Your browser doesn't support javascript.
loading
Model-based image reconstruction with wavelet sparsity regularization for through-plane resolution restoration in T2 -weighted spin-echo prostate MRI.
Borisch, Eric A; Froemming, Adam T; Grimm, Roger C; Kawashima, Akira; Trzasko, Joshua D; Riederer, Stephen J.
Afiliación
  • Borisch EA; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Froemming AT; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Grimm RC; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Kawashima A; Department of Radiology, Mayo Clinic, Phoenix, Arizona, USA.
  • Trzasko JD; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Riederer SJ; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Magn Reson Med ; 89(1): 454-468, 2023 Jan.
Article en En | MEDLINE | ID: mdl-36093998
ABSTRACT

PURPOSE:

The purpose is to develop a model-based image-reconstruction method using wavelet sparsity regularization for maintaining restoration of through-plane resolution but with improved retention of SNR versus linear reconstruction using Tikhonov (TK) regularization in high through-plane resolution (1 mm) T2 -weighted spin-echo (T2SE) images of the prostate.

METHODS:

A wavelet sparsity (WS)-regularized image reconstruction was developed that takes as input a set of ≈80 overlapped 3-mm-thick slices acquired using a T2SE multislice scan and typically 30 coil elements. After testing in contrast and resolution phantoms and calibration in 6 subjects, the WS reconstruction was evaluated in 16 consecutive prostate T2SE MRI exams. Results reconstructed with nominal 1-mm thickness were compared with those from the TK reconstruction with the same raw data. Results were evaluated radiologically. The ratio of magnitude of prostate signal to periprostatic muscle signal was used to assess the presence of noise reduction. Technical performance was also compared with a commercial 3D-T2SE sequence.

RESULTS:

The new WS reconstruction was assessed as superior statistically to TK for overall SNR, contrast, and multiple evaluation criteria related to sharpness while retaining the high (1 mm) through-plane resolution. Wavelet sparsity tended to provide improved overall diagnostic quality versus TK, but not significantly so. In all 16 studies, the prostate-to-muscle signal ratio increased.

CONCLUSIONS:

Model-based WS-regularized reconstruction consistently provides improved SNR in high (1 mm) through-plane resolution images of prostate T2SE MRI versus linear reconstruction using TK regularization.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Próstata / Imagen por Resonancia Magnética Tipo de estudio: Prognostic_studies Límite: Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Próstata / Imagen por Resonancia Magnética Tipo de estudio: Prognostic_studies Límite: Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos