient specific implants showed superior fatigue properties compared to miniplates in mandibular reconstruction. Additionally, the patient specific shape comes with intrinsic guiding properties to support the reconstruction process during surgery. This demonstrates that the combination of additive manufacturing and topology optimization can be beneficial for future maxillofacial surgery.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estrés Mecánico / Placas Óseas / Diseño de Equipo / Reconstrucción Mandibular / Fracturas Mandibulares Tipo de estudio: Prognostic_studies Límite: Aged / Female / Humans / Male Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2021 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Estrés Mecánico / Placas Óseas / Diseño de Equipo / Reconstrucción Mandibular / Fracturas Mandibulares Tipo de estudio: Prognostic_studies Límite: Aged / Female / Humans / Male Idioma: En Revista: PLoS One Asunto de la revista: CIENCIA / MEDICINA Año: 2021 Tipo del documento: Article País de afiliación: Alemania