Your browser doesn't support javascript.
loading
Hemodynamic characterization of the Realheart® total artificial heart with a hybrid cardiovascular simulator.
Fresiello, Libera; Najar, Azad; Brynedal Ignell, Nils; Zielinski, Krzysztof; Rocchi, Maria; Meyns, Bart; Perkins, Ina Laura.
Afiliación
  • Fresiello L; Department of Cardiovascular Sciences, Cardiac Surgery, Katholieke Universiteit Leuven, Leuven, Belgium.
  • Najar A; Institute of Clinical Physiology, National Research Council, Pisa, Italy.
  • Brynedal Ignell N; R&D, Scandinavian Real Heart AB, Västerås, Sweden.
  • Zielinski K; R&D, Scandinavian Real Heart AB, Västerås, Sweden.
  • Rocchi M; Nalecz Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, Warsaw, Poland.
  • Meyns B; Department of Cardiovascular Sciences, Cardiac Surgery, Katholieke Universiteit Leuven, Leuven, Belgium.
  • Perkins IL; Department of Cardiovascular Sciences, Cardiac Surgery, Katholieke Universiteit Leuven, Leuven, Belgium.
Artif Organs ; 46(8): 1585-1596, 2022 Aug.
Article en En | MEDLINE | ID: mdl-35231138
ABSTRACT

BACKGROUND:

Heart failure is a growing health problem worldwide. Due to the lack of donor hearts there is a need for alternative therapies, such as total artificial hearts (TAHs). The aim of this study is to evaluate the hemodynamic performance of the Realheart® TAH, a new 4-chamber cardiac prosthesis device.

METHODS:

The Realheart® TAH was connected to a hybrid cardiovascular simulator with inflow connections at the left/right atrium, and outflow connections at the ascending aorta/pulmonary artery. The Realheart® TAH was tested at different pumping rates and stroke volumes. Different systemic resistances (20.0-16.7-13.3-10.0 Wood units), pulmonary resistances (6.7-3.3-1.7 Wood units), and pulmonary/systemic arterial compliances (1.4-0.6 ml/mm Hg) were simulated. Tests were also conducted in static conditions, by imposing predefined values of preload-afterload across the artificial ventricle.

RESULTS:

The Realheart® TAH allows the operator to finely tune the delivered flow by regulating the pumping rate and stroke volume of the artificial ventricles. For a systemic resistance of 16.7 Wood units, the TAH flow ranges from 2.7 ± 0.1 to 6.9 ± 0.1 L/min. For a pulmonary resistance of 3.3 Wood units, the TAH flow ranges from 3.1 ± 0.0 to 8.2 ± 0.3 L/min. The Realheart® TAH delivered a pulse pressure ranging between ~25 mm Hg and ~50 mm Hg for the tested conditions.

CONCLUSIONS:

The Realheart® TAH offers great flexibility to adjust the output flow and delivers good pressure pulsatility in the vessels. Low sensitivity of device flow to the pressure drop across it was identified and a new version is under development to counteract this.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Trasplante de Corazón / Corazón Artificial Límite: Humans Idioma: En Revista: Artif Organs Año: 2022 Tipo del documento: Article País de afiliación: Bélgica

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Trasplante de Corazón / Corazón Artificial Límite: Humans Idioma: En Revista: Artif Organs Año: 2022 Tipo del documento: Article País de afiliación: Bélgica