Differences in human and minipig platelet number, volume and activation induced by borosilicate glass beads in a modified chandler loop-system.
Clin Hemorheol Microcirc
; 79(1): 149-155, 2021.
Article
en En
| MEDLINE
| ID: mdl-34487038
The pig is the most widely used large animal model in Europe, with cardiovascular research being one of the main areas of application. Adequate refinement of interventional studies in this field, meeting the requirements of Russel and Burchs' 3âR concept, can only be performed if blood-contacting medical devices are hemocompatible. Because most medical devices for cardiovascular interventional procedures are developed for humans they are tested mostly for compatibility with human blood. The aim of this study was therefore to determine whether there are differences in behavior of porcine and human platelets when they come into contact with glass, which was used as an exemplary thrombogenic material. For this purpose changes of platelet count, platelet volume and platelet expression of the activation markers CD61, CD62P and CD63 were measured using a modified chandler loop-system simulating the fluidic effects of the blood flow. Minipig and human platelets showed significant differences in number and volume, but not in activation after 4-8âmin exposure to glass.