Stable Pb(II) ion-selective electrodes with a low detection limit using silver nanoparticles/polyaniline as the solid contact.
Mikrochim Acta
; 188(11): 393, 2021 Oct 26.
Article
en En
| MEDLINE
| ID: mdl-34698939
ABSTRACT
Solid contact-based ion-selective electrodes (SC-ISEs) based on silver nanoparticles/polyaniline (Ag@PANI) as the solid contact (SC) were successfully prepared. The Ag@PANI SC showed high capacitance and excellent electron transport performance. Owing to the synergetic effects of the Ag nanoparticles and PANI, a GC/Ag@PANI-II/Pb2+-ISE (where II refers to a Ag content of 0.01 wt% in the SC layer) showed a low Pb2+ detection limit (6.31 × 10-10 M) with a slope of 29.1 ± 0.3 mV/dec, a fast response (< 5 s), and high stability. GC/Ag@PANI-II/Pb2+-ISE exhibited a Nernstian response for Pb2+ ions over a wide concentration range (10-3 to 10-9 M). After a 3-week operation, GC/Ag@PANI-II/Pb2+-ISE responded linearly to Pb2+ over the range of 10-7-10-3 M, demonstrating good long-term potential stability. Furthermore, the electrode showed excellent reproducibility and repeatability of the potential values and was successfully applied to detect the Pb2+ concentration in real samples with a recovery of 97 - 109%. Results suggest that Ag@PANI composites offer good transducer performance in trace ion detection sensors.
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Tipo de estudio:
Diagnostic_studies
Idioma:
En
Revista:
Mikrochim Acta
Año:
2021
Tipo del documento:
Article
País de afiliación:
China