Your browser doesn't support javascript.
loading
Automated field-in-field whole brain radiotherapy planning.
Huang, Kai; Hernandez, Soleil; Wang, Chenyang; Nguyen, Callistus; Briere, Tina Marie; Cardenas, Carlos; Court, Laurence; Xiao, Yao.
Afiliación
  • Huang K; The University of Texas MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, USA.
  • Hernandez S; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Wang C; The University of Texas MD Anderson Cancer Center UTHealth Graduate School of Biomedical Sciences, Houston, Texas, USA.
  • Nguyen C; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Briere TM; Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Cardenas C; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Court L; Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Xiao Y; Department of Radiation Oncology, The University of Alabama at Birmingham, Birmingham, Alabama, USA.
J Appl Clin Med Phys ; 24(2): e13819, 2023 Feb.
Article en En | MEDLINE | ID: mdl-36354957
ABSTRACT

PURPOSE:

We developed and tested an automatic field-in-field (FIF) solution for whole-brain radiotherapy (WBRT) planning that creates a homogeneous dose distribution by minimizing hotspots, resulting in clinically acceptable plans.

METHODS:

A configurable auto-planning algorithm was developed to automatically generate FIF WBRT plans independent of the treatment planning system. Configurable parameters include the definition of hotspots, target volume, maximum number of subfields, and minimum number of monitor units per field. This algorithm iteratively identifies a hotspot, creates two opposing subfields, calculates the dose, and optimizes the beam weight based on user-configured constraints of dose-volume histogram coverage and least-squared cost functions. The algorithm was retrospectively tested on 17 whole-brain patients. First, an in-house landmark-based automated beam aperture technique was used to generate the treatment fields and initial plans. Second, the FIF algorithm was employed to optimize the plans using physician-defined goals of 99.9% of the brain volume receiving 100% of the prescription dose (30 Gy in 10 fractions) and a target hotspot definition of 107% of the prescription dose. The final auto-optimized plans were assessed for clinical acceptability by an experienced radiation oncologist using a five-point scale.

RESULTS:

The FIF algorithm reduced the mean (± SD) plan hotspot percentage dose from 35.0 Gy (116.6%) ± 0.6 Gy (2.0%) to 32.6 Gy (108.8%) ± 0.4 Gy (1.2%). Also, it decreased the mean (± SD) hotspot V107% [cm3 ] from 959 ± 498 cm3 to 145 ± 224 cm3 . On average, plans were produced in 16 min without any user intervention. Furthermore, 76.5% of the auto-plans were clinically acceptable (needing no or minor stylistic edits), and all of them were clinically acceptable after minor clinically necessary edits.

CONCLUSIONS:

This algorithm successfully produced high-quality WBRT plans and can improve treatment planning efficiency when incorporated into an automatic planning workflow.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radioterapia de Intensidad Modulada Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radioterapia de Intensidad Modulada Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos